Aliquoting on the centrifugal microfluidic platform based on centrifugo-pneumatic valves

被引:79
|
作者
Mark, Daniel [1 ]
Weber, Patrick [2 ]
Lutz, Sascha [1 ]
Focke, Maximilian [2 ]
Zengerle, Roland [1 ,2 ,3 ]
von Stetten, Felix [1 ,2 ]
机构
[1] HSG IMIT, D-78052 Villingen Schwenningen, Germany
[2] Univ Freiburg, Dept Microsyst Engn IMTEK, Lab MEMS Applicat, D-79110 Freiburg, Germany
[3] Univ Freiburg, Ctr Biol Signalling Studies BIOSS, D-79110 Freiburg, Germany
关键词
Lab-on-a-chip; Centrifugal microfluidics; Aliquoting; Multiplexing; PCR; Pneumatic; POLYMERASE-CHAIN-REACTION; WHOLE-BLOOD ANALYZER; LAB;
D O I
10.1007/s10404-010-0759-0
中图分类号
TB3 [工程材料学];
学科分类号
0805 ; 080502 ;
摘要
We present a new method for aliquoting liquids on the centrifugal microfluidic platform. Aliquoting is an essential unit operation to perform multiple parallel assays ("geometric multiplexing") from one individual sample, such as genotyping by real-time polymerase chain reactions (PCR), or homogeneous immunoassay panels. Our method is a two-stage process with an initial metering phase and a subsequent transport phase initiated by switching a centrifugo-pneumatic valve. The method enables aliquoting liquids into completely separated reaction cavities. It includes precise metering that is independent on the volume of pre-stored reagents in the receiving cavities. It further excludes any cross-contamination between the receiving cavities. We characterized the performance for prototypes fabricated by three different technologies: micro-milling, thermoforming of foils, and injection molding. An initial volume of similar to 90 mu l was split into 8 aliquots of 10 mu l volume each plus a waste reservoir on a thermoformed foil disk resulting in a coefficient of variation (CV) of the metered volumes of 3.6%. A similar volume of similar to 105 mu l was split into 16 aliquots of 6 mu l volume each on micro-milled and injection-molded disks and the corresponding CVs were 2.8 and 2.2%, respectively. Thus, the compatibility of the novel aliquoting structure to the aforementioned prototyping and production technologies is demonstrated. Additionally, the important question of achievable volume precision of the aliquoting structure with respect to the production tolerances inherent to each of these production technologies is addressed experimentally and theoretically. The new method is amenable to low cost mass production, since it does not require any post-replication surface modifications like hydrophobic patches.
引用
收藏
页码:1279 / 1288
页数:10
相关论文
共 50 条
  • [11] A CENTRIFUGO-PNEUMATIC CASCADE FOR FULLY INTEGRATED AND MULTIPLEXED BIOLOGICAL ANALYSIS
    Godino, Neus
    Gorkin, Robert, III
    Linares, Ana V.
    Burger, Robert
    Ducree, Jens
    2012 IEEE 25TH INTERNATIONAL CONFERENCE ON MICRO ELECTRO MECHANICAL SYSTEMS (MEMS), 2012,
  • [12] Comprehensive integration of homogeneous bioassays via centrifugo-pneumatic cascading
    Godino, Neus
    Gorkin, Robert, III
    Linares, Ana V.
    Burger, Robert
    Ducree, Jens
    LAB ON A CHIP, 2013, 13 (04) : 685 - 694
  • [13] TRANSIENT MODELLING OF PNEUMATIC VALVES IN CENTRIFUGAL MICROFLUIDIC DEVICES
    Moon, Michael R.
    Lin, Lin
    PROCEEDINGS OF THE ASME INTERNATIONAL MECHANICAL ENGINEERING CONGRESS AND EXPOSITION, 2016, VOL. 7, 2017,
  • [14] Aliquoting structure for centrifugal microfluidics based on a new pneumatic valve
    Mark, D.
    Haeberle, S.
    Metz, T.
    Lutz, S.
    Ducree, J.
    Zengerle, R.
    von Stetten, F.
    MEMS 2008: 21ST IEEE INTERNATIONAL CONFERENCE ON MICRO ELECTRO MECHANICAL SYSTEMS, TECHNICAL DIGEST, 2008, : 611 - +
  • [15] WIRELESS CLOSED-LOOP CONTROL OF CENTRIFUGO-PNEUMATIC VALVING TOWARDS LARGE-SCALE MICROFLUIDIC PROCESS INTEGRATION
    Kinahan, David J.
    Delgado, Sarai M.
    Julius, Lourdes A. N.
    Mallette, Adam
    Saenz-Ardila, David
    Mishra, Rohit
    Miyazaki, Celina M.
    Korvink, Jan
    Mager, Dario
    Ducree, Jens
    2018 IEEE MICRO ELECTRO MECHANICAL SYSTEMS (MEMS), 2018, : 1213 - 1216
  • [16] Reversible thermo-pneumatic valves on centrifugal microfluidic platforms
    Aeinehvand, Mohammad Mahdi
    Ibrahim, Fatimah
    Harun, Sulaiman Wadi
    Kazemzadeh, Amin
    Rothan, Hussin A.
    Yusof, Rohana
    Madou, Marc
    LAB ON A CHIP, 2015, 15 (16) : 3358 - 3369
  • [17] Passive flow switching valves on a centrifugal microfluidic platform
    Kim, Jitae
    Kido, Horacio
    Rangel, Roger H.
    Madou, Marc J.
    SENSORS AND ACTUATORS B-CHEMICAL, 2008, 128 (02) : 613 - 621
  • [18] Baking Powder Actuated Centrifugo-Pneumatic Valving for Automation of Multi-Step Bioassays
    Kinahan, David J.
    Renou, Marine
    Kurzbuch, Dirk
    Kilcawley, Niamh A.
    Bailey, Eanna
    Glynn, Macdara T.
    McDonagh, Colette
    Ducree, Jens
    MICROMACHINES, 2016, 7 (10):
  • [19] Centrifugo-Pneumatic Reciprocating Flowing Coupled with a Spatial Confinement Strategy for an Ultrafast Multiplexed Immunoassay
    Qian, Chungen
    Li, Pengjie
    Wang, Jingjing
    Hong, Xianzhe
    Zhao, Xudong
    Wu, Liqiang
    Miao, Zeyu
    Du, Wei
    Feng, Xiaojun
    Li, Yiwei
    Chen, Peng
    Liu, Bi-Feng
    ANALYTICAL CHEMISTRY, 2024, 96 (18) : 7145 - 7154
  • [20] Fluidic Automation of Nitrate and Nitrite Bioassays in Whole Blood by Dissolvable-Film Based Centrifugo-Pneumatic Actuation
    Nwankire, Charles E.
    Chan, Di-Sien S.
    Gaughran, Jennifer
    Burger, Robert
    Gorkin, Robert, III
    Ducree, Jens
    SENSORS, 2013, 13 (09) : 11336 - 11349