Magnetic resonance imaging in patients with cardiac implantable electronic devices: A prospective study

被引:6
|
作者
Navarro-Valverde, Cristina [1 ]
Ramos-Maqueda, Javier [2 ]
Jose Romero-Reyes, M. [1 ]
Esteve-Ruiz, Irene [3 ]
Garcia-Medina, Dolores [1 ]
Pavon-Jimenez, Ricardo [1 ]
Rodriguez-Gomez, Cristobal [4 ]
Leal-Del-Ojo, Juan [5 ]
Cayuela, Aurelio [6 ]
Molano-Casimiro, Francisco J. [1 ]
机构
[1] Hosp Univ Valme, Dept Cardiol, Seville, Spain
[2] Hosp Clin Univ Lozano Blesa, Dept Cardiol, Arrhythmias Unit, Zaragoza, Spain
[3] Hosp Juan Ramon Jimenez, Dept Cardiol, Huelva, Spain
[4] Hosp Univ Valme, Dept Radiol, Seville, Spain
[5] Hosp Fatima, Dept Cardiol, Seville, Spain
[6] Hosp Univ Valme, Dept Publ Hlth Prevent & Hlth Promot, Seville, Spain
关键词
Pacemaker Implantable cardioverter-defibrillator; Magnetic resonance imaging; Safety; CARDIOVERTER-DEFIBRILLATORS; PACEMAKER; SAFETY; MRI; UTILITY; STRATEGIES; CARDIOLOGY; STATEMENT; ARTIFACT; PROTOCOL;
D O I
10.1016/j.mri.2022.05.004
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Background: The number of patients with cardiac implantable electronic devices in whom magnetic resonance imaging (MRI) is indicated is constantly increasing. The potential risk of electromagnetic interference has limited its use and it is still contraindicated by the Food and Drug Administration in some cases. The aim of this study is to evaluate the safety and efficacy of MRI in these patients. Methods: A prospective registry comprising patients with a pacemaker (PM) or implantable cardioverterdefibrillator (ICD), MRI-conditional or not, who were candidates for MRI (at 1.5 T) with no suitable alternative diagnostic technique. All devices were programmed before the procedure and patients were monitored throughout the test. Clinical, electrical, and technical parameters were evaluated before and after MRI. Results: 147 MRI examinations (132 PM and 15 ICD) were performed. There were no clinical events or significant differences in the electrical parameters of the leads after MRI. A variation in the impedance of the ventricular leads was detected, although the difference was not clinically relevant. In one patient with a PM, a failure in release of the safety impulse was detected in the auto-threshold test, although the threshold was correctly determined. In 11 of the 17 thoracic MRIs, image artifacts were detected, preventing the diagnosis in two of them. Conclusions: In patients with cardiac implantable electronic devices, MRIs performed under a specific protocol has been shown to be safe in the short term even in the thoracic region, as well as interpretable in most cases.
引用
收藏
页码:9 / 15
页数:7
相关论文
共 50 条
  • [1] Magnetic resonance imaging in patients with cardiac implantable electronic devices
    Maass, A. H.
    Hemels, M. E. W.
    Allaart, C. P.
    [J]. NETHERLANDS HEART JOURNAL, 2018, 26 (12) : 584 - 590
  • [2] Magnetic resonance imaging in patients with cardiac implantable electronic devices
    A. H. Maass
    M. E. W. Hemels
    C. P. Allaart
    [J]. Netherlands Heart Journal, 2018, 26 : 584 - 590
  • [3] Magnetic Resonance Imaging in Patients With Cardiac Implantable Electronic Devices
    Eyal, Allon
    Roguin, Ariel
    [J]. CIRCULATION, 2015, 132 (14) : E176 - E178
  • [4] Magnetic resonance imaging in patients with cardiac implantable electronic devices: a single-center prospective study
    Yadava, Mrinal
    Nugent, Matthew
    Krebsbach, Angela
    Minnier, Jessica
    Jessel, Peter
    Henrikson, Charles A.
    [J]. JOURNAL OF INTERVENTIONAL CARDIAC ELECTROPHYSIOLOGY, 2017, 50 (01) : 95 - 104
  • [5] Magnetic resonance imaging in patients with cardiac implantable electronic devices: a single-center prospective study
    Mrinal Yadava
    Matthew Nugent
    Angela Krebsbach
    Jessica Minnier
    Peter Jessel
    Charles A. Henrikson
    [J]. Journal of Interventional Cardiac Electrophysiology, 2017, 50 : 95 - 104
  • [6] Magnetic resonance imaging safety in patients with cardiac implantable electronic devices
    Yang, Eunice
    Suzuki, Masahito
    Nazarian, Saman
    Halperin, Henry R.
    [J]. TRENDS IN CARDIOVASCULAR MEDICINE, 2022, 32 (07) : 440 - 447
  • [7] The Safety of Cardiac and Thoracic Magnetic Resonance Imaging in Patients with Cardiac Implantable Electronic Devices
    Dandamudi, Sanjay
    Collins, Jeremy D.
    Carr, James C.
    Mongkolwat, Pat
    Rahsepar, Amir A.
    Tomson, Todd T.
    Verma, Nishant
    Arora, Rishi
    Chicos, Alex B.
    Kim, Susan S.
    Lin, Albert C.
    Passman, Rod S.
    Knight, Bradley P.
    [J]. ACADEMIC RADIOLOGY, 2016, 23 (12) : 1498 - 1505
  • [8] Safety and utility of magnetic resonance imaging in patients with cardiac implantable electronic devices
    Strom, Jordan B.
    Whelan, Jill B.
    Shen, Changyu
    Zheng, Shuang Qi
    Mortele, Koenraad J.
    Kramer, Daniel B.
    [J]. HEART RHYTHM, 2017, 14 (08) : 1138 - 1144
  • [9] Magnetic Resonance Imaging in Patients With Cardiac Implantable Electronic Devices With Abandoned Leads
    Schaller, Robert D.
    Brunker, Tamara
    Riley, Michael P.
    Marchlinski, Francis E.
    Nazarian, Saman
    Litt, Harold
    [J]. JAMA CARDIOLOGY, 2021, 6 (05) : 549 - 556
  • [10] Implantable Electronic Cardiac Devices and Compatibility With Magnetic Resonance Imaging
    Miller, Jared D.
    Nazarian, Saman
    Halperin, Henry R.
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2016, 68 (14) : 1590 - 1598