Impact of gap size and groove design of hydrodynamic bearing on plasma skimming effect for use in rotary blood pump

被引:2
|
作者
Jiang, Ming [1 ,2 ]
Sakota, Daisuke [2 ]
Kosaka, Ryo [2 ]
Hijikata, Wataru [1 ]
机构
[1] Tokyo Inst Technol, Sch Engn, Dept Mech Engn, Meguro Ku, 2-12-1 Ookayama, Tokyo 1528550, Japan
[2] Natl Inst Adv Ind Sci & Technol, Artificial Organ Res Grp, Hlth & Med Res Inst, 1-2-1 Namiki, Tsukuba, Ibaraki 3058564, Japan
关键词
Hemolysis; Hydrodynamic bearing; Plasma skimming; Rotary blood pump; Spiral groove pattern; HEMOCOMPATIBILITY; IMPROVEMENT; REDUCTION; CELLS; FLOW;
D O I
10.1007/s10047-021-01308-x
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Plasma skimming can exclude red blood cells from high shear regions in the gaps formed by hydrodynamic bearings in rotary blood pumps. We investigated the effect of the gap size and groove design on the plasma skimming efficiency. Spiral groove bearings (SGBs) were installed into a specially designed test rig for in vitro experiments performed using human blood. The measured gap between the ridges of the bearing and the rotor surface was 17-26 mu m at a flow rate of 150 ml/min and a rotor speed of 2400 rpm. Three different patterns of SGBs were designed (SGB-0, SGB-30, and SGB-60) with various degrees of the circumferential component. The hematocrit measured by a high-speed camera was compared with the hematocrit in the circuit, and the plasma skimming efficiency for the three bearing patterns was evaluated at hematocrits of 20%, 25%, and 30%. SGB-60, which had the strongest circumferential component, provided the best plasma skimming efficiency. When the gap size was less than 20 mu m, the red blood cells in the gaps between the ridges of the bearing and rotor surface reduced significantly and the efficiency became higher than 90%. The gap size had the strongest effect on producing a significant plasma skimming. The plasma skimming efficiency can be significantly improved by optimizing the bearing gap size and groove design, which facilitates the further development of SGBs for use in applications such as rotary blood pumps.
引用
收藏
页码:195 / 203
页数:9
相关论文
共 19 条
  • [1] Impact of gap size and groove design of hydrodynamic bearing on plasma skimming effect for use in rotary blood pump
    Ming Jiang
    Daisuke Sakota
    Ryo Kosaka
    Wataru Hijikata
    [J]. Journal of Artificial Organs, 2022, 25 : 195 - 203
  • [2] Analysis of Plasma Skimming within a Hydrodynamic Bearing Gap for Designing Spiral Groove Bearings in Rotary Blood Pumps
    Jiang, M.
    Sakota, D.
    Kosaka, R.
    Hijikata, W.
    [J]. 2021 43RD ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE & BIOLOGY SOCIETY (EMBC), 2021, : 1213 - 1217
  • [3] Spiral groove bearing design for improving plasma skimming in rotary blood pumps
    Jiang, Ming
    Hijikata, Wataru
    [J]. JOURNAL OF ARTIFICIAL ORGANS, 2024, 27 (03) : 212 - 221
  • [4] Plasma Skimming in a Spiral Groove Bearing of a Centrifugal Blood Pump
    Murashige, Tomotaka
    Sakota, Daisuke
    Kosaka, Ryo
    Nishida, Masahiro
    Kawaguchi, Yasuo
    Yamane, Takashi
    Maruyama, Osamu
    [J]. ARTIFICIAL ORGANS, 2016, 40 (09) : 856 - 866
  • [5] Plasma skimming efficiency of human blood in the spiral groove bearing of a centrifugal blood pump
    Sakota, Daisuke
    Kondo, Kazuki
    Kosaka, Ryo
    Nishida, Masahiro
    Maruyama, Osamu
    [J]. JOURNAL OF ARTIFICIAL ORGANS, 2021, 24 (02) : 126 - 134
  • [6] Plasma skimming efficiency of human blood in the spiral groove bearing of a centrifugal blood pump
    Daisuke Sakota
    Kazuki Kondo
    Ryo Kosaka
    Masahiro Nishida
    Osamu Maruyama
    [J]. Journal of Artificial Organs, 2021, 24 : 126 - 134
  • [7] A Novel Design of Spiral Groove Bearing in a Hydrodynamically Levitated Centrifugal Rotary Blood Pump
    Han, Qing
    Zou, Jun
    Ruan, Xiaodong
    Fu, Xin
    Yang, Huayong
    [J]. ARTIFICIAL ORGANS, 2012, 36 (08) : 739 - 746
  • [8] DEVELOPMENT OF DURABLE ROTARY BLOOD PUMP SYSTEMS USING HYDRODYNAMIC BEARING TECHNOLOGY
    Tsukiya, T.
    Mizuno, T.
    Takewa, Y.
    Tatsumi, E.
    Okubo, T.
    Osada, T.
    Yamane, T.
    Taenaka, Y.
    [J]. INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 2011, 34 (08): : 660 - 660
  • [9] A comparative study of structural parameters for spiral groove bearing in centrifugal rotary blood pump
    Xue, Cong
    Chen, Yu
    Zhu, Bin
    Wang, Xiuying
    [J]. HELIYON, 2023, 9 (03)
  • [10] The Spiral Groove Bearing as a Mechanism for Enhancing the Secondary Flow in a Centrifugal Rotary Blood Pump
    Amaral, Felipe
    Gross-Hardt, Sascha
    Timms, Daniel
    Egger, Christina
    Steinseifer, Ulrich
    Schmitz-Rode, Thomas
    [J]. ARTIFICIAL ORGANS, 2013, 37 (10) : 866 - 874