Plasma skimming efficiency of human blood in the spiral groove bearing of a centrifugal blood pump

被引:6
|
作者
Sakota, Daisuke [1 ]
Kondo, Kazuki [1 ]
Kosaka, Ryo [1 ]
Nishida, Masahiro [1 ]
Maruyama, Osamu [1 ]
机构
[1] Natl Inst Adv Ind Sci & Technol, Hlth & Med Res Inst, Artificial Organ Res Grp, 1-2-1 Namiki, Tsukuba, Ibaraki 3058564, Japan
关键词
Plasma skimming; Hydrodynamic bearing; Centrifugal blood pump; Hemolysis; Optics of blood; COMPUTATIONAL FLUID-DYNAMICS; THROMBUS GROWTH; HEMOLYSIS; CELLS; SCATTERING;
D O I
10.1007/s10047-020-01221-9
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
This work investigates the plasma skimming effect in a spiral groove bearing within a hydrodynamically levitated centrifugal blood pump when working with human blood having a hematocrit value from 0 to 40%. The present study assessed the evaluation based on a method that clarified the limitations associated with such assessments. Human blood was circulated in a closed-loop circuit via a pump operating at 4000 rpm at a flow rate of 5 L/min. Red blood cells flowing through a ridge area of the bearing were directly observed using a high-speed microscope. The hematocrit value in the ridge area was calculated using the mean corpuscular volume, the bearing gap, the cross-sectional area of a red blood cell, and the occupancy of red blood cells. The latter value was obtained from photographic images by dividing the number of pixels showing red blood cells in the evaluation area by the total number of pixels in this area. The plasma skimming efficiency was calculated as the extent to which the hematocrit of the working blood was reduced in the ridge area. For the hematocrit in the circuit from 0 to 40%, the plasma skimming efficiency was approximately 90%, meaning that the hematocrit in the ridge area became 10% as compared to that in the circuit. For a hematocrit of 20% and over, red blood cells almost completely occupied the ridge. Thus, a valid assessment of plasma skimming was only possible when the hematocrit was less than 20%.
引用
收藏
页码:126 / 134
页数:9
相关论文
共 50 条
  • [1] Plasma skimming efficiency of human blood in the spiral groove bearing of a centrifugal blood pump
    Daisuke Sakota
    Kazuki Kondo
    Ryo Kosaka
    Masahiro Nishida
    Osamu Maruyama
    [J]. Journal of Artificial Organs, 2021, 24 : 126 - 134
  • [2] Plasma Skimming in a Spiral Groove Bearing of a Centrifugal Blood Pump
    Murashige, Tomotaka
    Sakota, Daisuke
    Kosaka, Ryo
    Nishida, Masahiro
    Kawaguchi, Yasuo
    Yamane, Takashi
    Maruyama, Osamu
    [J]. ARTIFICIAL ORGANS, 2016, 40 (09) : 856 - 866
  • [3] Spiral groove bearing design for improving plasma skimming in rotary blood pumps
    Jiang, Ming
    Hijikata, Wataru
    [J]. JOURNAL OF ARTIFICIAL ORGANS, 2024, 27 (03) : 212 - 221
  • [4] A Novel Design of Spiral Groove Bearing in a Hydrodynamically Levitated Centrifugal Rotary Blood Pump
    Han, Qing
    Zou, Jun
    Ruan, Xiaodong
    Fu, Xin
    Yang, Huayong
    [J]. ARTIFICIAL ORGANS, 2012, 36 (08) : 739 - 746
  • [5] A comparative study of structural parameters for spiral groove bearing in centrifugal rotary blood pump
    Xue, Cong
    Chen, Yu
    Zhu, Bin
    Wang, Xiuying
    [J]. HELIYON, 2023, 9 (03)
  • [6] The Spiral Groove Bearing as a Mechanism for Enhancing the Secondary Flow in a Centrifugal Rotary Blood Pump
    Amaral, Felipe
    Gross-Hardt, Sascha
    Timms, Daniel
    Egger, Christina
    Steinseifer, Ulrich
    Schmitz-Rode, Thomas
    [J]. ARTIFICIAL ORGANS, 2013, 37 (10) : 866 - 874
  • [7] Analysis of Plasma Skimming within a Hydrodynamic Bearing Gap for Designing Spiral Groove Bearings in Rotary Blood Pumps
    Jiang, M.
    Sakota, D.
    Kosaka, R.
    Hijikata, W.
    [J]. 2021 43RD ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE & BIOLOGY SOCIETY (EMBC), 2021, : 1213 - 1217
  • [8] Impact of gap size and groove design of hydrodynamic bearing on plasma skimming effect for use in rotary blood pump
    Jiang, Ming
    Sakota, Daisuke
    Kosaka, Ryo
    Hijikata, Wataru
    [J]. JOURNAL OF ARTIFICIAL ORGANS, 2022, 25 (03) : 195 - 203
  • [9] Impact of gap size and groove design of hydrodynamic bearing on plasma skimming effect for use in rotary blood pump
    Ming Jiang
    Daisuke Sakota
    Ryo Kosaka
    Wataru Hijikata
    [J]. Journal of Artificial Organs, 2022, 25 : 195 - 203
  • [10] Experimental study of dynamic characteristics of a centrifugal blood pump with a conical spiral groove bearing for a ventricular assist device
    Nakamura, Youichi
    Tsukamoto, Hiroshi
    Miyazaki, Koji
    [J]. FEDSM 2007: PROCEEDINGS OF THE 5TH JOINT AMSE/JSME FLUIDS ENGINEERING SUMMER CONFERENCE VOL 1, PTS A AND B, 2007, : 1187 - 1194