Impurity toroidal rotation profile measurement using upgraded high-resolution visible spectroscopic diagnostic on ADITYA-U tokamak

被引:7
|
作者
Shukla, G. [1 ]
Chowdhuri, M. B. [2 ]
Shah, K. [1 ]
Manchanda, R. [2 ]
Ramaiya, N. [2 ]
Tanna, R. L. [2 ]
Mayya, K. B. K. [1 ]
Ghosh, J. [2 ]
机构
[1] Pandit Deendayal Petr Univ, Dept Sci, Gandhinagar 382007, India
[2] Inst Plasma Res, Bhat 382428, Gandhinagar, India
来源
REVIEW OF SCIENTIFIC INSTRUMENTS | 2021年 / 92卷 / 06期
关键词
ION TEMPERATURE;
D O I
10.1063/5.0043789
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
A high-resolution spectroscopic diagnostic for the measurement of spatial profiles of impurity ion toroidal rotation velocities on the ADITYA-U tokamak has been upgraded to cover the complete plasma minor radius. Earlier, the coverage of diagnostics toward the plasma edge was restricted due to the placement of collection optics on the tangential port outside the vacuum vessel. The coverage of the full plasma minor radius, from 0 to 24 cm, has been achieved using the newly designed and developed collection optics that have seven lines of sight to view the tokamak plasma mounted inside a customized re-entrant view port which is installed in the shadow of the limiter inside the vacuum vessel. The upgraded diagnostic also includes a faster charged coupled device detector with a smaller pixel size for the detection of a small wavelength shift. The complete spatial profile has been measured using the Doppler shifted passive change exchange spectral line at 529.0 nm from the C5+ ion. In this article, we present the collection optics' design, installation, calibration, and results obtained using the upgraded diagnostic. Published under an exclusive license by AIP Publishing.
引用
收藏
页数:5
相关论文
共 25 条
  • [21] Mapping an Extended Neurochemical Profile at 3 and 7 T Using Accelerated High-Resolution Proton Magnetic Resonance Spectroscopic Imaging
    Gruber, Stephan
    Heckova, Eva
    Strasser, Bernhard
    Povazan, Michal
    Hangel, Gilbert J.
    Minarikova, Lenka
    Trattnig, Siegfried
    Bogner, Wolfgang
    INVESTIGATIVE RADIOLOGY, 2017, 52 (10) : 631 - 639
  • [22] Compton-scattering measurement of diagnostic x-ray spectrum using high-resolution Schottky CdTe detector
    Maeda, K
    Matsumoto, M
    Taniguchi, A
    MEDICAL PHYSICS, 2005, 32 (06) : 1542 - 1547
  • [23] High-Resolution Magnetic Resonance Spectroscopic Imaging using a Multi-Encoder Attention U-Net with Structural and Adversarial Loss
    Dong, Siyuan
    Hangel, Gilbert
    Bogner, Wolfgang
    Trattnig, Siegfried
    Roessler, Karl
    Widhalm, Georg
    De Feyter, Henk M.
    de Graaf, Robin A.
    Duncan, James S.
    2021 43RD ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE & BIOLOGY SOCIETY (EMBC), 2021, : 2891 - 2895
  • [24] High-resolution and nondestructive profile measurement by spectral-domain optical coherence tomography with a visible broadband light source for optical-device fabrication
    Nishi, Tsuyoshi
    Ozaki, Nobuhiko
    Oikawa, Yoichi
    Miyaji, Kunio
    Ohsato, Hirotaka
    Watanabe, Eiichiro
    Ikeda, Naoki
    Sugimoto, Yoshimasa
    JAPANESE JOURNAL OF APPLIED PHYSICS, 2016, 55 (08)
  • [25] Measurement of Toxaphene Congeners in Pooled Human Serum Collected in Three U.S. Cities Using High-Resolution Mass Spectrometry
    John R. Barr
    Adrian R. Woolfitt
    Vincent L. Maggio
    Donald G. Patterson
    Archives of Environmental Contamination and Toxicology, 2004, 46 : 551 - 556