Evaluation of a commercial electron treatment planning system based on Monte Carlo techniques (eMC)

被引:73
|
作者
Pemler, Peter [1 ]
Besserer, Juergen [1 ]
Schneider, Uwe [1 ]
Neuenschwander, Hans [2 ]
机构
[1] Stadtspital Triemli, Klin Radioonkol & Nukl Med, CH-8063 Zurich, Switzerland
[2] Lindenhofspital, Klin Radioonkol, Bern, Switzerland
来源
ZEITSCHRIFT FUR MEDIZINISCHE PHYSIK | 2006年 / 16卷 / 04期
关键词
Monte Carlo; treatment planning; electrons; commissioning;
D O I
10.1078/0939-3889-00330
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A commercial electron beam treatment planning system on the basis of a Monte Carlo algorithm (Varian Eclipse, eMC V7.2.35) was evaluated. Measured dose distributions were used for comparison with dose distributions predicted by eMC calculations. Tests were carried out for various applicators and field sizes, irregular shaped cut outs and an inhomogeneity phantom for energies between 6 MeV and 22 MeV. Monitor units were calculated for all applicator/energy combinations and field sizes down to 3 cm diameter and source-to-surface distances of 100 cm and 110 cm. A mass-density-to-Hounsfield-Units calibration was performed to compare dose distributions calculated with a default and an individual calibration. The relationship between calculation parameters of the eMC and the resulting dose distribution was studied in detail. Finally, the algorithm was also applied to a clinical case (boost treatment of the breast) to reveal possible problems in the implementation. For standard geometries there was a good agreement between measurements and calculations, except for profiles for low energies (6 MeV) and high energies (18 MeV, 22 MeV), in which cases the algorithm overestimated the dose off-axis in the high-dose region. For energies of 12 MeV and higher, there were oscillations in the plateau region of the corresponding depth dose curves calculated with a grid size of 1 mm. With irregular cut outs, an overestimation of the dose was observed for small slits and low energies (4% for 6 MeV), as well as for asymmetric cases and extended source-to-surface distances (12% for SSD = 120 cm). While all monitor unit calculations for SSD = 100 cm were within 3% compared to measure-ments, there were large deviations for small cut outs and source-to-surface distances larger than 100 cm (7% for a 3 cm diameter cutout and a source-to-surface distance of 110 cm).
引用
收藏
页码:313 / 329
页数:17
相关论文
共 50 条
  • [21] Dosimetric validation of a commercial Monte Carlo based IMRT planning system
    Grofsmid, Dennis
    Dirkx, Maarten
    Marijnissen, Hans
    Woudstra, Evert
    Heijmen, Ben
    [J]. MEDICAL PHYSICS, 2010, 37 (02) : 540 - 549
  • [22] GPU-Based Monte Carlo Treatment Planning System for Electron FLASH Radiotherapy
    Zhou, B.
    Lu, W.
    Lai, Y.
    Rahman, M.
    Zhang, R.
    Jia, X.
    Wang, K.
    [J]. MEDICAL PHYSICS, 2022, 49 (06) : E127 - E127
  • [23] Evaluation of an electron Monte Carlo dose calculation algorithm for treatment planning
    Chamberland, Eve
    Beaulieu, Luc
    Lachance, Bernard
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2015, 16 (03): : 60 - 79
  • [24] Evaluation of the Monte Carlo Dose Calculation Engine of Eclipse Treatment Planning System for Electron Beams
    Kamta, G. Lagmago
    Reniers, B.
    [J]. MEDICAL PHYSICS, 2008, 35 (06)
  • [25] COMMISSIONING AND VERIFICATION OF A COMMERCIAL PHOTON MONTE CARLO TREATMENT PLANNING SYSTEM.
    Behrens, C.
    Ottosson, R.
    Karlsson, A.
    Samsoe, E.
    Sjostrom, D.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2009, 92 : S223 - S223
  • [26] RETROSPECTIVE ANALYSIS OF COMMERCIAL MONTE CARLO TREATMENT PLANNING SYSTEM FOR LUNG CANCER
    Miura, H.
    Norihisa, M.
    Oh, R. J.
    Shiomi, H.
    Norihiko, K.
    Toshihiko, I.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2011, 99 : S369 - S369
  • [27] Verification of a commercial implementation of VMC++ for Monte Carlo electron beam treatment planning
    Cygler, J
    Daskalov, G
    Chan, G
    Ding, G
    [J]. MEDICAL PHYSICS, 2002, 29 (06) : 1351 - 1352
  • [28] Commissioning of Electron Monte Carlo in Eclipse Treatment Planning System for TrueBeam
    Yang, X.
    Lasio, G.
    Zhou, J.
    Lin, M.
    Yi, B.
    Guerrero, M.
    [J]. MEDICAL PHYSICS, 2014, 41 (06) : 362 - 362
  • [29] Dosimetric verification of a commercial Monte Carlo treatment planning system (VMC++) for a 9 MeV electron beam
    Schiapparelli, P.
    Zefiro, D.
    Taccini, G.
    [J]. MEDICAL PHYSICS, 2009, 36 (05) : 1759 - 1767
  • [30] DEVELOPMENT OF A MULTIMODAL MONTE CARLO BASED TREATMENT PLANNING SYSTEM
    Kumada, Hiroaki
    Takada, Kenta
    Sakurai, Yoshinori
    Suzuki, Minoru
    Takata, Takushi
    Sakurai, Hideyuki
    Matsumura, Akira
    Sakae, Takeji
    [J]. RADIATION PROTECTION DOSIMETRY, 2018, 180 (1-4) : 286 - 290