Monte Carlo calculations and experimental measurements of dosimetric parameters of the IRA-103Pd brachytherapy source

被引:21
|
作者
Sadeghi, Mahdi [1 ,2 ]
Raisali, Gholamreza [3 ]
Hosseini, S. Hamed [2 ]
Shavar, Arzhang [4 ]
机构
[1] Agr Med & Ind Res Sch, Nucl Med Res Grp, Karaj, Iran
[2] Islamic Azad Univ, Fac Engn, Tehran, Iran
[3] Radiat Applicat Res Sch, Nucl Sci & Technol Res Inst, Tehran, Iran
[4] Med & Ind Res Sch, SSDL Grp, Karaj, Iran
关键词
Monte Carlo calculations; brachytherapy dosimetry; TG-43; MCNP; Pd-103;
D O I
10.1118/1.2870229
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
This article presents a brachytherapy source having Pd-103 adsorbed onto a cylindrical silver rod that has been developed by the Agricultural, Medical, and Industrial Research School for permanent implant applications. Dosimetric characteristics (radial dose function, anisotropy function, and anisotropy factor) of this source were experimentally and theoretically determined in terms of the updated AAPM Task group 43 (TG-43U1) recommendations. Monte Carlo simulations were used to calculate the dose rate constant. Measurements were performed using TLD-GR200A circular chip dosimeters using standard methods employing thermoluminescent dosimeters in a Perspex phantom. Precision machined bores in the phantom located the dosimeters and the source in a reproducible fixed geometry, providing for transverse-axis and angular dose profiles over a range of distances from 0.5 to 5 cm. The Monte Carlo N-particle (MCNP) code, version 4C simulation techniques have been used to evaluate the dose-rate distributions around this model Pd-103 source in water and Perspex phantoms. The Monte Carlo calculated dose rate constant of the IRA-Pd-103 source in water was found to be 0.678 cGy h(-1) U-1 with an approximate uncertainty of +/- 0.1%. The anisotropy function, F(r,theta), and the radial dose function, g(r), of the IRA-Pd-103 source were also measured in a Perspex phantom and calculated in both Perspex and liquid water phantoms. (C) 2008 American Association of Physicists in Medicine.
引用
收藏
页码:1288 / 1294
页数:7
相关论文
共 50 条
  • [1] Monte Carlo calculations and experimental measurements of dosimetric parameters of the IRA-103Pd source
    Sadeghi, M.
    Raisali, G.
    Hoseini, H.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2007, 84 : S193 - S194
  • [2] Experimental measurements and Monte Carlo calculations of dosimetric parameters of the IRA1-103Pd brachytherapy source
    Sadeghi, Mahdi
    Hosseini, S. Hamed
    Raisali, Gholamreza
    [J]. APPLIED RADIATION AND ISOTOPES, 2008, 66 (10) : 1431 - 1437
  • [3] Monte Carlo calculations and experimental measurements of dosimetry parameters of a new 103Pd source
    Li, ZF
    Palta, JR
    Fan, JJ
    [J]. MEDICAL PHYSICS, 2000, 27 (05) : 1108 - 1112
  • [4] Experimental and Monte Carlo dosimetric characterization of a 1 cm 103Pd brachytherapy source
    Reed, Joshua L.
    Rivard, Mark J.
    Micka, John A.
    Culberson, Wesley S.
    DeWerd, Larry A.
    [J]. BRACHYTHERAPY, 2014, 13 (06) : 657 - 667
  • [5] Dosimetric parameters of the new design 103Pd brachytherapy source based on Monte Carlo study
    Saidi, Pooneh
    Sadeghi, Mahdi
    Shirazi, Alireza
    Tenreiro, Claudio
    [J]. PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2012, 28 (01): : 13 - 18
  • [6] Dosimetric parameters of palladium-103 brachytherapy source with Monte Carlo simulation
    WANG JianHua
    [J]. Science Bulletin, 2008, (09) : 1304 - 1309
  • [7] Dosimetric parameters of palladium-103 brachytherapy source with Monte Carlo simulation
    Wang JianHua
    Liu Wei
    Xu XunJiang
    Gu JiaHui
    Cai Jun
    Hua ZhengDong
    Xu JiaQiang
    [J]. CHINESE SCIENCE BULLETIN, 2008, 53 (09): : 1304 - 1309
  • [8] Monte Carlo dosimetric characterization of the IsoAid ADVANTAGE 103Pd brachytherapy source
    Sowards, Keith T.
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2007, 8 (02): : 18 - 25
  • [9] Monte Carlo and experimental characterization of the first AMIRS 103Pd brachytherapy source
    Raisali, Gholamreza
    Ghonchehnazi, Maryam G.
    Shokrani, Parvaneh
    Sadeghi, Mahdi
    [J]. APPLIED RADIATION AND ISOTOPES, 2008, 66 (12) : 1856 - 1860
  • [10] Dosimetric investigation of 103Pd permanent breast seed implant brachytherapy based on Monte Carlo calculations
    Deering, Stephen G.
    Hilts, Michelle
    Batchelar, Deidre
    Crook, Juanita
    Thomson, Rowan M.
    [J]. BRACHYTHERAPY, 2021, 20 (03) : 686 - 694