Experiences with packed capillary electrochromatography at ambient pressure

被引:103
|
作者
vandenBosch, SE [1 ]
Heemstra, S [1 ]
Kraak, JC [1 ]
Poppe, H [1 ]
机构
[1] UNIV AMSTERDAM,ANALYT CHEM LAB,AMSTERDAM INST MOL STUDIES,NL-1018 WV AMSTERDAM,NETHERLANDS
关键词
air bubbles; electrochromatography; capillary columns; packed capillary columns; packing retainers; frits; phenols; polynuclear aromatic hydrocarbons;
D O I
10.1016/S0021-9673(96)00612-7
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
The practical execution of packed capillary electrochromatography is evaluated. The most serious problem is the generation of air bubbles in the system. This occurs especially in frits which were fabricated by glueing particles differing from the packing particles. No air bubbles were generated with frits which were prepared by locally sintering a narrow band of the reversed-phase packing by means of a heating filament, while flushing the column with water. In this way electrochromatography can be performed simply without pressurizing the buffer vials, providing the mobile phase is frequently degassed with helium. The performance under electrically driven conditions appears to be excellent and for 5 mu m reversed-phase particles reduced plate heights down to 1.8 were achieved. Compared to pressure driven conditions, 25-40% smaller plate heights were obtained under electrically driven conditions.
引用
收藏
页码:165 / 177
页数:13
相关论文
共 50 条
  • [1] Capillary electrochromatography in columns packed by centripetal forces
    Fermier, AM
    Colon, LA
    JOURNAL OF MICROCOLUMN SEPARATIONS, 1998, 10 (05) : 439 - 447
  • [2] Aspects of column fabrication for packed capillary electrochromatography
    Angus, PDA
    Demarest, CW
    Catalano, T
    Stobaugh, JF
    JOURNAL OF CHROMATOGRAPHY A, 2000, 887 (1-2) : 347 - 365
  • [3] Comparison of separation behavior of benzodiazepines in packed capillary electrochromatography and open-tubular capillary electrochromatography
    Jinno, K
    Sawada, H
    Catabay, AP
    Watanabe, H
    Sabli, NBH
    Pesek, JJ
    Matyska, MT
    JOURNAL OF CHROMATOGRAPHY A, 2000, 887 (1-2) : 479 - 487
  • [4] Preparation of fritless packed silica columns for capillary electrochromatography
    Fujimoto, C
    HRC-JOURNAL OF HIGH RESOLUTION CHROMATOGRAPHY, 2000, 23 (01): : 89 - 92
  • [5] Capillary electrochromatography with packed bead beds in microfluidic devices
    Jemere, Abebaw B.
    Martinez, Dolores
    Finot, Michael
    Harrison, D. Jed
    ELECTROPHORESIS, 2009, 30 (24) : 4237 - 4244
  • [6] Capillary electrochromatography without external pressure assistance Use of packed columns with a monolithic inlet frit
    Rocco, Anna
    Fanali, Salvatore
    JOURNAL OF CHROMATOGRAPHY A, 2008, 1191 (1-2) : 263 - 267
  • [7] Enantiomer separation by CECEnantiomer separation by packed-capillary electrochromatography
    Jun Zhu
    Hanfa Zou
    Wei Shi
    Jianzhong Rui
    Jianyi Ni
    Yukui Zhang
    Science in China Series B: Chemistry, 1999, 42 : 511 - 519
  • [9] Perfusive Electroosmotic Transport in Packed Capillary Electrochromatography: Mechanism and Utility
    Department of Chemistry, West Virginia University, Morgantown, WV 26506-6045, United States
    J Microcolumn Sep, 5 (389-397):
  • [10] Enantiomer separation by capillary electrochromatography using fritless packed columns
    Fujimoto, C
    ANALYTICAL SCIENCES, 2002, 18 (01) : 19 - 25