The effects of polyethylene cup thickness on wear of total hip prostheses

被引:18
|
作者
Oonishi, H
Iwaki, H
Kin, N
Kushitani, S
Murata, N
Wakitani, S
Imoto, K
机构
[1] Osaka Minami Natl Hosp, Biomat Res Lab, Kawachinagano, Osaka 586, Japan
[2] Osaka Minami Natl Hosp, Dept Orthopaed Surg, Artificial Joint Sect, Kawachinagano, Osaka 586, Japan
关键词
Polymer; Alumina; Molecular Weight; Polyethylene; Femoral Head;
D O I
10.1023/A:1008823518381
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The wear rate of the ultrahigh molecular weight polyethylene cups in combination with a 28 mm alumina femoral head was measured on the radiographs of patients without any complications, and on cups retrieved due to slight loosening of the prosthesis or due to late infection between bone and components. The wear rate on the radiographs did not include the initial wear, but the wear rate on the retrieved cups did include the initial wear. The wear rate on the retrieved cups was higher, by 50%, than that measured on the radiographs. In both cases, the thicker the polyethylene cups, the lower was the wear rate measured. The average wear rate of the 7 and 8 mm thick cups was about twice that of cups 10 and 11 mm thick. From these results, we conclude that: polyethylene cups more than 11 mm thick should be used.
引用
收藏
页码:475 / 478
页数:4
相关论文
共 50 条
  • [1] The effects of polyethylene cup thickness on wear of total hip prostheses
    H Oonishi
    H Iwaki
    N Kin
    S Kushitani
    N Murata
    S Wakitani
    K Imoto
    [J]. Journal of Materials Science: Materials in Medicine, 1998, 9 : 475 - 478
  • [2] Retrieved total hip prostheses - Part I - The effects of cup thickness, head sizes and fusion defects on wear
    Oonishi, H
    Tsuji, E
    Kim, YY
    [J]. JOURNAL OF MATERIALS SCIENCE-MATERIALS IN MEDICINE, 1998, 9 (07) : 393 - 401
  • [3] Retrieved total hip prostheses. Part I: The effects of cup thickness, head sizes and fusion defects on wear
    H. Oonishi
    E. Tsuji
    Y. Y. Kim
    [J]. Journal of Materials Science: Materials in Medicine, 1998, 9 : 393 - 401
  • [4] Wear behaviour of polyethylene cup against 28mm alumina ball in total hip prostheses
    Oonishi, H
    Murata, N
    Kushitani, S
    Wakitani, S
    Imoto, K
    Iwaki, Y
    Kin, N
    [J]. BIOCERAMICS, VOL 10, 1997, : 333 - 336
  • [5] The effect of total hip arthroplasty cup design on polyethylene wear rate
    Hamilton, WG
    Hopper, RH
    Ginn, SD
    Hammell, NP
    Engh, CA
    Engh, CA
    [J]. JOURNAL OF ARTHROPLASTY, 2005, 20 (07): : 63 - 72
  • [6] Polyethylene wear of the cup in total hip replacement with Muller reinforcement ring
    Sirka, Aurimas
    Clauss, Martin
    Tarasevicius, Sarunas
    Wingstrand, Hans
    Stucinskas, Justinas
    Ilchmann, Thomas
    [J]. SWISS MEDICAL WEEKLY, 2015, 145 : 37S - 37S
  • [7] Wear of the polyethylene acetabular cup - The effect of head material, head diameter, and cup thickness studied with a hip simulator
    Saikko, VO
    [J]. ACTA ORTHOPAEDICA SCANDINAVICA, 1995, 66 (06): : 501 - 506
  • [8] Polyethylene wear in total hip prostheses: the influence of direction of linear wear on volumetric wear determined from radiographic data
    Kosak, R
    Antolic, V
    Pavlovcic, V
    Kralj-Iglic, V
    Milosev, I
    Vidmar, G
    Iglic, A
    [J]. SKELETAL RADIOLOGY, 2003, 32 (12) : 679 - 686
  • [9] Polyethylene wear in total hip prostheses: the influence of direction of linear wear on volumetric wear determined from radiographic data
    R. Košak
    V. Antolič
    V. Pavlovčič
    V. Kralj-Iglič
    I. Milošev
    G. Vidmar
    A. Iglič
    [J]. Skeletal Radiology, 2003, 32 : 679 - 686
  • [10] Effects of acetabular abduction on cup wear rates in total hip arthroplasty
    Del Schutte, H
    Lipman, AJ
    Bannar, SM
    Livermore, JT
    Ilstrup, D
    Morrey, BF
    [J]. JOURNAL OF ARTHROPLASTY, 1998, 13 (06): : 621 - 626