Simultaneous point-of-care testing of blood lipid profile and glucose: Performance evaluation of the GCare Lipid Analyzer

被引:1
|
作者
Kim, Ha Nui [1 ]
Yoon, Soo-Young [1 ]
机构
[1] Korea Univ, Dept Lab Med, Guro Hosp, 148 Gurodong Ro, Seoul 08308, South Korea
关键词
diabetes mellitus; dyslipidemia; glucose; lipid; POC; point-of-care testing; EDUCATION-PROGRAM RECOMMENDATIONS; DENSITY-LIPOPROTEIN CHOLESTEROL;
D O I
10.1002/jcla.24055
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Background Point-of-care (POC) testing provides quick results and includes tests for blood glucose and lipid profiles. We evaluated the newly developed POC device, the GCare Lipid Analyzer, which is used to measure glucose, total cholesterol (TC), triglyceride (TG), and high-density lipoprotein cholesterol (HDL-C) levels. Methods Venous and capillary blood samples were collected from patients who visited Korea University Guro Hospital. The results obtained using the GCare Lipid Analyzer were compared with those obtained using the TBA 2000FR chemistry analyzer and YSI 2300 STAT Plus analyzer. The glucose system evaluation process was based on the International Organization for Standardization 15197:2013 guidelines. Results The correlation coefficients (R) for TC, TG, and HDL-C were 0.965, 0.969, and 0.943 in capillary blood and 0.969, 0.990, and 0.956 in venous blood, respectively. The total errors for TC, TG, and HDL-C of the lipid profile using venous blood were all acceptable at 6.6%, 9.3%, and 11.6%, respectively. For glucose concentrations <100 mg/dl, 96.1% of the measured glucose levels were within +/- 15 mg/dl in venous samples and 100% were within +/- 15 mg/dl in capillary samples. For glucose concentrations >= 100 mg/dl, 100% and 99.5% of the measured glucose levels were within 15% for venous and capillary blood, respectively. Conclusion The performance of the GCare Lipid Analyzer is acceptable for both blood glucose and lipid profile testing, indicating that it is reliable for use in patients with diabetic dyslipidemia.
引用
收藏
页数:11
相关论文
共 50 条
  • [21] Maltose Interference-Free Test Strips for Blood Glucose Testing at Point-of-Care: A Laboratory Performance Evaluation
    Ng, Wai-Yoong
    Tiong, Chiau Chiae
    Jacob, Edward
    DIABETES TECHNOLOGY & THERAPEUTICS, 2010, 12 (11) : 889 - 893
  • [22] Evaluation and quality assessment of glucose concentration measurement in blood by point-of-care testing devices
    Wehmeier, Michael
    Arndt, Burkhard T.
    Schumann, Gerhard
    Kulpmann, Wolf R.
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2006, 44 (07) : 888 - 893
  • [23] Multicenter evaluation of a new blood glucose monitoring system for point-of-care testing.
    Weber, S
    CLINICAL CHEMISTRY, 2000, 46 (06) : A7 - A7
  • [24] Evaluation of Student Confidence Level Pertaining to Waived Point-of-Care Blood Glucose Testing
    Paolella, Patrice
    Tomesko, Jennifer
    JOURNAL OF NUTRITION EDUCATION AND BEHAVIOR, 2024, 56 (04) : 265 - 270
  • [25] Effects of pH on glucose measurements with handheld glucose meters and a portable glucose analyzer for point-of-care testing
    Tang, ZP
    Du, XG
    Louie, RF
    Kost, GJ
    ARCHIVES OF PATHOLOGY & LABORATORY MEDICINE, 2000, 124 (04) : 577 - 582
  • [26] Evaluation of the Element point-of-care blood gas analyzer for use in horses
    Krueger, Clarisa R.
    Hackett, Eileen S.
    Hess, Ann M.
    Mama, Khursheed R.
    JOURNAL OF VETERINARY EMERGENCY AND CRITICAL CARE, 2020, 30 (03) : 279 - 285
  • [27] The Evaluation of a Point-of-Care Hematology Analyzer Hemoscreen for Abnormal Blood Samples
    Hradska, Katarina
    Pulcer, Martin
    Skorupova, Michaela
    Popkova, Tereza
    Kelbichova, Veronika
    Augustynek, Martin
    Hajek, Roman
    BLOOD, 2024, 144 : 7632 - 7633
  • [28] Critical Care Glucose Point-of-Care Testing
    Narla, S. N.
    Jones, M.
    Hermayer, K. L.
    Zhu, Y.
    ADVANCES IN CLINICAL CHEMISTRY, VOL 76, 2016, 76 : 97 - 121
  • [29] Method validation of GEM® 4000 blood gas analyzer for point-of-care testing
    Martinez Sanchez, L. M.
    Lafuente Muro, S.
    Castillo Ribelles, L.
    Martinez Casademont, M.
    CLINICA CHIMICA ACTA, 2019, 493 : S765 - S765
  • [30] Point-of-Care Glucose and Lipid Profile Measures Using a Human Point-of-Care Device in Mouse Models of Type 2 Diabetes Mellitus, Aging, and Alzheimer Disease
    Smyth, Brendan J.
    Polaski, Rachel S.
    Safer, Anton
    Boettcher, Flint A.
    Konrad-Martin, Dawn
    Gratton, Michael Anne
    JOURNAL OF THE AMERICAN ASSOCIATION FOR LABORATORY ANIMAL SCIENCE, 2021, 60 (06): : 609 - 615