Knee Muscles Composition Using Electrical Impedance Myography and Magnetic Resonance Imaging

被引:1
|
作者
Albano, Domenico [1 ]
Gitto, Salvatore [2 ]
Vitale, Jacopo [1 ]
Bernareggi, Susan [1 ]
Lamorte, Sveva [3 ]
Aliprandi, Alberto [4 ]
Sconfienza, Luca Maria [1 ,2 ]
Messina, Carmelo [1 ,2 ]
机构
[1] IRCCS Ist Ortoped Galeazzi, I-20161 Milan, Italy
[2] Univ Milan, Dipartimento Sci Biomed Salute, I-20122 Milan, Italy
[3] Univ Parma, Scuola Specializzaz Radiodiagnost, I-43121 Parma, Italy
[4] Ist Clin Zucchi, Unita Operat Radiol, I-20052 Monza, Italy
关键词
Skulpt Chisel (TM); electrical impedance myography; magnetic resonance imaging; body composition; sarcopenia; muscle; knee; SKELETAL-MUSCLE; SARC-F; SARCOPENIA;
D O I
10.3390/diagnostics12092217
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
We evaluated the correlation of electrical impedance myography (EIM) measurements of knee muscles composition using Skulpt Chisel (TM) with MRI data retrieved from muscles segmentation. A total of 140 patients (71 females, 52 +/- 21 years) underwent knee MRI, EIM with Skulpt (R) , and clinical evaluation (SARC-F questionnaire). MRIs were reviewed to assess the cross-sectional area (CSA) and skeletal muscle index (SMI = CSA/height(2)) of vastus medialis, vastus lateralis, biceps, semimembranosus, and sartorius. We tested the correlations of EIM-derived parameters [body fat-percentage (BF%) and muscle quality] with total CSA, CSA of each muscle, SMI, and SARC-F scores (0-10) using Pearson correlation coefficient. We found medium negative correlation of BF% with SMI (r = -0.430, p < 0.001) and total CSA (r = -0.445, p < 0.001), particularly with biceps (r = -0.479, p < 0.001), sartorius (r = -0.440, p < 0.001), and semimembranosus (r = -0.357, p < 0.001). EIM-derived muscle quality showed small-to-medium positive correlation with MRI measurements, ranging from r = 0.234 of biceps (p = 0.006) to r = 0.302 of total CSA (p < 0.001), except for vastus lateralis (r = 0.014, p = 0.873). SARC-F scores showed small correlations with EIM and MRI data, ranging from r = -0.132 (p = 0.121) with EIM muscle quality to r = -0.288 (p = 0.001) with CSA of vastus medialis. Hence, we observed small-to-medium correlations of muscle parameters derived from Skulpt Chisel (TM) with SARC-F scores and MRI parameters. We recommend using Skulpt Chisel (TM) with caution for assessing knee skeletal muscles composition.
引用
收藏
页数:9
相关论文
共 50 条
  • [21] Magnetic resonance-electrical impedance tomography in breast cancer imaging
    Birgul, Ozlem
    Hamamura, Mark J.
    Muftuler, L. Turgan
    Nalcioglu, Orhan
    2006 3RD IEEE INTERNATIONAL SYMPOSIUM ON BIOMEDICAL IMAGING: MACRO TO NANO, VOLS 1-3, 2006, : 686 - +
  • [22] Alterations in Localized Electrical Impedance Myography of Biceps Brachii Muscles Paralyzed by Spinal Cord Injury
    Li, Le
    Stampas, Argyrios
    Shin, Henry
    Li, Xiaoyan
    Zhou, Ping
    FRONTIERS IN NEUROLOGY, 2017, 8
  • [23] Magnetic resonance imaging of the knee
    Chien, Alexander
    Weaver, Jennifer S.
    Kinne, Erica
    Omar, Imran
    POLISH JOURNAL OF RADIOLOGY, 2020, 85 : E509 - E531
  • [24] Magnetic Resonance Imaging of the Knee
    Hash, Thomas W., II
    SPORTS HEALTH-A MULTIDISCIPLINARY APPROACH, 2013, 5 (01): : 78 - 107
  • [25] Magnetic resonance imaging of the knee
    Prickett, WD
    Ward, SI
    Matava, MJ
    SPORTS MEDICINE, 2001, 31 (14) : 997 - 1019
  • [26] Magnetic Resonance Imaging of the Knee
    William D. Prickett
    Sabrina I. Ward
    Matthew J. Matava
    Sports Medicine, 2001, 31 : 997 - 1019
  • [27] Magnetic resonance imaging of facial muscles
    Farrugia, M. E.
    Bydder, G. M.
    Francis, J. M.
    Robson, M. D.
    CLINICAL RADIOLOGY, 2007, 62 (11) : 1078 - 1086
  • [28] Focused Current Density Imaging Using Internal Electrode in Magnetic Resonance Electrical Impedance Tomography (MREIT)
    Jeong, Woo Chul
    Sajib, Saurav Z. K.
    Kim, Hyung Joong
    Kwon, Oh In
    IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2014, 61 (07) : 1938 - 1946
  • [29] Static conductivity imaging using variational gradient Bz algorithm in magnetic resonance electrical impedance tomography
    Park, C
    Park, EJ
    Woo, EJ
    Kwon, O
    Seo, JK
    PHYSIOLOGICAL MEASUREMENT, 2004, 25 (01) : 257 - 269
  • [30] An investigation of the accuracy and reliability of body composition assessed with a handheld electrical impedance myography device
    Mclester, Cherilyn N.
    Dewitt, Alex D.
    Rooks, Rasmus
    Mclester, John R.
    EUROPEAN JOURNAL OF SPORT SCIENCE, 2018, 18 (06) : 763 - 771