Evaluation of B/A nonlinear parameter using an acoustic self-calibrated pulse-echo method

被引:34
|
作者
Meulen, F. Vander [1 ]
Haumesser, L. [1 ]
机构
[1] Univ Tours, INSERM, U930,IC CUP,ENI Loire, CNRS,FRE2448, F-41034 Blois, France
关键词
D O I
10.1063/1.2937838
中图分类号
O59 [应用物理学];
学科分类号
摘要
The objective of this work is to develop an easy-to-build and robust setup for measuring the nonlinearity parameter B/A in fluids using ultrasound. The method is based on the pulse-echo technique, using a single element broadband acoustic transducer, and requires electrical signal measurements. Results obtained in water and denatured alcohol validate the proposed procedure. The choice of a suitable primary wave frequency is discussed with regard to the transducer sensitivity. Further, the influence of the perturbations introduced by the experimental device nonlinearities, and the role of the reflector on the measured second harmonic field amplitude are investigated. (C) 2008 American Institute of Physics.
引用
收藏
页数:3
相关论文
共 50 条
  • [41] Enhancing acoustic signal quality by rapidly switching between pulse-echo and through-transmission using diplexers
    Valencia, Juan D.
    Diaz, Aaron A.
    Tucker, Brian J.
    NONDESTRUCTIVE CHARACTERIZATION FOR COMPOSITE MATERIALS, AEROSPACE ENGINEERING, CIVIL INFRASTRUCTURE, AND HOMELAND SECURITY 2008, 2008, 6934
  • [42] Assessment of heat-treated INCONEL alloy 690 using third-order ultrasonic nonlinearity parameter measured by the pulse-echo method
    Bang, Sung-Jun
    Song, Dong-Gi
    Jhang, Kyung-Young
    NDT & E INTERNATIONAL, 2025, 151
  • [43] Improving the performance of speckle correlation imaging by using a speckle refinement method with self-calibrated homomorphic filtering
    Liu, Yang
    Cui, Guangmang
    Shi, Shigong
    Liao, Fu
    Cui, Weize
    Zhao, Jufeng
    OPTICS AND LASER TECHNOLOGY, 2024, 179
  • [44] Thickness measurements of curved multi-layered polymer system using ultrasonic pulse-echo method
    Maev, RG
    Shao, H
    Maeva, EY
    REVIEW OF PROGRESS IN QUANTITATIVE NONDESTRUCTIVE EVALUATION, VOLS 17A AND 17B, 1998, : 1665 - 1672
  • [45] Modeling and experimental detection of damage in various materials using the pulse-echo method and piezoelectric sensors/actuators
    Yang, MJ
    Qiao, PZ
    SMART MATERIALS AND STRUCTURES, 2005, 14 (06) : 1083 - 1100
  • [46] Integrity evaluation of pipe welding zones using wavelet transforms, and specific sensitivities based on SH-EMAT pulse-echo method
    Jin-Hyuk Lee
    Dae-Hyun Kim
    International Journal of Precision Engineering and Manufacturing, 2014, 15 : 2051 - 2057
  • [47] Integrity Evaluation of Pipe Welding Zones using Wavelet Transforms, and Specific Sensitivities based on SH-EMAT Pulse-Echo Method
    Lee, Jin-Hyuk
    Kim, Dae-Hyun
    INTERNATIONAL JOURNAL OF PRECISION ENGINEERING AND MANUFACTURING, 2014, 15 (10) : 2051 - 2057
  • [48] Fluid compressional properties sensing at microscale using a longitudinal bulk acoustic wave transducer operated in a pulse-echo scheme
    Yanez, Jesus
    Uranga, Arantxa
    Barniol, Nuria
    SENSORS AND ACTUATORS A-PHYSICAL, 2022, 334
  • [49] Towards a simple acoustic method to evaluate the nonlinear parameter B/A of fluids
    Vander Meulen, Francois
    Haumesser, Lionel
    2008 IEEE ULTRASONICS SYMPOSIUM, VOLS 1-4 AND APPENDIX, 2008, : 1908 - 1911
  • [50] A radial self-calibrated (RASCAL) generalized autocalibrating partially parallel acquisition (GRAPPA) method using weight interpolation
    Codella, Noel C. F.
    Spincemaille, Pascal
    Prince, Martin
    Wang, Yi
    NMR IN BIOMEDICINE, 2011, 24 (07) : 844 - 854