摘要
We describe a setup for the calibration of the transresistance gain of low-current amplifiers, based on the capacitance-charging method. The calibration can be performed in the current range of typical interest for electron-counting experiments. The setup implementation is simple and rugged, and is suitable to be embedded in larger experiments, where the amplifier is employed. The calibrated transresistance is traceable to the units of capacitance and time. Two different calibration modes were tested: with dc current (obtained using a custom-made piecewise linear ramp generator) and with low-frequency sinewave current (using a commercial generator). The relative base accuracy of the implementation is in the 10(-5) range.
引用
收藏
页码:2676 / 2683
页数:8
相关论文
共 50 条
  • [41] A SETUP FOR THE CALIBRATION OF THE HPC READOUT CHAMBERS
    CAMPORESI, T
    GEARA, C
    NAVARRIA, FL
    SEUFERT, R
    VALENTI, G
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1989, 277 (01): : 77 - 82
  • [42] High-order linear transformation MOSFET-C filters using operational transresistance amplifiers
    Hwang, YS
    Chen, JJ
    Lee, WT
    2005 IEEE INTERNATIONAL SYMPOSIUM ON CIRCUITS AND SYSTEMS (ISCAS), VOLS 1-6, CONFERENCE PROCEEDINGS, 2005, : 3275 - 3278
  • [43] Ultralow threshold current lasers
    Dapkus, P.Daniel
    MacDougal, Michael H.
    Yang, Gye Mo
    Bond, Aaron E.
    Lin, Chao-Kun
    Tishinin, Denis
    Pudikov, Vasily
    Cheng, Yong
    Uppal, Kushant
    Conference on Quantum Electronics and Laser Science (QELS) - Technical Digest Series, 1996, 9 : 168 - 169
  • [44] Calibration procedure in microchannel amplifiers design
    Ivanov, V
    Barnyakov, A.
    Barnyakov, M.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2018, 903 : 170 - 174
  • [45] A Calibration Method of the Linearity of Lock In Amplifiers
    Raso, F.
    Hortelano, A.
    Izquierdo, M. M.
    2016 CONFERENCE ON PRECISION ELECTROMAGNETIC MEASUREMENTS (CPEM 2016), 2016,
  • [46] Current Differencing Transresistance Amplifier (CDTRA) and Its Application for Analog Signal Processing
    Siripruchyanun, Montree
    Poonnoy, Nitchamon
    Upathamkuekool, Chairat
    Payakkakul, Kangwal
    2016 INTERNATIONAL ELECTRICAL ENGINEERING CONGRESS, IEECON2016, 2016, : 184 - 187
  • [47] A high precision calibration setup for low-current meters in the range of 10 pA to 1 fA
    Kim, Wan-Seop
    Kim, Mun-Seog
    Kim, Kyu-Tae
    Chong, Yonuk
    Kim, Han-Jun
    Kim, Myungsoo
    Tauscher, H. Norbert
    2008 CONFERENCE ON PRECISION ELECTROMAGNETIC MEASUREMENTS DIGEST, 2008, : 358 - +
  • [48] VERY-HIGH-OUTPUT-IMPEDANCE CASCODE CURRENT SOURCES CURRENT MIRRORS TRANSRESISTANCE STAGES AND THEIR APPLICATIONS
    ZARABADIL, SR
    ISMAIL, M
    INTERNATIONAL JOURNAL OF CIRCUIT THEORY AND APPLICATIONS, 1992, 20 (06) : 639 - 648
  • [49] A high sensitivity ultralow temperature RF conductance and noise measurement setup
    Parmentier, F. D.
    Mahe, A.
    Denis, A.
    Berroir, J. -M.
    Glattli, D. C.
    Placais, B.
    Feve, G.
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2011, 82 (01):
  • [50] Wire setup calibration of beam position monitors
    Wang, D
    Binns, B
    Kogan, M
    Zolfaghari, A
    PROCEEDINGS OF THE 1995 PARTICLE ACCELERATOR CONFERENCE, VOLS 1-5, 1996, : 2536 - 2538