Quantitative measurement of spastic ankle joint stiffness using a manual device: A preliminary study

被引:9
|
作者
Kobayashi, Toshiki [1 ]
Leung, Aaron K. L. [1 ]
Akazawa, Yasushi [2 ]
Tanaka, Masao [3 ]
Hutchins, Stephen W. [4 ]
机构
[1] Hong Kong Polytech Univ, Dept Hlth Technol & Informat, Kowloon, Hong Kong, Peoples R China
[2] Hyogo Inst Assist Technol, Kobe, Hyogo, Japan
[3] Osaka Univ, Grad Sch Engn Sci, Osaka, Japan
[4] Univ Salford, Ctr Rehabil & Human Performance Res, Salford M5 4WT, Lancs, England
关键词
Spasticity; Ankle; Quantification; Manual; Stroke; OF-THE-LITERATURE;
D O I
10.1016/j.jbiomech.2010.02.024
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
Quantitative measurement of ankle joint stiffness following stroke could prove useful in monitoring the progress of a rehabilitation programme. The objective of this study was to design a manual device for use in the clinical setting. Manual measurement of spastic ankle joint stiffness has historically been conducted using hand-held dynamometers or alternative devices, but some difficulties have been reported in controlling the velocity applied to the ankle during the measurement. In this study, a manually operated device was constructed with a footplate, a torquemeter and a potentiometer. It was mechanically designed to rotate around an approximated axis of the ankle joint and to measure ankle joint angular position and its corresponding resistive torque. Two stroke hemiplegic subjects pariticapted in a pilot study. The results suggested that difficulty in controlling the applied velocity might be complemented by presenting torque data as a function of peak angular velocity in each stretching cycle. Moreover, the results demonstrated that the device could potentially apply a wide range of angular velocities and provide potentially useful clinical information. Quantitative data successfully acquired using this method included the approximate ankle angular position, where the velocity-dependent characteristics of stiffness was notably initiated and its corresponding torque and velocity. (C) 2010 Elsevier Ltd. All rights reserved.
引用
收藏
页码:1831 / 1834
页数:4
相关论文
共 50 条
  • [31] Ankle Joint Stretching Device Using Tension Rod for Self Rehabilitation
    Toda, Hideki
    Sugihara, Shin
    JOURNAL OF ROBOTICS AND MECHATRONICS, 2021, 33 (04) : 843 - 850
  • [32] Measurement of ankle spasticity in children with cerebral palsy using a manual spasticity evaluator
    Peng, QY
    Shah, P
    Selles, RW
    Gaebler-Spira, DJ
    Zhang, LQ
    PROCEEDINGS OF THE 26TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-7, 2004, 26 : 4896 - 4899
  • [33] QUANTITATIVE ASSESSMENT OF THE EFFECT OF CYPROHEPTADINE ON SPASTIC PARETIC GAIT - A PRELIMINARY-STUDY
    WAINBERG, M
    BARBEAU, H
    GAUTHIER, S
    JOURNAL OF NEUROLOGY, 1986, 233 (05) : 311 - 314
  • [34] Methods of Joint Stiffness Measurement Using a Manually Actuated Dynamometer
    Kutilek, Patrik
    Volf, Petr
    Hybl, Jan
    Hejda, Jan
    Viteckova, Slavka
    Krivanek, Vaclav
    Doskocil, Radek
    Smrcka, Pavel
    MECHATRONICS 2019: RECENT ADVANCES TOWARDS INDUSTRY 4.0, 2020, 1044 : 125 - 132
  • [35] Joint and Myofascial Manual Therapy Techniques in Haemophilic Ankle Arthropathy: A Randomized Pilot Study
    Truque-Diaz, Carlos
    Merono-Gallut, Javier
    Cuesta-Barriuso, Ruben
    Perez-Llanes, Raul
    HAEMOPHILIA, 2025,
  • [36] A preliminary quantitative EEG study on Augmented Reality Guidance of Manual Tasks
    Rho, Gianluca
    Callara, Alejandro Luis
    Condino, Sara
    Ghiasi, Shadi
    Nardelli, Mimma
    Carbone, Marina
    Ferrari, Vincenzo
    Greco, Alberto
    Riling, Enzo Pasquale
    2020 IEEE INTERNATIONAL SYMPOSIUM ON MEDICAL MEASUREMENTS AND APPLICATIONS (MEMEA), 2020,
  • [37] Knee Stiffness in Assistive Device Control at Quiet Stance: A Preliminary Study
    Sreenivasan, Gayatri
    Zhu, Chunchu
    Yi, Jingang
    IFAC PAPERSONLINE, 2023, 56 (03): : 163 - 168
  • [38] Knee joint stiffness in individuals with and without knee osteoarthritis: A preliminary study
    Oatis, Carol A.
    Wolff, Edward F.
    Lennon, Sandra K.
    JOURNAL OF ORTHOPAEDIC & SPORTS PHYSICAL THERAPY, 2006, 36 (12): : 935 - 941
  • [39] RELIABILITY OF TAPE MEASUREMENT OF ANKLE JOINT SWELLING USING A FIGURE OF 8
    PERRY, SB
    FELLMAN, SK
    PHYSICAL THERAPY, 1986, 66 (05): : 802 - 802
  • [40] Experimental and computational analysis of soft tissue stiffness in forearm using a manual indentation device
    Iivarinen, Jarkko T.
    Korhonen, Rami K.
    Julkunen, Petro
    Jurvelin, Jukka S.
    MEDICAL ENGINEERING & PHYSICS, 2011, 33 (10) : 1245 - 1253