Quantitative measurement of spastic ankle joint stiffness using a manual device: A preliminary study

被引:9
|
作者
Kobayashi, Toshiki [1 ]
Leung, Aaron K. L. [1 ]
Akazawa, Yasushi [2 ]
Tanaka, Masao [3 ]
Hutchins, Stephen W. [4 ]
机构
[1] Hong Kong Polytech Univ, Dept Hlth Technol & Informat, Kowloon, Hong Kong, Peoples R China
[2] Hyogo Inst Assist Technol, Kobe, Hyogo, Japan
[3] Osaka Univ, Grad Sch Engn Sci, Osaka, Japan
[4] Univ Salford, Ctr Rehabil & Human Performance Res, Salford M5 4WT, Lancs, England
关键词
Spasticity; Ankle; Quantification; Manual; Stroke; OF-THE-LITERATURE;
D O I
10.1016/j.jbiomech.2010.02.024
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
Quantitative measurement of ankle joint stiffness following stroke could prove useful in monitoring the progress of a rehabilitation programme. The objective of this study was to design a manual device for use in the clinical setting. Manual measurement of spastic ankle joint stiffness has historically been conducted using hand-held dynamometers or alternative devices, but some difficulties have been reported in controlling the velocity applied to the ankle during the measurement. In this study, a manually operated device was constructed with a footplate, a torquemeter and a potentiometer. It was mechanically designed to rotate around an approximated axis of the ankle joint and to measure ankle joint angular position and its corresponding resistive torque. Two stroke hemiplegic subjects pariticapted in a pilot study. The results suggested that difficulty in controlling the applied velocity might be complemented by presenting torque data as a function of peak angular velocity in each stretching cycle. Moreover, the results demonstrated that the device could potentially apply a wide range of angular velocities and provide potentially useful clinical information. Quantitative data successfully acquired using this method included the approximate ankle angular position, where the velocity-dependent characteristics of stiffness was notably initiated and its corresponding torque and velocity. (C) 2010 Elsevier Ltd. All rights reserved.
引用
收藏
页码:1831 / 1834
页数:4
相关论文
共 50 条
  • [1] Assessment of a portable device for the quantitative measurement of ankle joint stiffness in spastic individuals
    Lorentzen, Jakob
    Grey, Michael J.
    Geertsen, Svend S.
    Biering-Sorensen, Fin
    Brunton, Kelly
    Gorassini, Monica
    Nielsen, Jens Bo
    CLINICAL NEUROPHYSIOLOGY, 2012, 123 (07) : 1371 - 1382
  • [2] Design of a manual device to measure ankle joint stiffness and range of motion
    Kobayashi, Toshiki
    Leung, Aaron K. L.
    Hutchins, Stephen W.
    PROSTHETICS AND ORTHOTICS INTERNATIONAL, 2011, 35 (04) : 478 - 481
  • [3] Measurement of the Passive Stiffness of Ankle Joint in 3 DOF Using Stewart Platform Type Ankle Foot Device
    Nomura, Kenta
    Yonezawa, Teru
    Mizoguchi, Hiroshi
    Takemura, Hiroshi
    2016 38TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY (EMBC), 2016, : 5011 - 5014
  • [4] QUANTITATIVE MEASUREMENT OF JOINT STIFFNESS
    WRIGHT, V
    JOHNS, RJ
    JOURNAL OF CLINICAL INVESTIGATION, 1959, 38 (06): : 1056 - 1056
  • [5] Quantitative evaluations of ankle spasticity and stiffness in neurological disorders using manual spasticity evaluator
    Peng, Qiyu
    Park, Hyung-Soon
    Shah, Parag
    Wilson, Nicole
    Ren, Yupeng
    Wu, Yi-Ning
    Liu, Jie
    Gaebler-Spira, Deborah J.
    Zhang, Li-Qun
    JOURNAL OF REHABILITATION RESEARCH AND DEVELOPMENT, 2011, 48 (04): : 473 - 481
  • [6] Validity and reliability of a new in vivo ankle stiffness measurement device
    Zinder, Steven M.
    Granata, Kevin P.
    Padua, Darin A.
    Gansneder, Bruce M.
    JOURNAL OF BIOMECHANICS, 2007, 40 (02) : 463 - 467
  • [7] A Novel Method for Field Measurement of Ankle Joint Stiffness in Hopping
    Ghorbani Faal, Sanubar
    Shirzad, Elham
    Sharifnezhad, Ali
    Ashrostaghi, Mojtaba
    Naemi, Roozbeh
    APPLIED SCIENCES-BASEL, 2021, 11 (24):
  • [8] Measurement device for ankle joint kinematic and dynamic characterisation
    C. Giacomozzi
    S. Cesinaro
    F. Basile
    G. De Angelis
    D. Giansanti
    G. Maccioni
    E. Masci
    A. Panella
    M. Paolizzi
    M. Torre
    P. Valentini
    V. Macellari
    Medical and Biological Engineering and Computing, 2003, 41 : 486 - 493
  • [9] A measurement device for anterior laxity of the ankle joint complex
    Kerkhoffs, GMMJ
    Blankevoort, L
    van Dijk, CN
    CLINICAL BIOMECHANICS, 2005, 20 (02) : 218 - 222
  • [10] Measurement device for ankle joint kinematic and dynamic characterisation
    Giacomozzi, C
    Cesinaro, S
    Basile, F
    De Angelis, G
    Giansanti, D
    Maccioni, G
    Masci, E
    Panella, A
    Paolizzi, M
    Torre, M
    Valentini, P
    Macellari, V
    MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 2003, 41 (04) : 486 - 493