Interactions of medical implants with the magnetic fields in MRI

被引:4
|
作者
Nyenhuis, JA [1 ]
机构
[1] Purdue Univ, Sch Elect & Comp Engn, W Lafayette, IN 47907 USA
关键词
MRI; magnetic resonance Imaging; implant; safety;
D O I
10.1109/IEMBS.2003.1280980
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Interactions of the static, gradient and radio frequency (RF) magnetic fields in a magnetic resonance (MR) imager with implanted medical devices are qualitatively and quantitatively reviewed. The static field has typical intensity of 1.5 T, but may be as large as 8 T. Spatial variations in the static field induce potentially large forces on magnetic components. Torque is exerted on magnetic objects, tending to align the long axis along the field. Moving conducting objects will experience force and torque due to induced eddy currents. The gradient fields are of audio frequency and peak intensity of the time-varying magnetic field dB/dt may exceed 100 T/s. Metallic implants tend to locally focus the gradient currents, enhancing the potential for peripheral nerve stimulation. The frequency of the RF magnetic field is proportional to the intensity of the static field and is 64 MHz for 1.5 T. The principal bioeffect of the RF field is tissue heating. Elongated metallic implants will locally focus the RF currents, resulting in the potential for dangerous temperature rises. From calculation and measurement of the interactions of an implant with the fields in an MR imager, the safe operating conditions for a patient with an implant may be determined.
引用
收藏
页码:3767 / 3770
页数:4
相关论文
共 50 条
  • [1] MRI interactions with medical implants
    Nyenhuis, J
    [J]. 2002 IEEE INTERNATIONAL SYMPOSIUM ON ELECTROMAGNETIC COMPATIBILITY, VOLS 1 AND 2, SYMPOSIUM RECORD, 2002, : 920 - 924
  • [2] Interactions of magnetic resonance imaging radio frequency magnetic fields with elongated medical implants
    Smith, CD
    Kildishev, AV
    Nyenhuis, JA
    Foster, KS
    Bourland, JD
    [J]. JOURNAL OF APPLIED PHYSICS, 2000, 87 (09) : 6188 - 6190
  • [3] Interaction of medical implants with the MRI electromagnetic fields
    Mohsin, S.A.
    Nyenhuis, J.
    Masood, R.
    [J]. Progress In Electromagnetics Research C, 2010, 13 : 195 - 202
  • [4] MRI Image Distortion due to Magnetic Materials in Medical Implants
    Harberts, Dick W.
    van Helvoort, Mark
    [J]. 2015 IEEE INTERNATIONAL SYMPOSIUM ON ELECTROMAGNETIC COMPATIBILITY (EMC), 2015, : 1463 - 1466
  • [5] Techniques to demonstrate MRI safety of medical implants
    Oberle, Michael
    [J]. Microwave Journal, 2009, 52 (12): : 60 - 68
  • [6] TECHNIQUES TO DEMONSTRATE MRI SAFETY OF MEDICAL IMPLANTS
    Oberle, Michael
    [J]. MICROWAVE JOURNAL, 2009, 52 (12) : 60 - +
  • [7] MRI safety and compatibility of implants and medical devices
    Woods, TO
    [J]. STAINLESS STEELS FOR MEDICAL AND SURGICAL APPLICATIONS, 2003, 1438 : 82 - 90
  • [8] On interactions of static magnetic fields
    Zor, Omer
    [J]. JOURNAL OF ELECTRICAL ENGINEERING-ELEKTROTECHNICKY CASOPIS, 2019, 70 (03): : 253 - 255
  • [9] Understanding magnetic fields, for NMR/MRI
    Conradi, Mark S.
    [J]. JOURNAL OF MAGNETIC RESONANCE OPEN, 2024, 20
  • [10] Neurofunctional MRI at high magnetic fields
    Speck, O.
    Turner, R.
    [J]. RADIOLOGE, 2013, 53 (05): : 415 - 421