Characterization of a multi-axis ion chamber array

被引:23
|
作者
Simon, Thomas A. [1 ,2 ,3 ]
Kozelka, Jakub [3 ]
Simon, William E. [3 ]
Kahler, Darren [2 ]
Li, Jonathan [2 ]
Liu, Chihray [2 ]
机构
[1] Univ Florida, Dept Nucl & Radiol Engn, Gainesville, FL 32611 USA
[2] Univ Florida, Dept Radiat Oncol, Hlth Sci Ctr, Gainesville, FL 32610 USA
[3] Sun Nucl Corp, Melbourne, FL 32940 USA
关键词
linear accelerator; ionization chamber; quality assurance; detector array; characterization; DIODE-ARRAY; DOSIMETRY;
D O I
10.1118/1.3505452
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: The aim of this work was to characterize a multi-axis ion chamber array (IC PROFILER (TM); Sun Nuclear Corporation, Melbourne, FL USA) that has the potential to simplify the acquisition of LINAC beam data. Methods: The IC PROFILER (TM) (or panel) measurement response was characterized with respect to radiation beam properties, including dose, dose per pulse, pulse rate frequency (PRF), and energy. Panel properties were also studied, including detector-calibration stability, power-on time, backscatter dependence, and the panel's agreement with water tank measurements [profiles, fractional depth dose (FDD), and output factors]. Results: The panel's relative deviation was typically within (+/-) 1% of an independent (or nominal) response for all properties that were tested. Notable results were (a) a detectable relative field shape change of similar to 1% with linear accelerator PRF changes; (b) a large range in backscatter thickness had a minimal effect on the measured dose distribution (typically less than 1%); (c) the error spread in profile comparison between the panel and scanning water tank (Blue Phantom (TM), CC13 (TM); IBA Schwarzenbruck, DE) was approximately (+/-) 0.75%. Conclusions: The ability of the panel to accurately reproduce water tank profiles, FDDs, and output factors is an indication of its abilities as a dosimetry system. The benefits of using the panel versus a scanning water tank are less setup time and less error susceptibility. The same measurements (including device setup and breakdown) for both systems took 180 min with the water tank versus 30 min with the panel. The time-savings increase as the measurement load is increased. (C) 2010 American Association of Physicists in Medicine. [DOI: 10.1118/1.3505452]
引用
收藏
页码:6101 / 6111
页数:11
相关论文
共 50 条
  • [21] Segmentation theory for design of a multi-axis actuator array using segmented binary control
    Cho, KJ
    Asada, HH
    ACC: PROCEEDINGS OF THE 2005 AMERICAN CONTROL CONFERENCE, VOLS 1-7, 2005, : 1969 - 1974
  • [22] Multi-Axis Electric Stepper Motor
    Groenhuis, Vincent
    Rolff, Gijs
    Bosman, Koen
    Abelmann, Leon
    Stramigioli, Stefano
    IEEE ROBOTICS AND AUTOMATION LETTERS, 2021, 6 (04): : 7201 - 7208
  • [23] Alignment of optical axis parallelism in multi-axis system
    Wang, Shurong
    Li, Zhanfeng
    Huang, Yu
    Lin, Guanyu
    Yang, Xiaohu
    Chinese Optics Letters, 2015, 13
  • [24] Multi-axis machines make sense
    Machinery and Production Engineering, 2000, 158 (4006):
  • [25] Dyson invest in multi-axis robots
    不详
    INDUSTRIAL ROBOT, 2002, 29 (05): : 455 - 457
  • [26] Creating a Multi-axis Machining Postprocessor
    Vavruska, Petr
    ACTA POLYTECHNICA, 2012, 52 (04) : 113 - 117
  • [27] MULTI-AXIS VIBRATION ENVIRONMENT AND MAN
    LOVESEY, EJ
    APPLIED ERGONOMICS, 1970, 1 (05) : 258 - 261
  • [28] Multi-axis inertial sensing system
    不详
    AIRCRAFT ENGINEERING AND AEROSPACE TECHNOLOGY, 2002, 74 (01): : 88 - 89
  • [29] A Soft Multi-Axis Force Sensor
    Vogt, Daniel
    Park, Yong-Lae
    Wood, Robert J.
    2012 IEEE SENSORS PROCEEDINGS, 2012, : 897 - 900
  • [30] MAGNETIC TRANSITIONS IN A MULTI-AXIS ANTIFERROMAGNET
    KEEN, BE
    LANDAU, DP
    WOLF, WP
    PHYSICS LETTERS, 1966, 23 (03): : 202 - &