Controllable Wireless Power Delivery for Biomedical Implants

被引:0
|
作者
Venkateswaran, Madhav [1 ]
Lee, Fred T., Jr. [2 ]
van der Weide, Daniel [1 ]
机构
[1] Univ Wisconsin, Dept Elect & Comp Engn, Madison, WI 53706 USA
[2] Univ Wisconsin, Dept Radiol, Madison, WI 53706 USA
关键词
D O I
暂无
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
We demonstrate a method, independent of coil geometry, orientation and separation, to measure inductive wireless power transfer to an implanted coil, without a bidirectional communication link. Load switching on the implantable coil at 1 kHz modulates the 13.56 MHz carrier signal. A lock-in amplifier detects current perturbation in the excitation coil at the switching frequency producing an output proportional to the induced current in the implantable coil. Measurements show good agreement with analytical values at different locations in a 5mm x 5mm grid for parallel and orthogonal coil orientations. Observed standard deviations are within 10% and 15% for parallel and orthogonal orientations.
引用
收藏
页码:125 / 128
页数:4
相关论文
共 50 条
  • [21] Biosafety Considerations of a Capacitive Link for Wireless Power Transfer to Biomedical Implants
    Erfani, Reza
    Marefat, Fatemeh
    Mohseni, Pedram
    2018 IEEE BIOMEDICAL CIRCUITS AND SYSTEMS CONFERENCE (BIOCAS): ADVANCED SYSTEMS FOR ENHANCING HUMAN HEALTH, 2018, : 355 - 358
  • [22] Modeling and Design of Wireless Data Telemetry and Power Transfer for Biomedical Implants
    Agarwal, Kush
    Jegadeesan, Rangarajan
    Guo, Yong-Xin
    Thakor, Nitish V.
    2015 IEEE INTERNATIONAL CONFERENCE ON COMPUTATIONAL ELECTROMAGNETICS (ICCEM), 2015, : 64 - 66
  • [23] Midfield Wireless Power Transfer for Deep-Tissue Biomedical Implants
    Nguyen, Ngan
    Ha-Van, Nam
    Seo, Chulhun
    IEEE ANTENNAS AND WIRELESS PROPAGATION LETTERS, 2020, 19 (12): : 2270 - 2274
  • [24] Efficient Wireless Power Delivery and Reliable Telemetry for Implants and Safety
    Jegadeesan, Rangarajan
    Guo, Yong-Xin
    Yen, Shih-Cheng
    Thakor, Nitish Vyomesh
    2015 IEEE MTT-S INTERNATIONAL MICROWAVE WORKSHOP SERIES ON RF AND WIRELESS TECHNOLOGIES FOR BIOMEDICAL AND HEALTHCARE APPLICATIONS (IMWS-BIO), 2015, : 25 - 26
  • [25] A Multiband Antenna Associating Wireless Monitoring and Nonleaky Wireless Power Transfer System for Biomedical Implants
    Das, Rupam
    Yoo, Hyoungsuk
    IEEE TRANSACTIONS ON MICROWAVE THEORY AND TECHNIQUES, 2017, 65 (07) : 2485 - 2495
  • [26] Modeling of Capacitive Resonant Wireless Power and Data Transfer to Deep Biomedical Implants
    Narayanamoorthi, R.
    IEEE TRANSACTIONS ON COMPONENTS PACKAGING AND MANUFACTURING TECHNOLOGY, 2019, 9 (07): : 1253 - 1263
  • [27] Backscattering-based Wireless Communication and Power Transfer to Small Biomedical Implants
    Ukkonen, Leena
    Sydanheimo, Lauri
    Ma, Shubin
    Bjorninen, Toni
    MICROFLUIDICS, BIOMEMS, AND MEDICAL MICROSYSTEMS XVIII, 2020, 11235
  • [28] A Low-Frequency Versatile Wireless Power Transfer Technology for Biomedical Implants
    Jiang, Hao
    Zhang, Junmin
    Lan, Di
    Chao, Kelvin K.
    Liou, Shyshenq
    Shahnasser, Hamid
    Fechter, Richard
    Hirose, Shinjiro
    Harrison, Michael
    Roy, Shuvo
    IEEE TRANSACTIONS ON BIOMEDICAL CIRCUITS AND SYSTEMS, 2013, 7 (04) : 526 - 535
  • [29] Transcutaneous Capacitive Wireless Power Transfer (C-WPT) for Biomedical Implants
    Erfani, Reza
    Marefat, Fatemeh
    Sodagar, Amir M.
    Mohseni, Pedram
    2017 IEEE INTERNATIONAL SYMPOSIUM ON CIRCUITS AND SYSTEMS (ISCAS), 2017, : 2561 - 2564
  • [30] Optimization of the Coupling Coefficient of the Inductive Link for Wireless Power Transfer to Biomedical Implants
    Bao, Jiarui
    Hu, Shuyan
    Xie, Zibin
    Hu, Guangxi
    Lu, Ye
    Zheng, Lirong
    INTERNATIONAL JOURNAL OF ANTENNAS AND PROPAGATION, 2022, 2022