Further developments in an ac-based capacitance tomography system

被引:54
|
作者
Yang, WQ [1 ]
机构
[1] Univ Manchester, Inst Sci & Technol, Dept Elect Engn & Elect, Manchester M60 1QD, Lancs, England
来源
REVIEW OF SCIENTIFIC INSTRUMENTS | 2001年 / 72卷 / 10期
关键词
D O I
10.1063/1.1403012
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
An ac-based electrical capacitance tomography (ECT) system has been developed at UMIST, with sinusoidal excitation and phase-sensitive demodulation (PSD). In this article the overall system design is briefly described and then discussed in detail are the key design features, needed to achieve a high signal-to-noise ratio and a high data acquisition rate, including transient process analysis of the first stage, design of an analog Butterworth low-pass filter and optimization of the data acquisition process. The experiments show that the ac-based system has signal variance 19 times better than the previous design, the charge/discharge ECT system, and that a data acquisition rate of 140 frames per second is achieved for a 12-electrode system. (C) 2001 American Institute of Physics.
引用
收藏
页码:3902 / 3907
页数:6
相关论文
共 50 条
  • [1] New AC-based capacitance tomography system
    Yang, WQ
    York, TA
    IEE PROCEEDINGS-SCIENCE MEASUREMENT AND TECHNOLOGY, 1999, 146 (01) : 47 - 53
  • [2] Improvement of AC-based Electrical Capacitance Tomography Hardware
    Styra, D.
    Babout, L.
    ELEKTRONIKA IR ELEKTROTECHNIKA, 2010, (07) : 47 - 50
  • [3] Analysis of the effect of stray capacitance on an AC-based capacitance tomography transducer
    Yang, WQ
    Stott, AL
    Gamio, JC
    IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 2003, 52 (05) : 1674 - 1681
  • [4] AC-based capacitance tomography system with small-diameter and high-pressure pipe
    Yang, Daoye
    Xu, Chuanlong
    Zhou, Bin
    Wang, Shimin
    Wu, Xin
    MULTIPHASE FLOW: THE ULTIMATE MEASUREMENT CHALLENGE, PROCEEDINGS, 2007, 914 : 725 - +
  • [5] Analysis of non-ideal characteristics of an ac-based capacitance transducer for tomography
    Gamio, JC
    Yang, WQ
    Stott, AL
    MEASUREMENT SCIENCE AND TECHNOLOGY, 2001, 12 (08) : 1076 - 1082
  • [6] The dynamic response of a Butterworth low-pass filter in an ac-based electrical capacitance tomography system
    Chen, Dixiang
    Yang, Wuqiang
    Pan, Mengchun
    MEASUREMENT SCIENCE AND TECHNOLOGY, 2010, 21 (10)
  • [7] Influence of particle electrification on AC-based capacitance measurement and its elimination
    Li, Jian
    Kong, Ming
    Xu, Chuanlong
    Wang, Shengnan
    Wang, Shimin
    MEASUREMENT, 2015, 76 : 93 - 103
  • [8] Investigation of droplet distribution in electrohydrodynamic atomization (EHDA) using an ac-based electrical capacitance tomography (ECT) system with an internal-external electrode sensor
    Rezvanpour, Alireza
    Wang, Chi-Hwa
    Liang, Yung C.
    Yang, Wuqiang
    MEASUREMENT SCIENCE AND TECHNOLOGY, 2012, 23 (01)
  • [9] AC-based brushless dc motor
    Kochergin V.V.
    Zakrevskaya B.V.
    Russian Electrical Engineering, 2009, 80 (10) : 539 - 541
  • [10] Modeling collapse and blackout of power system via AC-based model
    Mei Shengwei
    Xue Ancheng
    Weng Xiaofeng
    He Fei
    IPEC: 2005 INTERNATIONAL POWER ENGINEERING CONFERENCE, VOLS 1 AND 2, 2005, : 747 - 751