Comparison of Three High Resolution Real-Time Spectrometers for Microwave Ozone Profiling Instruments

被引:5
|
作者
Sauvageat, Eric [1 ,2 ]
Albers, Roland [1 ]
Kotiranta, Mikko [1 ]
Hocke, Klemens [1 ,2 ]
Gomez, R. Michael [3 ]
Nedoluha, Gerald [3 ]
Murk, Axel [1 ,2 ]
机构
[1] Univ Bern, Inst Appl Phys, CH-3012 Bern, Switzerland
[2] Univ Bern, Oeschger Ctr Climate Change Res, CH-3012 Bern, Switzerland
[3] US Naval Res Lab, Remote Sensing Div, Washington, DC 20375 USA
关键词
Gases; Atmospheric measurements; Radiometers; Instruments; Microwave radiometry; Time series analysis; Real-time systems; digital real-time spectrometers; microwave radiometry; microwave spectroscopy; ozone; remote sensing; water vapor; WATER-VAPOR;
D O I
10.1109/JSTARS.2021.3114446
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
In this contribution, we present a comparison of three digital real-time spectrometers used in passive remote sensing of ozone and other trace gases in the middle atmosphere. During a period of six months, we connected the spectrometers to the same radiometric front-end to perform parallel observations of the ozone emission line at 110.836 GHz. This allowed us to better characterize a bias previously observed on the integrated spectra of the Acqiris AC240, a widely used digital spectrometer which has been used for more than a decade in many operational microwave radiometers. We investigated the bias under different atmospheric conditions and found that it is caused by multiple sources. Nonlinearities in the calibration are responsible for part of the bias, but a larger contribution stems from a second effect in the AC240. Although this error source is still partly unexplained, we found that a simple correction scheme simulating a spectral leakage can be applied to the integrated spectra of the AC240 and worked well on our range of observations. We also show that by applying our bias correction to the spectra, we can correct the bias in the ozone retrievals. There is still a need for further measurements to validate this approximate correction, but it could help to correct the numerous time series of ozone and other atmospheric constituents recorded by the AC240.
引用
收藏
页码:10045 / 10056
页数:12
相关论文
共 50 条
  • [1] Comparison of three high resolution real-time spectrometers for microwave ozone profiling instruments
    Sauvageat, Eric
    Kotiranta, Mikko
    Hocke, Klemens
    Gomez, R. Michael
    Nedoluha, Gerald
    Murk, Axel
    16TH SPECIALIST MEETING ON MICROWAVE RADIOMETRY AND REMOTE SENSING OF THE ENVIRONMENT (RAD 2020), 2020,
  • [2] Real-Time Picometer Resolution Spectrometers
    Beaton, Alexander
    SPECTROSCOPY, 2023, 38 (02) : 40 - 41
  • [3] Microwave Photonic Bistatic Radar for Real-Time and High-Resolution Imaging
    Fan, Beichen
    Zhang, Fangzheng
    Ma, Cong
    Yang, Yue
    Pan, Shilong
    Wang, Xiangchuan
    IEEE PHOTONICS TECHNOLOGY LETTERS, 2020, 32 (21) : 1397 - 1400
  • [4] THE REAL-TIME OPTIMIZATION OF ELECTRON SPECTROMETERS
    HOWELL, SK
    BOWRING, NJ
    HAMILTON, TDS
    JOURNAL OF PHYSICS E-SCIENTIFIC INSTRUMENTS, 1987, 20 (03): : 293 - 314
  • [5] High-resolution, real-time three-dimensional shape measurement
    Zhang, Song
    Huang, Peisen S.
    OPTICAL ENGINEERING, 2006, 45 (12)
  • [6] Ultra-high resolution real-time radar imaging based on microwave photonics
    Pan, Shilong
    Zhang, Fangzheng
    23RD OPTO-ELECTRONICS AND COMMUNICATIONS CONFERENCE (OECC2018), 2018,
  • [7] Ultra-high Resolution Real-time Radar Imaging Based on Microwave Photonics
    Zhang, Fangzheng
    Pan, Shilong
    2018 PROGRESS IN ELECTROMAGNETICS RESEARCH SYMPOSIUM (PIERS-TOYAMA), 2018, : 973 - 977
  • [8] REAL-TIME HIGH RESOLUTION MASS SPECTROMETRY
    BURLINGA.AL
    SMITH, DH
    OLSEN, RW
    MERREN, TO
    ORGANIC MASS SPECTROMETRY, 1968, 1 (03): : 491 - &
  • [9] REAL-TIME OPTIMISATION OF ELECTRON SPECTROMETERS.
    Howell, S.K.
    Bowring, N.J.
    Hamilton, T.D.S.
    1600, (20):