Efficiencies and noise levels of portable surgical smoke evacuation systems

被引:24
|
作者
Seipp, Hans-Martin [1 ]
Steffens, Thomas [1 ]
Weigold, Janine [1 ]
Lahmer, Armin [1 ]
Maier-Hasselmann, Andreas [2 ]
Herzog, Torsten [3 ]
Herzog-Niescery, Jennifer [4 ]
机构
[1] Univ Appl Sci, Dept Life Sci Engn, Giessen, Germany
[2] Stadt Klinikum Bogenhausen, Dept Vasc Surg, Munich, Germany
[3] Ruhr Univ Bochum, St Josef Hosp, Dept Surg, Bochum, Germany
[4] Ruhr Univ Bochum, St Josef Hosp, Dept Anesthesiol, Bochum, Germany
关键词
Electrocauterization; healthcare workers; local exhaust ventilation; occupational exposure; surgical smoke; EXPOSURE; ROOM;
D O I
10.1080/15459624.2018.1513134
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
Surgical smoke resulting from electrocauterization is a health risk for operating room personnel. The U.S. National Institute for Occupational Safety and Health recommends the use of local exhaust ventilation such as a portable smoke evacuation system to reduce surgical smoke, but its efficiency has never been assessed under experimental conditions. In this study, particle filtration efficiencies of five commercially available smoke evacuation systems were investigated in a model operating room. Two cutting angles, the devices' suction capacities, three unidirectional displacement flow rates, and the noise exposures were considered. Results demonstrated that portable smoke evacuation systems reduce surgical smoke up to 99% under optimal conditions. A cutting angle of 45 degrees, the device's maximum suction capacity, and a unidirectional displacement flow rate of 10,500 m(3)/hr were advantageous. Sound levels ranged between 51-69 dBA and exceeded recommended threshold limits, if used with medium or maximum suction capacity. Hence, portable smoke evacuation systems are beneficial and are recommended. However, a combination with general unidirectional room ventilation and a strict limitation of the use of electrocauterization is strongly advised.
引用
收藏
页码:773 / 781
页数:9
相关论文
共 50 条
  • [41] Optimizing the Muffler of a Portable Lawn Mower Engine to reduce Noise Pollution Levels
    Sedighi, Sh.
    Kalantari, D.
    Redl, J.
    JOURNAL OF AGRICULTURAL MACHINERY, 2021, 11 (02)
  • [42] Multiagent Systems and Potential Fields to Smoke Dispersion Applied to Evacuation Simulations: The Case of Kiss Nightclub
    Correa, Bruna A.
    Bicho, Alessandro L.
    Adamatti, Diana F.
    APPLIED ARTIFICIAL INTELLIGENCE, 2019, 33 (11) : 1008 - 1021
  • [43] Low-cost, safe, and effective smoke evacuation device for surgical procedures in the COVID-19 age
    Andrade, Wesley Pereira
    Goncalves, Graziella Galvao
    Medeiros, Luciana Cortelazzi
    Miyamoto Araujo, Danielle Cristina
    Gomes Pereira, Gisele Tolaini
    Pereira Moraes, Daniela Martins
    Spencer, Ranyell M. S. S. B.
    JOURNAL OF SURGICAL ONCOLOGY, 2020, 122 (05) : 844 - 847
  • [44] NOISE-LEVELS IN COTTON GINNING SYSTEMS
    ANTHONY, WS
    WESLEY, RA
    MCCASKILL, OL
    TRANSACTIONS OF THE ASAE, 1978, 21 (03): : 574 - &
  • [45] Noise Levels in Surgical ICUs Are Consistently Above Recommended Standards
    Tainter, Christopher R.
    Levine, Alexander R.
    Quraishi, Sadeq A.
    Butterly, Arielle D.
    Stahl, David L.
    Eikermann, Matthias
    Kaafarani, Haytham M.
    Lee, Jarone
    CRITICAL CARE MEDICINE, 2016, 44 (01) : 147 - 152
  • [46] Escape of surgical smoke particles, comparing conventional and valveless trocar systems
    Daniel Robertson
    Frank Sterke
    Willem van Weteringen
    Alberto Arezzo
    Yoav Mintz
    Tim Horeman
    Surgical Endoscopy, 2023, 37 (11) : 8552 - 8561
  • [47] Escape of surgical smoke particles, comparing conventional and valveless trocar systems
    Robertson, Daniel
    Sterke, Frank
    van Weteringen, Willem
    Arezzo, Alberto
    Mintz, Yoav
    Horeman, Tim
    SURGICAL ENDOSCOPY AND OTHER INTERVENTIONAL TECHNIQUES, 2023, 37 (11): : 8552 - 8561
  • [48] Sound levels from headphone/portable compact disc player systems
    Keith, SE
    Bly, SHP
    Chiu, V
    Hussey, RG
    INTER-NOISE 99: PROCEEDINGS OF THE 1999 INTERNATIONAL CONGRESS ON NOISE CONTROL ENGINEERING, VOLS 1-3, 1999, : 1865 - 1870
  • [49] TEMPORARY PORTABLE AIRCRAFT-NOISE MONITORING INSTRUMENTATION SYSTEMS - PRACTICAL CONSIDERATIONS
    GREENE, R
    ENGINEERING FOR ENVIRONMENTAL NOISE CONTROL, VOLS 1 AND 2: INTER-NOISE 89, 1989, : 931 - 934
  • [50] Sound levels from headphone/portable compact disc player systems
    Radiation Protection Bur, Ottawa, Canada
    Can Acoust, 3 (74-75):