Time-resolved fluorescence immunoassay (TRFIA) for the simultaneous detection of hs-CRP and lipoprotein(a) in serum

被引:7
|
作者
Ma, Hongyan [1 ]
Mao, Qian [1 ]
Zhu, Yanbin [1 ]
Cong, Chunli [1 ]
Zheng, Shiyu [1 ]
Zhang, Qi [1 ]
Chen, Cuicui [2 ]
Li, Laiqing [2 ]
机构
[1] Beihua Univ, Dept Cardiol, Affiliated Hosp, Jilin, Jilin, Peoples R China
[2] Guangzhou Youdi Biotechnol Co Ltd, 3 Juquan Rd, Guangzhou 510663, Peoples R China
关键词
atherosclerosis; biomarker; high-sensitivity C-reactive protein; lipoprotein(a); serum; time-resolved fluorescence immunoassay; C-REACTIVE PROTEIN; ATHEROSCLEROSIS; MARKERS; INFLAMMATION;
D O I
10.1002/bab.2310
中图分类号
Q5 [生物化学]; Q7 [分子生物学];
学科分类号
071010 ; 081704 ;
摘要
Elevated serum high-sensitivity C-reactive protein (hs-CRP) and lipoprotein(a) (Lp(a)) levels are associated with the development of native coronary atherosclerosis. We aimed to establish a new method for the simultaneous detection of hs-CRP and Lp(a) to predict the development of atherosclerosis. A one- step time-resolved fluorescence immunoassay (TRFIA) with europium(III) (Eu3+) or samarium(III) (Sm3+) labelswas established, and the performance of this TRFIA (in terms of sensitivity, specificity, accuracy, and cutoff values) was evaluated using clinical serum samples and comparedwith those of registered kits. The sensitivity was 0.052 mu g/ml for hs-CRP and 0.64 mu g/ml for Lp(a). The intra-assay and inter-assay cross-reactivities (CVs) were very low, ranging from 2.05% to 4.67% for hs- CRP and from 2.42% to 6.43% for Lp(a). The CVs were very low (<0.34% and <2.65%, respectively) with five interferents. Additionally, there was a high Pearson coefficient between the present TRFIA method and the registered kits (R-2 = 0.9967 and 0.9906, respectively). These data indicate that this study developed a TRFIA method that can be used for the quantitative detection of hs- CRP and Lp(a) in serum with high sensitivity, specificity, and accuracy. This TRFIA provides a new method for predicting the development of atherosclerosis.
引用
收藏
页码:2617 / 2623
页数:7
相关论文
共 50 条
  • [1] Time-resolved Fluorescence Immunoassay (TRFIA) for the Simultaneous Detection of MMP-9 and Lp-PLA2 in Serum
    Hong-yan Ma
    Qian Mao
    Yan-bin Zhu
    Chun-li Cong
    Shi-yu Zheng
    Qi Zhang
    Cui-cui Chen
    Lai-qing Li
    Journal of Fluorescence, 2021, 31 : 1771 - 1777
  • [2] Time-resolved Fluorescence Immunoassay (TRFIA) for the Simultaneous Detection of MMP-9 and Lp-PLA2 in Serum
    Ma, Hong-Yan
    Mao, Qian
    Zhu, Yan-Bin
    Cong, Chun-Li
    Zheng, Shi-Yu
    Zhang, Qi
    Chen, Cui-Cui
    Li, Lai-Qing
    JOURNAL OF FLUORESCENCE, 2021, 31 (06) : 1771 - 1777
  • [3] Time-resolved fluorescence immunoassay of estradiol in serum
    Tang, D
    Wang, YC
    Chang, WB
    Ci, YX
    Guo, ZQ
    CHINESE JOURNAL OF ANALYTICAL CHEMISTRY, 1999, 27 (08) : 899 - 903
  • [4] SENSITIVE TIME-RESOLVED FLUORESCENCE IMMUNOASSAY OF SOMATOTROPIN IN SERUM
    KAHAN, I
    PAPANASTASIOUDIAMANDI, A
    ELLIS, G
    MAKELA, SK
    MCLAURIN, JA
    DCOSTA, M
    DIAMANDIS, EP
    CLINICAL CHEMISTRY, 1990, 36 (03) : 503 - 508
  • [5] Simultaneous detection of deoxynivalenol and zearalenone by dual-label time-resolved fluorescence immunoassay
    Zhang, Jue
    Gao, Lei
    Zhou, Bin
    Zhu, Lan
    Zhang, Yi
    Huang, Biao
    JOURNAL OF THE SCIENCE OF FOOD AND AGRICULTURE, 2011, 91 (02) : 193 - 197
  • [6] A dual-label time-resolved fluorescence immunoassay (TRFIA) for screening of Coronary atherosclerosis based on simultaneous detection of Lp-PLA2 and HsCRP
    Zhang, Zhong
    Liu, Xiaozhu
    Li, Yinfeng
    Huang, Ruolan
    Wang, Ling
    Li, Laiqing
    Chen, Cuicui
    Ou, Lijun
    Chang, Xiao
    Qiao, Qiujie
    Chen, Mingtai
    IMMUNOLOGY LETTERS, 2017, 182 : 12 - 17
  • [7] Time-resolved fluorescence immunoassay (TRFIA) for the detection of Escherichia coli O157:H7 in apple cider
    Yu, LSL
    Reed, SA
    Golden, MH
    JOURNAL OF MICROBIOLOGICAL METHODS, 2002, 49 (01) : 63 - 68
  • [8] Simultaneous detection of sulfamethazine and sulfaquinoxaline using a dual-label time-resolved fluorescence immunoassay
    Le, Tao
    Yan, Peifeng
    Liu, Jin
    Wei, Shu
    FOOD ADDITIVES AND CONTAMINANTS PART A-CHEMISTRY ANALYSIS CONTROL EXPOSURE & RISK ASSESSMENT, 2013, 30 (07): : 1264 - 1269
  • [9] Evaluation of fluorescence hs-CRP immunoassay for point-of-care testing
    Oh, SW
    Moon, JD
    Park, SY
    Jang, HJ
    Kim, JH
    Nahm, KB
    Choi, EY
    CLINICA CHIMICA ACTA, 2005, 356 (1-2) : 172 - 177
  • [10] Time-resolved fluorescence immunoassay of thyroxine in serum: Immobilized antigen approach
    Wu, FB
    Xu, YY
    Xu, T
    Wang, YZ
    Han, SQ
    ANALYTICAL BIOCHEMISTRY, 1999, 276 (02) : 171 - 176