CONTEMPORARY RADIATION DOSES IN INTERVENTIONAL CARDIOLOGY: A NATIONWIDE STUDY OF PATIENT SKIN DOSES IN FINLAND

被引:4
|
作者
Jarvinen, Jukka [1 ,2 ,3 ]
Sierpowska, Joanna [4 ]
Siiskonen, Teemu [5 ]
Husso, Minna [6 ]
Jarvinen, Hannu [5 ]
Kiviniemi, Tuomas [1 ]
Rissanen, Tuomas T. [4 ]
Lindholm, Carita [5 ]
Matikka, Hanna [6 ]
Larjava, Heli R. S. [7 ]
Makela, Timo J. [8 ]
Strengell, Satu [9 ]
Eskola, Markku [10 ,11 ]
Parviainen, Teuvo [5 ]
Hallinen, Elina [5 ]
Pirinen, Markku [5 ]
Kivela, Antti [6 ]
Teras, Mika [3 ,12 ]
机构
[1] Turku Univ Hosp, Dept Cardiol, Turku Heart Ctr, Turku 20521, Finland
[2] Turku Univ Hosp, Dept Radiol, Turku 20521, Finland
[3] Turku Univ Hosp, Dept Med Phys, Turku 20521, Finland
[4] Cent Hosp Northern Karelia, Dept Radiol, Joensuu 80210, Finland
[5] STUK Radiat & Nucl Safety Author, Helsinki 00880, Finland
[6] Kuopio Univ Hosp, Dept Radiol, Imaging Ctr, Kuopio 70210, Finland
[7] Cent Finland Hlth Care Dist, Dept Med Imaging, Jyvaskyla 40620, Finland
[8] Oulu Univ Hosp, Div Cardiol, Dept Internal Med, Oulu 90220, Finland
[9] Helsinki Univ Hosp, Div Cardiol, Heart & Lung Ctr, Helsinki 00029, Finland
[10] Tampere Univ Hosp, Heart Hosp, Dept Cardiol, Tampere, Finland
[11] Univ Tampere, Fac Med & Life Sci, Tampere 33520, Finland
[12] Univ Turku, Inst Biomed, Turku 20521, Finland
关键词
CORONARY-ANGIOGRAPHY; RADIOCHROMIC FILM; SPECTRA; TRENDS;
D O I
10.1093/rpd/ncz273
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
In contemporary interventional cardiology, for typical elderly patients, the most severe radiation-related harm to patients can be considered to come from skin exposures. In this paper, maximum local skin doses in cardiological procedures are explored with Gafchromic film dosimetry. Film and reader calibrations and reading were performed at the Secondary Standards Dosimetry Laboratory of the Radiation and Nuclear Safety Authority (STUK), and data were gathered from seven hospitals in Finland. As alert levels for early transient erythema, 200 Gycm(2) kerma area product (KAP) and 2000 mGy air kerma levels for transcatheter aortic valve implantations (TAVI) procedures are proposed. The largest doses were measured in TAVI (4158.8 mGy) and percutaneous coronary interventions (PCI) (941.68 mGy). Accuracies of the GE DoseWatch and Siemens CareMonitor skin dose estimates were reasonable, but more results are needed to reliably assess and validate the tools' capabilities and reliabilities. Uncertainty of the Gafchromic dosimetry was estimated as 9.1% for a calibration with seven data points and 19.3% for a calibration with five data points.
引用
收藏
页码:181 / 190
页数:10
相关论文
共 50 条
  • [1] CONTEMPORARY RADIATION DOSES IN INTERVENTIONAL CARDIOLOGY: A NATIONWIDE STUDY OF PATIENT DOSES IN FINLAND
    Jarvinen, Jukka
    Sierpowska, Joanna
    Siiskonen, Teemu
    Jarvinen, Hannu
    Kiviniemi, Tuomas
    Rissanen, Tuomas T.
    Matikka, Hanna
    Niskanen, Eini
    Hurme, Saija
    Larjava, Heli R. S.
    Makela, Timo J.
    Strengell, Satu
    Eskola, Markku
    Parviainen, Teuvo
    Hallinen, Elina
    Pirinen, Markku
    Kivela, Antti
    Teras, Mika
    RADIATION PROTECTION DOSIMETRY, 2019, 185 (04) : 483 - 493
  • [2] Patient Radiation Doses in Interventional Cardiology Procedures
    Pantos, Ioannis
    Patatoukas, Georgios
    Katritsis, Demosthenes G.
    Efstathopoulos, Efstathios
    CURRENT CARDIOLOGY REVIEWS, 2009, 5 (01) : 1 - 11
  • [3] Patient Radiation Doses in Interventional Cardiology Procedure
    Hiswara, E.
    Kartikasari, D.
    Nuraeni, N.
    Sofyan, H.
    Sandy, K. Y. P.
    ATOM INDONESIA, 2022, 48 (02) : 81 - 85
  • [4] Radiation doses to patients in interventional cardiology
    Benini, A.
    Pedersen, F.
    Helqvist, S.
    Kastrup, J.
    Kelbaek, H.
    Saunamaki, K.
    Jorgensen, E.
    EUROPEAN HEART JOURNAL, 2009, 30 : 537 - 537
  • [5] Staff Radiation Doses in Interventional Cardiology: Correlation With Patient Exposure
    Vano, Eliseo
    Ubeda, Carlos
    Leyton, Fernando
    Miranda, Patricia
    Gonzalez, Luciano
    PEDIATRIC CARDIOLOGY, 2009, 30 (04) : 409 - 413
  • [6] Patient radiation doses in paediatric interventional cardiology and optimization actions
    Ubeda, Carlos
    Vano, Eliseo
    Riquelme, Nemorino
    Aguirre, Daniel
    Vasquez, Hector
    Chavez, Cristian
    Dalmazzo, Dandaro
    RADIATION PHYSICS AND CHEMISTRY, 2020, 168
  • [7] Patient radiation doses in paediatric interventional cardiology procedures: a review
    Harbron, R. W.
    Dreuil, S.
    Bernier, M-O
    Pearce, M. S.
    Thierry-Chef, I.
    Chapple, C-L
    Baysson, H.
    JOURNAL OF RADIOLOGICAL PROTECTION, 2016, 36 (04) : R131 - R144
  • [8] Staff Radiation Doses in Interventional Cardiology: Correlation With Patient Exposure
    Eliseo Vano
    Carlos Ubeda
    Fernando Leyton
    Patricia Miranda
    Luciano Gonzalez
    Pediatric Cardiology, 2009, 30 : 409 - 413
  • [9] Evaluation of patient doses and lens radiation doses to interventional cardiologists in a nationwide survey in Chile
    Ubeda, Carlos
    Vano, Eliseo
    Gonzalez, Luciano
    Miranda, Patricia
    Valenzuela, Edith
    Vergara, Francisco
    Guarda, Eduardo
    RADIATION PROTECTION DOSIMETRY, 2013, 157 (01) : 36 - 43
  • [10] Patients skin radiation doses in a contemporary cohort of patients undergoing percutaneous interventional cardiology procedures.
    Nogales-Asensio, Juan
    Sanchez, Sara
    Martinez, Ana
    Yuste, Maria
    Ordiales, Jose
    Lopez-Minguez, Jose
    Gonzalez, Reyes
    Merchan, Antonio
    Martinez, Gines
    Diaz-Cortegana, Francisco
    Garcia-Serrano, Laura
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2012, 60 (17) : B92 - B92