DOSES TO OPERATORS DURING INTERVENTIONAL RADIOLOGY PROCEDURES: FOCUS ON EYE LENS AND EXTREMITY DOSIMETRY

被引:50
|
作者
Koukorava, C. [1 ,2 ]
Carinou, E. [1 ]
Simantirakis, G. [1 ,2 ]
Vrachliotis, T. G. [3 ]
Archontakis, E. [4 ]
Tierris, C. [3 ]
Dimitriou, P. [2 ]
机构
[1] GAEC, Athens 15310, Greece
[2] Univ Athens, Sch Med, Dept Med Phys, GR-11527 Athens, Greece
[3] Henry Dunant Hosp, Athens 11526, Greece
[4] Red Cross Hosp, Athens, Greece
关键词
RADIATION-EXPOSURE; PATIENT;
D O I
10.1093/rpd/ncq328
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
The present study is focused on the personnel doses during several types of interventional radiology procedures. Apart from the use of the official whole body dosemeters (thermoluminescence dosemeter type), measurements were performed to the extremities and the eyes using thermoluminescent loose pellets. The mean doses per kerma area product were calculated for the monitored anatomic regions and for the most frequent types of procedures. Higher dose values were measured during therapeutic procedures, especially embolisations. The maximum recorded doses during a single procedure were 1.8 mSv to the finger (nephrostomy), 2.1 mSv to the wrist (liver chemoembolisation), 0.6 mSv to the leg (brain embolisation) and 2.4 mSv to the eye (brain embolisation). The annual doses estimated for the operator with the highest workload according to the measurements and the system's log book were 90.4 mSv to the finger, 107.9 mSv to the wrist, 21.6 mSv to the leg and 49.3 mSv to the eye. Finally, the effect of the beam angulation (i.e. projection) and shielding equipment on the personnel doses was evaluated. The measurements were performed within the framework of the ORAMED (Optimization of RAdiation Protection for MEDical staff) project.
引用
收藏
页码:482 / 486
页数:5
相关论文
共 50 条
  • [1] ASSESSMENT OF EYE LENS DOSES FOR WORKERS DURING INTERVENTIONAL RADIOLOGY PROCEDURES
    Urboniene, A.
    Sadzeviciene, E.
    Ziliukas, J.
    [J]. RADIATION PROTECTION DOSIMETRY, 2015, 165 (1-4) : 299 - 303
  • [2] MONTE CARLO CALCULATIONS ON EXTREMITY AND EYE LENS DOSIMETRY FOR MEDICAL STAFF AT INTERVENTIONAL RADIOLOGY PROCEDURES
    Carinou, E.
    Ferrari, P.
    Koukorava, C.
    Krim, S.
    Struelens, L.
    [J]. RADIATION PROTECTION DOSIMETRY, 2011, 144 (1-4) : 492 - 496
  • [3] EXTREMITY AND EYE LENS DOSES OF INTERVENTIONAL RADIOLOGY AND CARDIOLOGY WORKERS
    Urboniene, Ausra
    Griciene, Birute
    [J]. MEDICAL PHYSICS IN THE BALTIC STATES, 2010, (02): : 109 - 111
  • [4] Assessment of occupational radiation doses to eye lens during interventional radiology procedures
    Ali, Amir
    Al-Nuaimi, Meshari
    Khan, Banaras
    Burki, Shamsullah
    Nishtar, Tahira
    Yaseen, Muhammad
    Shah, Syed Sher Azam
    Irfan, Muhammad
    [J]. RADIATION PHYSICS AND CHEMISTRY, 2024, 225
  • [5] EXTREMITY AND EYE LENS DOSES IN INTERVENTIONAL RADIOLOGY AND CARDIOLOGY PROCEDURES: FIRST RESULTS OF THE ORAMED PROJECT
    Domienik, J.
    Brodecki, M.
    Carinou, E.
    Donadille, L.
    Jankowski, J.
    Koukorava, C.
    Krim, S.
    Nikodemova, D.
    Ruiz-Lopez, N.
    Sans-Merce, M.
    Struelens, L.
    Vanhavere, F.
    [J]. RADIATION PROTECTION DOSIMETRY, 2011, 144 (1-4) : 442 - 447
  • [6] Recommendations to reduce extremity and eye lens doses in interventional radiology and cardiology
    Carinou, E.
    Brodecki, M.
    Domienik, J.
    Donadille, L.
    Koukorava, C.
    Krim, S.
    Nikodemova, D.
    Ruiz-Lopez, N.
    Sans-Merce, M.
    Struelens, L.
    Vanhavere, F.
    [J]. RADIATION MEASUREMENTS, 2011, 46 (11) : 1324 - 1329
  • [7] EXTREMITY DOSES DURING INTERVENTIONAL RADIOLOGY
    ODURNY, A
    [J]. CLINICAL RADIOLOGY, 1990, 42 (02) : 142 - 142
  • [8] EXTREMITY DOSES DURING INTERVENTIONAL RADIOLOGY
    RAMSDALE, ML
    WALKER, WJ
    HORTON, PW
    [J]. CLINICAL RADIOLOGY, 1990, 41 (01) : 34 - 36
  • [9] Occupational doses to the eye lens in pediatric and adult noncardiac interventional radiology procedures
    Morcillo, Ana Belen
    Alejo, Luis
    Huerga, Carlos
    Bayon, Jose
    Marin, Alberto
    Corredoira, Eva
    Novo, Joan Ricardo
    Hernandez, Teresa
    Ponce, Maria Dolores
    Garzon, Gonzalo
    Vano, Eliseo
    Guibelalde, Eduardo
    [J]. MEDICAL PHYSICS, 2021, 48 (04) : 1956 - 1966
  • [10] Eye lens dosimetry of workers during medical interventional procedures and surgery
    Bera, G.
    Gellie, G.
    Michel, X.
    [J]. EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2017, 44 : S810 - S811