In vitro evaluation of 56 coronary artery stents by 256-slice multi-detector coronary CT

被引:9
|
作者
Steen, Henning [1 ]
Andre, Florian [1 ]
Korosoglou, Grigorios [1 ]
Mueller, Dirk [2 ]
Hosch, Waldemar [3 ]
Kauczor, Hans-Ulrich [3 ]
Giannitsis, Evangelos [1 ]
Katus, Hugo A. [1 ]
机构
[1] Univ Heidelberg, Dept Cardiol, D-69120 Heidelberg, Germany
[2] Philips GmbH, Healthcare Div, D-20099 Hamburg, Germany
[3] Univ Heidelberg, Dept Diagnost & Intervent Radiol, D-69120 Heidelberg, Germany
关键词
Multislice CT; Coronary stents; CT angiography; Artifacts; Coronary artery disease; In-stent; Restenosis; SPIRAL COMPUTED-TOMOGRAPHY; DUAL-SOURCE CT; ROW CT; ANGIOGRAPHY; RESTENOSIS; VISUALIZATION;
D O I
10.1016/j.ejrad.2010.08.007
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objective: We sought to investigate stent lumen visibility of 56 coronary stents with the newest 256-multi-slice-CT (256-MDCT) technology for different reconstruction algorithms in an in vitro model. Background: Early identification of in-stent restenosis (ISR) is important to avoid recurrent ischemia and prevent acute myocardial infarction (AMI). Since angiography has the disadvantage of high costs and its invasiveness, MDCT could be a convenient and safe non-invasive alternative for detection of ISR. Material and methods: Percentages of in-stent lumen diameter and in-stent signal attenuation (measured as contrast-to-noise ratio (CNR)) of 56 coronary stents (group A <= 2.5 mm; group B = 2.75-3.0 mm; group C = 3.5-4.0 mm) were evaluated in a coronary vessel in vitro phantom (iodine-filled plastic tubes) employing four different reconstruction algorithms (XCD, CC, CD, XCB) on a novel 256-MDCT (Philips-iCT, collimation = 128 mm x 0.625 mm; rotation time = 270 ms; tube current = 800 mA s with 120 kV). Analysis was conducted with the semi-automatical full-width-at-half-maximum (FWHM) method. P-values <0.05 were regarded statistically significant. Results: In-stent lumen diameter >60% for group C stents was significantly larger and CNR was significantly lower (both p < 0.05) for sharp kernels (CD; XCD) when compared to groups A/B. The FWHM-method showed significantly smaller in-stent lumen diameter (p < 0.05) when compared to the manual method. Conclusion: 256-MDCT could potentially be employed for clinical assessment of stent patency in stents >3.0 mm when analysed with cardio-dedicated sharp kernels, although clinical studies corroborating this claim should be performed. However, stents <= 3.0 mm reconstructed by soft kernels revealed insufficient in-stent lumen visualisation and should not be used in clinical practice. Further improvements in spatial and temporal image resolution as well as reductions of radiation exposure and image noise have to be accomplished for the ambitious goal of characterising both CT coronary artery anatomy and in-stent lumen. (C) 2010 Elsevier Ireland Ltd. All rights reserved.
引用
收藏
页码:143 / 150
页数:8
相关论文
共 50 条
  • [21] Noninvasive Assessment Of Coronary Artery Bypass Graft With 64-Slice Multi-Detector Row CT: Comparison With Invasive Coronary Angiography
    Urdiales, Pedro L.
    De La Vega, Alejandro
    AMERICAN JOURNAL OF CARDIOLOGY, 2009, 104 (6A): : 73D - 73D
  • [22] Improved image quality at 256-slice coronary CT angiography in patients with a high heart rate and coronary artery disease: comparison with 64-slice CT imaging
    Oda, Seitaro
    Katahira, Kazuhiro
    Utsunomiya, Daisuke
    Takaoka, Hiroko
    Honda, Keiichi
    Noda, Katsuo
    Oshima, Shuichi
    Yuki, Hideaki
    Namimoto, Tomohiro
    Yamashita, Yasuyuki
    ACTA RADIOLOGICA, 2015, 56 (11) : 1308 - 1314
  • [23] Evaluation of patency of coronary-artery bypass graft using multi-detector row CT
    Kim, M
    Choi, B
    Choi, DH
    Choe, KO
    Choi, H
    Kim, J
    RADIOLOGY, 2002, 225 : 390 - 390
  • [24] The predictive value of 64-slice multi-detector computed tomography coronary angiography in the evaluation of coronary artery stent in a comparison to the invasive coronary angiography
    Das, K. M.
    El-Menyar, A.
    Salam, A.
    Dabdoob, W.
    Singh, R.
    Albinali, H.
    Al Suwaidi, J.
    EUROPEAN HEART JOURNAL, 2007, 28 : 880 - 881
  • [25] Coronary artery calcium measurement with multi-detector row CT: In vitro assessment of effect of radiation dose
    Hong, C
    Bae, KT
    Pilgram, TK
    Suh, J
    Bradley, D
    RADIOLOGY, 2002, 225 (03) : 901 - 906
  • [26] Evaluation of image quality and radiation dose at prospective ECG-triggered axial 256-slice multi-detector CT in infants with congenital heart disease
    Huang, Mei-ping
    Liang, Chang-hong
    Zhao, Zhen-jun
    Liu, Hui
    Li, Jing-lei
    Zhang, Jin-e
    Cui, Yan-hai
    Yang, Lin
    Liu, Qi-shun
    Ivanc, Thomas B.
    Vembar, Mani
    PEDIATRIC RADIOLOGY, 2011, 41 (07) : 858 - 866
  • [27] Evaluation of image quality and radiation dose at prospective ECG-triggered axial 256-slice multi-detector CT in infants with congenital heart disease
    Mei-ping Huang
    Chang-hong Liang
    Zhen-jun Zhao
    Hui Liu
    Jing-lei Li
    Jin-e Zhang
    Yan-hai Cui
    Lin Yang
    Qi-shun Liu
    Thomas B. Ivanc
    Mani Vembar
    Pediatric Radiology, 2011, 41 : 858 - 866
  • [28] Optimal Systolic and Diastolic Image Reconstruction Windows for Coronary 256-Slice CT Angiography
    Mok, Greta S. P.
    Yang, Ching-Ching
    Chen, Liang-Kuang
    Lu, Kun-Mu
    Law, Wei-Yip
    Wu, Tung-Hsin
    ACADEMIC RADIOLOGY, 2010, 17 (11) : 1386 - 1393
  • [29] The utility of multi-detector row spiral CT for detection of coronary artery stenoses
    Moon, JY
    Chung, N
    Choi, BW
    Choe, KO
    Seo, HS
    Ko, YG
    Kang, SM
    Ha, JW
    Rim, SJ
    Jang, YS
    Shim, WH
    Cho, SY
    YONSEI MEDICAL JOURNAL, 2005, 46 (01) : 86 - 94
  • [30] Multi-Detector Coronary CT Imaging for the Identification of Coronary Artery Stenoses in a "Real-World" Population
    Makaryus, Amgad N.
    Henry, Sonia
    Loewinger, Lee
    Makaryus, John N.
    Boxt, Lawrence
    CLINICAL MEDICINE INSIGHTS-CARDIOLOGY, 2014, 8 : 13 - 22