In vitro evaluation of 56 coronary artery stents by 256-slice multi-detector coronary CT

被引:9
|
作者
Steen, Henning [1 ]
Andre, Florian [1 ]
Korosoglou, Grigorios [1 ]
Mueller, Dirk [2 ]
Hosch, Waldemar [3 ]
Kauczor, Hans-Ulrich [3 ]
Giannitsis, Evangelos [1 ]
Katus, Hugo A. [1 ]
机构
[1] Univ Heidelberg, Dept Cardiol, D-69120 Heidelberg, Germany
[2] Philips GmbH, Healthcare Div, D-20099 Hamburg, Germany
[3] Univ Heidelberg, Dept Diagnost & Intervent Radiol, D-69120 Heidelberg, Germany
关键词
Multislice CT; Coronary stents; CT angiography; Artifacts; Coronary artery disease; In-stent; Restenosis; SPIRAL COMPUTED-TOMOGRAPHY; DUAL-SOURCE CT; ROW CT; ANGIOGRAPHY; RESTENOSIS; VISUALIZATION;
D O I
10.1016/j.ejrad.2010.08.007
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objective: We sought to investigate stent lumen visibility of 56 coronary stents with the newest 256-multi-slice-CT (256-MDCT) technology for different reconstruction algorithms in an in vitro model. Background: Early identification of in-stent restenosis (ISR) is important to avoid recurrent ischemia and prevent acute myocardial infarction (AMI). Since angiography has the disadvantage of high costs and its invasiveness, MDCT could be a convenient and safe non-invasive alternative for detection of ISR. Material and methods: Percentages of in-stent lumen diameter and in-stent signal attenuation (measured as contrast-to-noise ratio (CNR)) of 56 coronary stents (group A <= 2.5 mm; group B = 2.75-3.0 mm; group C = 3.5-4.0 mm) were evaluated in a coronary vessel in vitro phantom (iodine-filled plastic tubes) employing four different reconstruction algorithms (XCD, CC, CD, XCB) on a novel 256-MDCT (Philips-iCT, collimation = 128 mm x 0.625 mm; rotation time = 270 ms; tube current = 800 mA s with 120 kV). Analysis was conducted with the semi-automatical full-width-at-half-maximum (FWHM) method. P-values <0.05 were regarded statistically significant. Results: In-stent lumen diameter >60% for group C stents was significantly larger and CNR was significantly lower (both p < 0.05) for sharp kernels (CD; XCD) when compared to groups A/B. The FWHM-method showed significantly smaller in-stent lumen diameter (p < 0.05) when compared to the manual method. Conclusion: 256-MDCT could potentially be employed for clinical assessment of stent patency in stents >3.0 mm when analysed with cardio-dedicated sharp kernels, although clinical studies corroborating this claim should be performed. However, stents <= 3.0 mm reconstructed by soft kernels revealed insufficient in-stent lumen visualisation and should not be used in clinical practice. Further improvements in spatial and temporal image resolution as well as reductions of radiation exposure and image noise have to be accomplished for the ambitious goal of characterising both CT coronary artery anatomy and in-stent lumen. (C) 2010 Elsevier Ireland Ltd. All rights reserved.
引用
收藏
页码:143 / 150
页数:8
相关论文
共 50 条
  • [1] Coronary Artery Stenosis in Moyamoya Disease Tissue Characterization by 256-Slice Multi-Detector CT and Virtual Histology
    Lee, Ju-Hee
    Youn, Tae-Jin
    Yoon, Yeonyee E.
    Park, Jin Joo
    Hong, Su Jin
    Chun, Eun Ju
    Choi, Sang Il
    Cho, Young-Seok
    Cho, Goo-Yeong
    Chae, In-Ho
    Choi, Dong-Ju
    CIRCULATION, 2013, 127 (20) : 2063 - 2065
  • [2] Performance of dual source versus 256-slice multi-slice CT in the evaluation of 16 coronary artery stents
    Andre, Florian
    Korosoglou, Grigorios
    Hosch, Waldemar
    Giannitsis, Evangelos
    Kauczor, Hans-Ulrich
    Katus, Hugo A.
    Steen, Henning
    EUROPEAN JOURNAL OF RADIOLOGY, 2013, 82 (04) : 601 - 607
  • [3] Effect of Body Mass Index in Coronary CT Angiography Performed on a 256-Slice Multi-Detector CT Scanner
    Law, Wei-Yip
    Huang, Guan-Lin
    Yang, Ching-Ching
    DIAGNOSTICS, 2022, 12 (02)
  • [4] Reproducibility in the assessment of noncalcified coronary plaque with 256-slice multi-detector CT and automated plaque analysis software
    Lee, Min Su
    Chun, Eun Ju
    Kim, Kil Joong
    Kim, Jeong A.
    Vembar, Mani
    Choi, Sang Il
    INTERNATIONAL JOURNAL OF CARDIOVASCULAR IMAGING, 2010, 26 : 237 - 244
  • [5] Reproducibility in the assessment of noncalcified coronary plaque with 256-slice multi-detector CT and automated plaque analysis software
    Min Su Lee
    Eun Ju Chun
    Kil Joong Kim
    Jeong A Kim
    Mani Vembar
    Sang Il Choi
    The International Journal of Cardiovascular Imaging, 2010, 26 : 237 - 244
  • [6] Diagnostic accuracy of low-dose 256-slice multi-detector coronary CT angiography using iterative reconstruction in patients with suspected coronary artery disease
    Hou, Yang
    Ma, Yue
    Fan, Weipeng
    Wang, Yuke
    Yu, Mei
    Vembar, Mani
    Guo, Qiyong
    EUROPEAN RADIOLOGY, 2014, 24 (01) : 3 - 11
  • [7] Diagnostic accuracy of low-dose 256-slice multi-detector coronary CT angiography using iterative reconstruction in patients with suspected coronary artery disease
    Yang Hou
    Yue Ma
    Weipeng Fan
    Yuke Wang
    Mei Yu
    Mani Vembar
    Qiyong Guo
    European Radiology, 2014, 24 : 3 - 11
  • [8] CT coronary angiography: 256-slice and 320-detector row scanners
    Hsiao E.M.
    Rybicki F.J.
    Steigner M.
    Current Cardiology Reports, 2010, 12 (1) : 68 - 75
  • [9] Image quality of coronary CT angiography in 256-slice CT
    Motoyama, Sadako
    Sarai, Masayoshi
    Anno, Hirofumi
    Sato, Takahisa
    Inoue, Kaori
    Harigaya, Hiroto
    Sanda, Yoshihiro
    Hishida, Hitoshi
    Ozaki, Yukio
    Katada, Kazuhiro
    CIRCULATION, 2007, 116 (16) : 612 - 612
  • [10] The evaluation of coronary artery-to-pulmonary artery fistula in adulthood on 256-slice CT coronary angiography: Comparison with coronary catheter angiography and transthoracic echocardiography
    Li, Jing-Lei
    Huang, Lei
    Zhu, Wei
    Ye, Wei-Tao
    Yan, Li-Fen
    Zhong, Xiao-Mei
    Luo, Hai-Ying
    Saboo, Sachin
    Huang, Mei-Ping
    Liang, Chang-Hong
    JOURNAL OF CARDIOVASCULAR COMPUTED TOMOGRAPHY, 2019, 13 (01) : 75 - 80