Implant-abutment screw removal torque values between customized titanium abutment, straight titanium abutment, and hybrid zirconia abutment after a million cyclic loading: an in vitro comparative study

被引:7
|
作者
Klongbunjit, Disayut [1 ]
Aunmeungtong, Weerapan [1 ]
Khongkhunthian, Pathawee [1 ]
机构
[1] Chiang Mai Univ, Ctr Excellence Dent Implantol, Fac Dent, Chiang Mai 50200, Thailand
关键词
Bending moment; Removal torque; Cyclic loading; Abutment; DENTAL IMPLANT; COMPLICATION RATES; FRACTURE LOAD; FATIGUE; CROWNS; SURVIVAL; CONNECTION; RESISTANCE; TESTS;
D O I
10.1186/s40729-021-00378-z
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Purpose The aim of this study was to compare removal torque values after mechanical cyclic loading and bending moment after the static compression testing of customized titanium abutment compared with prefabricated and hybrid abutments. Materials and methods The study was developed according to ISO 14801:2016. Sixty implants were divided into three groups equally: Straight titanium abutment group, Customized titanium abutment group, and Hybrid zirconia abutment group. Abutments were fabricated with zirconia restoration. Forty five implants underwent for cyclic loading. The removal torque values were measured after a fatigue test was conducted at 0 cycles (control), 50,000 cycles and 1,000,000 cycles. In the second experiment, 15 implants were divided into the same groups. Then, bending moments were investigated. Results The mean initial removal torque value was significantly higher than 50,000 cycles and 1,000,000 cycles (P < 0.001). The comparison of mean removal torque value between types of abutments was not significantly different (P > 0.05), and the bending moments of all abutments were not significantly different (P > 0.05). Conclusions From the boundary of this in-vitro study, it could be concluded that customized titanium abutment and hybrid abutment were not significantly different in terms of removal torque values after the fatigue test. The bending moment between types of abutment were not significantly different. Thus, it could be concluded that abutment type does not significantly influence abutment stability or fracture strength.
引用
收藏
页数:9
相关论文
共 50 条
  • [21] Reverse Torque Value of Angulated Screw Channel Abutment before and after Cyclic Loading: An In Vitro Study
    Chen, Yu-Hsuan
    Wu, Yu-Ling
    Chen, Hung-Shyong
    Lin, Ching-Ping
    Wu, Aaron Yu-Jen
    JOURNAL OF FUNCTIONAL BIOMATERIALS, 2023, 14 (03)
  • [22] Comparative study of bacterial microfiltration in the implant-abutment interface, with straight and conical internal connections, in vitro
    Carlos, Larrucea, V
    Carlos, Navarro C.
    Karina, Larrucea S. M.
    Sunil, Boda K.
    Carlos, Padilla E.
    Olga, Lobos G.
    CLINICAL AND EXPERIMENTAL DENTAL RESEARCH, 2021, 7 (06): : 1014 - 1024
  • [23] Effect of Repeated Moist Heat Sterilization on Titanium Implant-Abutment Interface-An In Vitro Study
    Morsy, Mohamed S. M.
    Hassan, Ali Abdel-Halim Abdel-Azim
    Alshawkani, Hamed A.
    Mattoo, Khurshid A.
    Mathur, Ankita
    Fiorillo, Luca
    EUROPEAN JOURNAL OF DENTISTRY, 2024, 18 (03) : 860 - 868
  • [24] Sealing capability of implant-abutment junction under cyclic loading: a toluidine blue in vitro study
    Scarano, Antonio
    Perrotti, Vittoria
    Piattelli, Adriano
    Iaculli, Flavia
    Iezzi, Giovanna
    JOURNAL OF APPLIED BIOMATERIALS & FUNCTIONAL MATERIALS, 2015, 13 (03): : E293 - E295
  • [25] Bending Moments and Types of Failure of Zirconia and Titanium Abutments with Internal Implant-Abutment Connections: A Laboratory Study
    Leutert, Christian R.
    Stawarczyk, Bogna
    Truninger, Thomas C.
    Haemmerle, Christoph H. F.
    Sailer, Irena
    INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS, 2012, 27 (03) : 505 - 512
  • [26] Implant-Abutment Connection Deformation After Prosthetic Procedures: An In Vitro Study
    Micarelli, Costanza
    Canullo, Luigi
    Iannello, Giuliano
    INTERNATIONAL JOURNAL OF PROSTHODONTICS, 2015, 28 (03) : 282 - 286
  • [27] Three-Dimensional Deformation and Wear of Internal Implant-Abutment Connection: A Comparative Biomechanical Study Using Titanium and Zirconia
    Nam, Rae-Kyeong
    Lee, Sang J.
    Park, Eun-Jin
    Kwon, Ho-Beom
    Yoon, Hyung-In
    INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS, 2018, 33 (06) : 1279 - 1286
  • [28] Removal torque pattern of a combined cone and octalobule index implant-abutment connection at different cyclic loading: an in-vitro experimental study
    Benjaboonyazit, Kanyarin
    Chaijareenont, Pisaisit
    Khongkhunthian, Pathawee
    INTERNATIONAL JOURNAL OF IMPLANT DENTISTRY, 2019, 5 (1)
  • [29] Removal torque pattern of a combined cone and octalobule index implant-abutment connection at different cyclic loading: an in-vitro experimental study
    Kanyarin Benjaboonyazit
    Pisaisit Chaijareenont
    Pathawee Khongkhunthian
    International Journal of Implant Dentistry, 5
  • [30] Implications of Wear at the Titanium-Zirconia Implant-Abutment Interface on the Health of Peri-implant Tissues
    Tawse-Smith, Andrew
    Ma, Sunyoung
    Duncan, Warwick J.
    Gray, Andrew
    Reid, Malcolm R.
    Rich, Alison M.
    INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS, 2017, 32 (03) : 599 - 609