Implant-abutment screw removal torque values between customized titanium abutment, straight titanium abutment, and hybrid zirconia abutment after a million cyclic loading: an in vitro comparative study

被引:7
|
作者
Klongbunjit, Disayut [1 ]
Aunmeungtong, Weerapan [1 ]
Khongkhunthian, Pathawee [1 ]
机构
[1] Chiang Mai Univ, Ctr Excellence Dent Implantol, Fac Dent, Chiang Mai 50200, Thailand
关键词
Bending moment; Removal torque; Cyclic loading; Abutment; DENTAL IMPLANT; COMPLICATION RATES; FRACTURE LOAD; FATIGUE; CROWNS; SURVIVAL; CONNECTION; RESISTANCE; TESTS;
D O I
10.1186/s40729-021-00378-z
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Purpose The aim of this study was to compare removal torque values after mechanical cyclic loading and bending moment after the static compression testing of customized titanium abutment compared with prefabricated and hybrid abutments. Materials and methods The study was developed according to ISO 14801:2016. Sixty implants were divided into three groups equally: Straight titanium abutment group, Customized titanium abutment group, and Hybrid zirconia abutment group. Abutments were fabricated with zirconia restoration. Forty five implants underwent for cyclic loading. The removal torque values were measured after a fatigue test was conducted at 0 cycles (control), 50,000 cycles and 1,000,000 cycles. In the second experiment, 15 implants were divided into the same groups. Then, bending moments were investigated. Results The mean initial removal torque value was significantly higher than 50,000 cycles and 1,000,000 cycles (P < 0.001). The comparison of mean removal torque value between types of abutments was not significantly different (P > 0.05), and the bending moments of all abutments were not significantly different (P > 0.05). Conclusions From the boundary of this in-vitro study, it could be concluded that customized titanium abutment and hybrid abutment were not significantly different in terms of removal torque values after the fatigue test. The bending moment between types of abutment were not significantly different. Thus, it could be concluded that abutment type does not significantly influence abutment stability or fracture strength.
引用
收藏
页数:9
相关论文
共 50 条
  • [1] Implant-abutment screw removal torque values between customized titanium abutment, straight titanium abutment, and hybrid zirconia abutment after a million cyclic loading: an in vitro comparative study
    Disayut Klongbunjit
    Weerapan Aunmeungtong
    Pathawee Khongkhunthian
    International Journal of Implant Dentistry, 7
  • [2] The behavior of a zirconia or metal abutment on the implant-abutment interface during cyclic loading
    Queiroz, Daher Antonio
    Hagee, Natalia
    Lee, Damian J.
    Zheng, Fengyuan
    JOURNAL OF PROSTHETIC DENTISTRY, 2020, 124 (02): : 211 - 216
  • [3] Survival of Four Conical Implant Abutment Connections After Removal of the Abutment Screw and Simulated Cyclic Loading: An In Vitro Comparative Study
    Yao, Kuang-Ta
    Chang, Tsai-Yu
    Huang, Guan-Jhong
    Fang, Hsu-Wei
    Wang, Ding-Han
    Hsu, Ming-Lun
    JOURNAL OF ORAL IMPLANTOLOGY, 2023, 49 (04) : 393 - 400
  • [4] Wear at the titanium-titanium and the titanium-zirconia implant-abutment interface: A comparative in vitro study
    Stimmelmayr, Michael
    Edelhoff, Daniel
    Gueth, Jan-Frederik
    Erdelt, Kurt
    Happe, Arndt
    Beuer, Florian
    DENTAL MATERIALS, 2012, 28 (12) : 1215 - 1220
  • [5] In Vitro Evaluation of Manual Torque Values Applied to Implant-Abutment Complex by Different Clinicians and Abutment Screw Loosening
    Kose, Onur Dincer
    Karatasli, Burcin
    Demircan, Sabit
    Kose, Taha Emre
    Cene, Erhan
    Aya, Serhan Aydjn
    Erdem, Mehmet Ali
    Cankaya, Abdulkadir Burak
    BIOMED RESEARCH INTERNATIONAL, 2017, 2017
  • [6] Implant-Abutment Misfit after Cyclic Loading: An In Vitro Experimental Study
    de Vasconcelos, John Eversong Lucena
    De Matos, Jefferson David Melo
    Queiroz, Daher Antonio
    Lopes, Guilherme da Rocha Scalzer
    De Lacerda, Bruna Caroline Goncalves Vasconcelos
    Bottino, Marco Antonio
    Turssi, Cecilia Pedroso
    Basting, Roberta Tarkany
    do Amaral, Flavia Lucisano Botelho
    Franca, Fabiana Mantovani Gomes
    MATERIALS, 2022, 15 (15)
  • [7] Wear at the Titanium-Zirconia Implant-Abutment Interface: A Pilot Study
    Klotz, Michael W.
    Taylor, Thomas D.
    Goldberg, A. Jon
    INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS, 2011, 26 (05) : 970 - 975
  • [8] Effect of Serrating Abutment-Implant Mating Surface on Torque Stability of Implant-Abutment Connection, Before and After Cyclic Loading
    Rezayani, Vida
    Alikhasi, Marzieh
    Monzavi, Abbas
    JOURNAL OF ORAL IMPLANTOLOGY, 2020, 46 (06) : 555 - 561
  • [9] Screw loosening and changes in removal torque relative to abutment screw length in a dental implant with external abutment connection after oblique cyclic loading
    Lee, Joo-Hee
    Cha, Hyun-Suk
    JOURNAL OF ADVANCED PROSTHODONTICS, 2018, 10 (06): : 415 - 421
  • [10] The Evaluation of Unscrewing Torque Values of Implant-Abutment Connections: An In Vitro Study
    Bruna, Ezio
    Fabianelli, Andrea
    Mastriforti, Giacomo
    Papacchini, Federica
    INTERNATIONAL JOURNAL OF PROSTHODONTICS, 2017, 30 (01) : 30 - 32