The holding power of Schanz pins in external fixation in the lower extremity

被引:7
|
作者
Pommer, A [1 ]
David, A [1 ]
Barczik, P [1 ]
Muhr, G [1 ]
机构
[1] Ruhr Univ Bochum, Berufsgenossenschaftliche Kliniken Bergmannsheil, Chirurg Klin & Poliklin, D-44789 Bochum, Germany
来源
UNFALLCHIRURG | 1998年 / 101卷 / 09期
关键词
Schanz pins; external fixation;
D O I
10.1007/s001130050327
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
We analyzed the failure of Schanz pins in 234 patients with fractures of the lower extremity. All patients had the AO external fixator and stainless steel pins (ISO 58/32-1). The mean duration of treatment was 14 weeks. During this time 88 of a total of 1147 pins became loose. The first failures occurred 2 weeks after insertion. Most pins failed after 5-6 weeks. Early loosening was not accompanied signs of local infection, while late failures regularly developed pin-tract infections. Pins next to the fracture gap exhibited the highest rate of loosening. Intermediate pins had the lowest one. We could not detect an effect of the configuration of the external fixation. Eighty-one percent of the failed pins were exchanged consequently. We conclude that the early occurrence of pin loosening is due to mechanical reasons. Significant factors for this complications were the location close to the fracture gap and the number of pins in that segment.
引用
收藏
页码:708 / 712
页数:5
相关论文
共 50 条
  • [21] Treatment of external fixation pins about the wrist - Reply
    Egol, Kenneth A.
    Hiebert, Rudi
    Paksima, Nader
    Koval, Kenneth J.
    JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2006, 88A (11): : 2535 - 2535
  • [22] EXTERNAL FIXATION OF EXTREMITY FLAPS AND GRAFTS
    BINDIGER, A
    ECKER, J
    KUSCHNER, S
    SHERMAN, R
    ANNALS OF PLASTIC SURGERY, 1991, 26 (02) : 198 - 199
  • [23] BIOMECHANICAL STUDIES ON FRAMEWORK AND INSERTION OF PINS OF EXTERNAL FIXATION
    OONISHI, H
    TATSUMI, M
    HASEGAWA, T
    ORTHOPEDICS, 1984, 7 (04) : 658 - 668
  • [24] CORACOACROMIAL PINS TO AUGMENT EXTERNAL FIXATION OF THE UPPER HUMERUS
    PATANKAR, H
    KAKATKAR, V
    SHAH, C
    AGASHE, V
    JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1995, 77B (04): : 660 - 661
  • [25] An in Vivo Experimental Comparison of Stainless Steel and Titanium Schanz Screws for External Fixation
    Antonia Ganser
    Rosemary E. Thompson
    Ivan Tami
    Dirk Neuhoff
    Adrian Steiner
    Keita Ito
    European Journal of Trauma and Emergency Surgery, 2007, 33 : 59 - 68
  • [26] Compression External Fixation With Transosseous Pins for Arthrodesis of the Ankle
    Bushnell, Brandon D.
    Byram, Ian R.
    Dahners, Laurence E.
    TECHNIQUES IN FOOT AND ANKLE SURGERY, 2006, 5 (02): : 74 - 83
  • [27] Heat conduction of fixator pins with polymethylmethacrylate external fixation
    Williams, N
    Tomlinson, JL
    Hahn, AW
    Constantinescu, GM
    WagnerMann, C
    VETERINARY AND COMPARATIVE ORTHOPAEDICS AND TRAUMATOLOGY, 1997, 10 (03) : 153 - 159
  • [28] Temporary external fixation for the management of complex intra- and periarticular fractures of the lower extremity
    Haidukewych, GJ
    JOURNAL OF ORTHOPAEDIC TRAUMA, 2002, 16 (09) : 678 - 685
  • [29] Comminuted tibial fracture treatment with type II frame external fixators with Maynard clamps and Schanz pins
    Adamiak, Z.
    Piorek, A.
    POLISH JOURNAL OF VETERINARY SCIENCES, 2009, 12 (02): : 275 - 277
  • [30] Hydroxyapatite coated external fixation pins - An experimental study
    Caja, VL
    Moroni, A
    CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 1996, (325) : 269 - 275