Neutrons absorbed dose rate calculations for interstitial brachytherapy with 252Cf sources

被引:7
|
作者
Paredes, L.
Azorin, J.
Balcazar, M.
Francois, J. L.
机构
[1] Inst Nacl Invest Nucl, Ocoyoacac 52750, Mex, Mexico
[2] Univ Autonoma Estado Mex, Fac Med, Toluca 50180, Mex, Mexico
[3] Univ Autonoma Metropolitana Iztapalapa, Dept Fis, Mexico City 09340, DF, Mexico
[4] Univ Nacl Autonoma Mexico, Fac Ingn, Dept Sist Energet, Mexico City 04510, DF, Mexico
关键词
californium; neutron dosimetry; brachytherapy; Monte Carlo code;
D O I
10.1016/j.nima.2007.05.239
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The AAPM TG-43 protocol was used to evaluate the neutron absorbed dose rate from Cf-252 sources for two tissue substitute materials, five normal tissues and six malignant tumours. Three Cf-252 source models (AT, VariSource and mu Selectron) were simulated using the Monte Carlo MCNPX code in spherical geometry for obtaining the following factors: (a) conversion coefficient of neutron air kerma strength, (b) neutron dose rate constant, (c) neutron radial dose function, (d) geometry function, (e) anisotropy function and (f) neutron absorbed dose rate. For the Cf-252 AT source and a Watt fission spectrum, the calculated reference neutron dose rate in water was D-N(r(0), theta(0)) = 1.916cGy h(-1) mu g(-1). The results for the reference neutron dose rate in normal tissues and malignant tumours show how small variations in their elemental composition produce variations in the neutron absorbed dose rate up to 14.9% and 8.4%, respectively, when water and muscle are the reference medium. (C) 2007 Elsevier B.V. All rights reserved.
引用
收藏
页码:582 / 585
页数:4
相关论文
共 50 条
  • [1] Burst calculations for 252Cf brachytherapy sources
    Rivard, MJ
    MEDICAL PHYSICS, 2000, 27 (12) : 2816 - 2820
  • [2] Neutron absorbed dose rate with 252Cf sources for medical applications
    Paredes, L.
    Azorin, J.
    Balcazar, M.
    Francois, J. L.
    REVISTA MEXICANA DE FISICA, 2010, 56 (01) : 44 - 46
  • [3] Boron dose enhancement for 252Cf brachytherapy
    Allen, BJ
    Ralston, A
    ADVANCES IN NEUTRON CAPTURE THERAPY, VOLS I AND II: VOL I: MEDICINE AND PHYSICS, VOL II: CHEMISTRY AND BIOLOGY, 1997, 1132 : A271 - A274
  • [4] Measurement of dose enhancement by neutron capture of boron in 252Cf interstitial brachytherapy
    Abd El-Hafez, AI
    Eissa, HM
    Eman, SA
    Fadel, MA
    RADIATION MEASUREMENTS, 1997, 28 (1-6) : 441 - 444
  • [5] GAMMA DOSE RATE NEAR A NEW 252Cf BRACHYTHERAPY SOURCE
    Fortune, Eugene C.
    Gauld, Ian C.
    Wang, C. -K. Chris
    NUCLEAR TECHNOLOGY, 2011, 175 (01) : 73 - 76
  • [6] Dosimetry for 252Cf neutron emitting brachytherapy sources: Protocol, measurements, and calculations
    Department of Radiation Oncology, Tufts University School of Medicine, New England Medical Center, #246, 750 Washington Street, Boston, MA 02111, United States
    Med. Phys., 8 (1503-1514):
  • [7] CALCULATIONS OF ANISOTROPY FACTORS AND DOSE EQUIVALENTS FOR UNMODERATED 252CF SOURCES
    HERTEL, NE
    MCDONALD, JC
    RADIATION PROTECTION DOSIMETRY, 1990, 32 (02) : 77 - 82
  • [8] Derivation of the relative biological effectiveness of high dose rate 252Cf brachytherapy
    Rivard, M
    Zinkin, H
    Melhus, C
    MEDICAL PHYSICS, 2005, 32 (06) : 2063 - 2063
  • [9] Measured gadolinium neutron capture dose enhancement using 252Cf brachytherapy sources
    Rivard, MJ
    Waid, DS
    Wierzbicki, JG
    Yudelev, M
    ADVANCES IN NEUTRON CAPTURE THERAPY, VOLS I AND II: VOL I: MEDICINE AND PHYSICS, VOL II: CHEMISTRY AND BIOLOGY, 1997, 1132 : B430 - B435
  • [10] Development of miniature high-dose-rate 252Cf sources for boron-enhanced and fast neutron brachytherapy
    Martin, RC
    Halpern, DS
    RESEARCH AND DEVELOPMENT IN NEUTRON CAPTURE THERAPY, 2002, : 111 - 114