A Facile Fabrication of a Potentiometric Arrayed Glucose Biosensor Based on Nafion-GOx/GO/AZO

被引:26
|
作者
Chou, Jung-Chuan [1 ]
Lin, Si-Hong [1 ]
Lai, Tsu-Yang [1 ]
Kuo, Po-Yu [1 ]
Lai, Chih-Hsien [1 ]
Nien, Yu-Hsun [2 ]
Su, Tzu-Yu [2 ]
机构
[1] Natl Yunlin Univ Sci & Technol, Grad Sch Elect Engn, Touliu 64002, Yunlin, Taiwan
[2] Natl Yunlin Univ Sci & Technol, Grad Sch Chem & Mat Engn, Touliu 64002, Yunlin, Taiwan
关键词
glucose; zinc oxide (ZnO); aluminum-doped zinc oxide (AZO); graphene oxide (GO); potentiometric biosensor; arrayed electrodes; ELECTROCHEMICAL IMPEDANCE SPECTROSCOPY; GRAPHENE OXIDE; THIN-FILM; AL; IMMOBILIZATION; OXIDATION; NANOWIRES; PLATFORM; DEVICES; SENSORS;
D O I
10.3390/s20040964
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
In this study, the potentiometric arrayed glucose biosensors, which were based on zinc oxide (ZnO) or aluminum-doped zinc oxide (AZO) sensing membranes, were fabricated by using screen-printing technology and a sputtering system, and graphene oxide (GO) and Nafion-glucose oxidase (GOx) were used to modify sensing membranes by using the drop-coating method. Next, the material properties were characterized by using a Raman spectrometer, a field-emission scanning electron microscope (FE-SEM), and a scanning probe microscope (SPM). The sensing characteristics of the glucose biosensors were measured by using the voltage-time (V-T) measurement system. Finally, electrochemical impedance spectroscopy (EIS) was conducted to analyze their charge transfer abilities. The results indicated that the average sensitivity of the glucose biosensor based on Nafion-GOx/GO/AZO was apparently higher than that of the glucose biosensor based on Nafion-GOx/GO/ZnO. In addition, the glucose biosensor based on Nafion-GOx/GO/AZO exhibited an excellent average sensitivity of 15.44 mV/mM and linearity of 0.997 over a narrow range of glucose concentration range, a response time of 26 s, a limit of detection (LOD) of 1.89 mM, and good reproducibility. In terms of the reversibility and stability, the hysteresis voltages (V-H) were 3.96 mV and 2.42 mV. Additionally, the glucose biosensor also showed good anti-inference ability and reproducibility. According to these results, it is demonstrated that AZO is a promising material, which could be used to develop a reliable, simple, and low-cost potentiometric glucose biosensor.
引用
收藏
页数:20
相关论文
共 50 条
  • [1] Fabrication and Investigation of Arrayed Glucose Biosensor Based on Microfluidic Framework
    Chou, Jung-Chuan
    Cheng, Tsung-Yi
    Ye, Guan-Chen
    Liao, Yi-Hung
    Yang, Shu-Ying
    Chou, Hsueh-Tao
    IEEE SENSORS JOURNAL, 2013, 13 (11) : 4180 - 4187
  • [2] A Novel and High Performance Potentiometric Arrayed Flexible Glucose Biosensor Based on Microfluidic Device
    Chou, Jung-Chuan
    Tsai, Ya-Li
    Cheng, Tsung-Yi
    Liao, Yi-Hung
    Lin, Jyun-Wun
    Chen, Jie-Ting
    Chou, Hsueh-Tao
    2013 13TH IEEE CONFERENCE ON NANOTECHNOLOGY (IEEE-NANO), 2013, : 561 - 565
  • [3] Fabrication of Potentiometric Enzymatic Glucose Biosensor Based on Graphene and Magnetic Beads
    Chou, Jung-Chuan
    Chen, Ruei-Ting
    Liao, Yi-Hung
    Lin, Jyun-Wun
    Lin, Chin-Yi
    Jhang, Chiung-Yun
    Chou, Hsueh-Tao
    IEEE SENSORS JOURNAL, 2015, 15 (09) : 5278 - 5284
  • [4] Fabrication of Arrayed Flexible Screen-Printed Glucose Biosensor Based on Microfluidic Framework
    Chou, Jung-Chuan
    Tsai, Ya-Li
    Cheng, Tsung-Yi
    Liao, Yi-Hung
    Ye, Guan-Chen
    Yang, Shu-Ying
    IEEE SENSORS JOURNAL, 2014, 14 (01) : 178 - 183
  • [5] Facile Fabrication of a Graphene-based Electrochemical Biosensor for Glucose Detection
    Zhang Yanqin
    Fan Youjun
    Wang Shanshan
    Tan Yiliang
    Shen Xingcan
    Shi Zujin
    CHINESE JOURNAL OF CHEMISTRY, 2012, 30 (05) : 1163 - 1167
  • [6] Simple Fabrication of Glucose Biosensor Based on Graphene-Nafion Composite by Amperometric Detections
    Lee, Wei-Che
    Kuo, Chun-Chuan
    Chiu, Nan-Fu
    2012 IEEE SENSORS PROCEEDINGS, 2012, : 1625 - 1628
  • [7] Fabrication and Research of Flexible Arrayed Glucose Biosensor Combined with the Magnetic Beads
    Lin, Jyun-Wun
    Chou, Jung-Chuan
    Liao, Yi-Hung
    Cheng, Tsung-Yi
    Tsai, Ya-Li
    Jhang, Chiung-Yun
    Chou, Hsueh-Tao
    2013 13TH IEEE CONFERENCE ON NANOTECHNOLOGY (IEEE-NANO), 2013, : 289 - 292
  • [8] A potentiometric polypyrrole-based glucose biosensor
    Trojanowicz, M
    Hitchman, ML
    ELECTROANALYSIS, 1996, 8 (03) : 263 - 266
  • [9] An Electrochemical Enzymatic Biosensor Based on Au/FGs/sol-gel-GOx Composite/Nafion
    Hossain, Mohammad Faruk
    Heo, Min
    Shin, Jae Ho
    Park, Jae Yeong
    INTERNATIONAL JOURNAL OF ELECTROCHEMICAL SCIENCE, 2015, 10 (08): : 6803 - 6819
  • [10] Design and Fabrication of Enzymatic Potentiometric Biosensor Based on Flexible Printed Circuit Board for Glucose Detection
    Kuo, Po-Yu
    Liao, Chi-Han
    Wang, Tai-Hui
    Hsu, Ming-Tai
    IEEE TRANSACTIONS ON NANOBIOSCIENCE, 2024, 23 (02) : 283 - 290