Influence of electromagnetic fields on function of automated external defibrillators

被引:6
|
作者
Fleischhackl, R
Singer, F
Nitsche, W
Gamperl, G
Roessler, B
Arrich, J
Fleischhackl, S
Losert, H
Sterz, F
Mittlboeck, M
Hoerauf, K
机构
[1] Med Univ Vienna, Dept Emergency Med, A-1090 Vienna, Austria
[2] Med Univ Vienna, Dept Anaesthesia & Intens Care, A-1090 Vienna, Austria
[3] Med Univ Vienna, Core Unit Med Stat & Informat, A-1090 Vienna, Austria
[4] Vienna Red Cross Res Inst, Vienna, Austria
[5] Fed Inst Technol State Inst Elect & Elect Engn, Vienna, Austria
关键词
automated external defibrillator; electromagnetic interference; public access defibrillation;
D O I
10.1197/j.aem.2005.07.036
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Objectives: In this study, the authors tested whether electromagnetic interference (EMI) is able to impair correct electrocardiogram analysis and produce false-positive shock advice from automated external defibrillators (AEDs) when the true rhythm is sinus. Methods: Nineteen healthy subjects were used to test five AEDs available on the Austrian market in a prospective, open, and sequence-randomized study. The primary outcome variable was the absolute number of shocks advised in the presence of EMI. The secondary outcome was the number of impaired analyses caused by incorrectly detected patient movements or electrode failure. Results: Of 760 tests run, 18 (2.37%) cases or false-positive results occurred, and two of five AFDs recommended shocks in the presence of sinus rhythm. Of 760 tests run, no electrode failures occurred. There were 27 occurrences (3.55%) of motion detected by an AED in the presence of strong electromagnetic fields. Conclusions: AED models differ in their response to EMI; it may be useful to consider specific safety requirements for areas with such fields present. Working personnel and emergency medical services staff should be informed about potential risks and the possible need for patient evacuation before AEDs are attached and shock recommendations are followed.
引用
收藏
页码:1 / 6
页数:6
相关论文
共 50 条
  • [31] Are there disparities in the location of automated external defibrillators in England?
    Brown, Terry P.
    Perkins, Gavin D.
    Smith, Christopher M.
    Deakin, Charles D.
    Fothergill, Rachael
    RESUSCITATION, 2022, 170 : 28 - 35
  • [32] The shocking truth about automated external defibrillators
    Brown, J
    Kellermann, AL
    JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 2000, 284 (11): : 1438 - 1441
  • [33] PERFORMANCE OF AUTOMATED EXTERNAL DEFIBRILLATORS IN RURAL COMMUNITIES
    STULTS, K
    BROWN, D
    KERBER, R
    CIRCULATION, 1985, 72 (04) : 9 - 9
  • [34] Automated External Defibrillators and the Law of Unintended Consequences
    Haines, David E.
    JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 2010, 304 (19): : 2178 - 2179
  • [35] Importance of proper management of automated external defibrillators
    Oh, Je Hyeok
    RESUSCITATION, 2022, 178 : 8 - 9
  • [36] Automated External Defibrillators in Health and Fitness Facilities
    Drezner, Jonathan A.
    Asif, Irfan M.
    Harmon, Kimberly G.
    PHYSICIAN AND SPORTSMEDICINE, 2011, 39 (02): : 114 - 118
  • [37] Automated External Defibrillators Utility, Rational, and Controversies
    Berger, Stuart
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2009, 32 : S75 - S79
  • [38] Automated external defibrillators in the Australian fitness industry
    Norton, Kevin I.
    Norton, Lynda H.
    JOURNAL OF SCIENCE AND MEDICINE IN SPORT, 2008, 11 (02) : 86 - 89
  • [39] Automated external defibrillators in health/fitness facilities
    Balady, GJ
    Chaitman, B
    Foster, C
    Froelicher, E
    Gordon, N
    Van Camp, S
    MEDICINE AND SCIENCE IN SPORTS AND EXERCISE, 2002, 34 (03): : 561 - 564
  • [40] Charging up the public for automated external defibrillators
    Paula Willoughby DeJesus
    Critical Care, 9