Evaluation of a Nasal Cannula in Noninvasive Ventilation Using a Lung Simulator

被引:35
|
作者
Iyer, Narayan P. [1 ]
Chatburn, Robert [2 ,3 ]
机构
[1] Case Western Reserve Univ, Dept Pediat, MetroHlth Med Ctr, Div Neonatol, Cleveland, OH 44106 USA
[2] Cleveland Clin, Resp Inst, Cleveland, OH 44121 USA
[3] Case Western Reserve Univ, Dept Med, Lerner Coll Med, Cleveland, OH 44106 USA
关键词
noninvasive ventilation; bronchopulmonary dysplasia; lung distending pressure; nasal cannula; RAM cannula; POSITIVE AIRWAY PRESSURE; BIRTH-WEIGHT INFANTS; CONTROLLED-TRIAL; EXTUBATION; CPAP;
D O I
10.4187/respcare.03560
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
BACKGROUND: Nasal noninvasive ventilation (NIV) is a common form of noninvasive respiratory mode used in newborn infants. A next-generation nasal cannula (Neotech RAM cannula) has recently been used to provide nasal NIV. The impact of the Neotech RAM cannula on the delivery of pressure needs to be studied. METHODS: In this ex vivo experimental design, a lung simulator (IngMar ASL 5000, version 3.4) was programmed to model a neonate (similar to 1-3 kg of body weight) with normal-to-moderately affected lungs. We used a Covidien PB840 ventilator with NIV software activated to compensate for leaks. Nasal NIV was set at peak airway pressures of 15, 20, and 25 cm H2O and PEEP of 5, 6, and 7 cm H2O. Three sizes of the Neotech RAM cannula were used (prong outer diameters of 3.0, 3.5, and 4.0 mm). The nose was designed to keep the leak of the nares by the prongs to 30%. We also created a worst case leak (58% leak) by using the largest simulated nostril diameter with the smallest diameter Neotech RAM cannula prong. The outcome measure was the difference in pressures, referred to as leak effect, measured by the lung simulator relative to the set peak airway pressure and PEEP on the ventilator. RESULTS: For the interface with 30% leak, leak effects of peak airway pressure during simulated nasal NIV were similar with all Neotech RAM cannula sizes, with 63-75% of peak airway pressure and 70-90% of PEEP being transmitted across the nasal interface. The worst case scenario produced a 92% leak effect in peak airway pressure and PEEP. CONCLUSIONS: When used with <= 30% leak, the Neotech RAM cannula interface results in clinically acceptable transmission of pressures. With > 50% leak, a clinically negligible amount of pressure is transmitted to the artificial lungs.
引用
收藏
页码:508 / 512
页数:5
相关论文
共 50 条
  • [21] High-Efficiency Generation and Delivery of Aerosols Through Nasal Cannula During Noninvasive Ventilation
    Longest, P. Worth
    Walenga, Ross L.
    Son, Yoen-Ju
    Hindle, Michael
    JOURNAL OF AEROSOL MEDICINE AND PULMONARY DRUG DELIVERY, 2013, 26 (05) : 266 - 279
  • [22] Respiratory support after extubation: noninvasive ventilation or high-flow nasal cannula, as appropriate
    Mauri, Tommaso
    Grasselli, Giacomo
    Jaber, Samir
    ANNALS OF INTENSIVE CARE, 2017, 7
  • [23] Bench Testing of a Bubble Noninvasive Ventilation Device in an Infant Lung Simulator
    John, Stephen C.
    John, Anna, V
    Moss, Albert W.
    Gustafson, Peter A.
    Fernando-Silva, Luis
    John, Sunil P.
    RESPIRATORY CARE, 2020, 65 (09) : 1339 - 1345
  • [24] Noninvasive mechanical ventilation of mice by using a balloon-tipped cannula
    Winning, J
    Huwer, H
    Redjai, J
    Naumann, C
    Winning, I
    Schneider, G
    Bauer, M
    CONTEMPORARY TOPICS IN LABORATORY ANIMAL SCIENCE, 2004, 43 (06): : 39 - 41
  • [25] Pediatric noninvasive nasal ventilation
    Birnkrant, DJ
    Pope, JF
    Eiben, RM
    JOURNAL OF CHILD NEUROLOGY, 1997, 12 (04) : 231 - 236
  • [26] Using the ROX Index to Predict Treatment Outcome for High-Flow Nasal Cannula and/or Noninvasive Ventilation in Patients With COPD Exacerbations
    Schaeffer, Brett Z.
    Fazio, Sarina A.
    Stocking, Jacqueline C.
    Adams, Jason Y.
    Liu, Anna
    Black, Hugh B.
    Harper, Richart W.
    Cortes-Puch, Irene
    Albertson, Timothy E.
    Kuhn, Brooks
    RESPIRATORY CARE, 2024, 69 (09) : 1100 - 1107
  • [27] Face mask leak with nasal cannula during noninvasive positive pressure ventilation: A randomized crossover trial
    Brown, Derek J.
    Carroll, Stephen M.
    April, Michael D.
    AMERICAN JOURNAL OF EMERGENCY MEDICINE, 2018, 36 (06): : 942 - 948
  • [28] THE ROLE OF NONINVASIVE VENTILATION AND HIGH-FLOW NASAL CANNULA IN CRITICAL COVID-19 PNEUMONIA
    Perez-Gutierrez, Victor
    Ocrospoma, Sebastian
    Bhaskaran, Aditya
    Menon, Vidya
    CRITICAL CARE MEDICINE, 2022, 50 (01) : 92 - 92
  • [29] GENERATION AND EFFECTIVE DELIVERY OF SUBMICROMETER AEROSOLS THROUGH HIGH FLOW NASAL CANNULA DURING NONINVASIVE VENTILATION
    Walenga, R.
    Longest, P. W.
    Son, Y.
    Hindle, M.
    JOURNAL OF AEROSOL MEDICINE AND PULMONARY DRUG DELIVERY, 2013, 26 (02) : A43 - A43
  • [30] Outcomes of Children With Bronchiolitis Treated With High-Flow Nasal Cannula or Noninvasive Positive Pressure Ventilation
    Clayton, Jason A.
    McKee, Bryan
    Slain, Katherine N.
    Rotta, Alexandre T.
    Shein, Steven L.
    PEDIATRIC CRITICAL CARE MEDICINE, 2019, 20 (02) : 128 - 135