Evaluation of a Nasal Cannula in Noninvasive Ventilation Using a Lung Simulator

被引:35
|
作者
Iyer, Narayan P. [1 ]
Chatburn, Robert [2 ,3 ]
机构
[1] Case Western Reserve Univ, Dept Pediat, MetroHlth Med Ctr, Div Neonatol, Cleveland, OH 44106 USA
[2] Cleveland Clin, Resp Inst, Cleveland, OH 44121 USA
[3] Case Western Reserve Univ, Dept Med, Lerner Coll Med, Cleveland, OH 44106 USA
关键词
noninvasive ventilation; bronchopulmonary dysplasia; lung distending pressure; nasal cannula; RAM cannula; POSITIVE AIRWAY PRESSURE; BIRTH-WEIGHT INFANTS; CONTROLLED-TRIAL; EXTUBATION; CPAP;
D O I
10.4187/respcare.03560
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
BACKGROUND: Nasal noninvasive ventilation (NIV) is a common form of noninvasive respiratory mode used in newborn infants. A next-generation nasal cannula (Neotech RAM cannula) has recently been used to provide nasal NIV. The impact of the Neotech RAM cannula on the delivery of pressure needs to be studied. METHODS: In this ex vivo experimental design, a lung simulator (IngMar ASL 5000, version 3.4) was programmed to model a neonate (similar to 1-3 kg of body weight) with normal-to-moderately affected lungs. We used a Covidien PB840 ventilator with NIV software activated to compensate for leaks. Nasal NIV was set at peak airway pressures of 15, 20, and 25 cm H2O and PEEP of 5, 6, and 7 cm H2O. Three sizes of the Neotech RAM cannula were used (prong outer diameters of 3.0, 3.5, and 4.0 mm). The nose was designed to keep the leak of the nares by the prongs to 30%. We also created a worst case leak (58% leak) by using the largest simulated nostril diameter with the smallest diameter Neotech RAM cannula prong. The outcome measure was the difference in pressures, referred to as leak effect, measured by the lung simulator relative to the set peak airway pressure and PEEP on the ventilator. RESULTS: For the interface with 30% leak, leak effects of peak airway pressure during simulated nasal NIV were similar with all Neotech RAM cannula sizes, with 63-75% of peak airway pressure and 70-90% of PEEP being transmitted across the nasal interface. The worst case scenario produced a 92% leak effect in peak airway pressure and PEEP. CONCLUSIONS: When used with <= 30% leak, the Neotech RAM cannula interface results in clinically acceptable transmission of pressures. With > 50% leak, a clinically negligible amount of pressure is transmitted to the artificial lungs.
引用
收藏
页码:508 / 512
页数:5
相关论文
共 50 条
  • [1] Asynchrony During Pediatric Noninvasive Ventilation With a Nasal Cannula Interface: A Lung Model Study
    Studeny, Scott R.
    Chatburn, Robert L.
    Liu, Wei
    Hanna, William J.
    RESPIRATORY CARE, 2021, 66 (07) : 1087 - 1095
  • [2] Outpatient Utilization of the RAM Cannula for Nasal Noninvasive Ventilation in Children
    Truitt, Brittany A.
    Kallam, Erin F.
    Price, Eric W.
    Shah, Amit S.
    Simon, Dawn M.
    Kasi, Ajay S.
    CLINICAL MEDICINE INSIGHTS-PEDIATRICS, 2023, 17
  • [3] High-flow nasal cannula versus noninvasive ventilation on lung re-aeration and diaphragm function
    Schreiber, Annia Fleur
    Lu, Qin
    Adam, Nicolas
    Langeron, Olivier
    EUROPEAN RESPIRATORY JOURNAL, 2017, 50
  • [4] COMPARISON OF PERCENT LEAK USING THE RAM CANNULA IN NONINVASIVE VENTILATION IN AN INFANT LUNG MODEL
    Podgorak, Toni
    Vines, David
    Dapkus, Julianne
    Manaila, Denis
    CRITICAL CARE MEDICINE, 2012, 40 (12) : U127 - U127
  • [5] High-Flow Nasal Cannula Versus Noninvasive Ventilation: A Matter of Confusion
    Modesto i Alapont, Vicent
    Pons-Odena, Marti
    Medina, Alberto
    PEDIATRIC CRITICAL CARE MEDICINE, 2019, 20 (12) : 1210 - 1211
  • [6] High-flow nasal cannula is superior to noninvasive ventilation to prevent reintubation?
    Nishimura, Masaji
    ANNALS OF TRANSLATIONAL MEDICINE, 2017, 5 (05)
  • [7] Aerosol Therapy During Noninvasive Ventilation or High-Flow Nasal Cannula
    Hess, Dean R.
    RESPIRATORY CARE, 2015, 60 (06) : 880 - 891
  • [8] Weaning from noninvasive ventilation and high flow nasal cannula in bronchiolitis: A survey of practice
    Suzanne, Marie
    Amaddeo, Alessandro
    Pin, Isabelle
    Milesi, Christophe
    Mortamet, Guillaume
    PEDIATRIC PULMONOLOGY, 2020, 55 (11) : 3104 - 3109
  • [9] High-flow nasal cannula and noninvasive ventilation: effects on alveolar recruitment and overdistension
    Saini, Amandeep S.
    Meredith, Simon
    Esquinas, Antonio M.
    Mina, Bushra A.
    ERJ OPEN RESEARCH, 2022, 8 (02)
  • [10] Nasal cannula during noninvasive positive pressure ventilation: Combining preoxygenation with apneic oxygen
    Brown, Derek J.
    Carroll, Stephen M.
    April, Michael D.
    AMERICAN JOURNAL OF EMERGENCY MEDICINE, 2018, 36 (05): : 878 - 879