Intracranial aneurysms treated with Guglielmi detachable coils: Is contrast material necessary in the follow-up with 3D time-of-flight MR angiography?

被引:0
|
作者
Cottier, JP [1 ]
Bleuzen-Couthon, A [1 ]
Gallas, S [1 ]
Vinikoff-Sonier, CB [1 ]
Bertrand, P [1 ]
Domengie, F [1 ]
Barantin, L [1 ]
Herbreteau, D [1 ]
机构
[1] Hop Bretonneau, Dept Neuroradiol, F-37044 Tours 1, France
关键词
D O I
暂无
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
BACKGROUND AND PURPOSE: Three-dimensional time-of-flight (TOF) MR angiography has been evaluated in the follow-up of intracranial aneurysms treated with Guglielmi detachable coils (GDCs) with good results. Some of the studies used contrast material in addition to the 3D TOF MR technique and others did not. We assessed the usefulness of contrast material with 3D TOF MR angiography by comparing this sequence before and after contrast material injection. METHODS: Fifty-eight patients harboring a total of 71 cerebral aneurysms previously treated with GDCs were included in the prospective study. MR angiography (at 1.5 T) was performed with a 3D TOF sequence before and after injection of gadolinium-based contrast material. Features evaluated were presence and size of a neck remnant, parent and adjacent vessel patency, and venous overlap. Digital subtraction angiography was the standard of reference. RESULTS: Comparison of the techniques showed a good agreement in the detection of residual flow. Six cases of small residual neck were not detected with either the 3D TOF or the contrast-enhanced 3D TOF sequence. In one case of giant aneurysm, the extent of recanalization was more evident after contrast material administration. The use of contrast material did not help to show the parent and adjacent arteries. Venous overlap on contrast-enhanced 3D TOF angiograms did not affect image interpretation. CONCLUSION. In this series, the use of intravenous contrast material did not improve the ability of 3D TOF MR angiography to depict the presence of residual or recurrent aneurysms previously treated with endovascular coiling. In one giant aneurysm, use of intravenous contrast material did result in improved visualization of a residual aneurysm.
引用
收藏
页码:1797 / 1803
页数:7
相关论文
共 50 条
  • [21] Three-dimensional time-of-flight MR angiography for evaluation of intracranial aneurysms after endosaccular packing with Guglielmi detachable coils: comparison with 3D digital subtraction angiography
    Okahara, M
    Kiyosue, H
    Hori, Y
    Yamashita, M
    Nagatomi, H
    Mori, H
    [J]. EUROPEAN RADIOLOGY, 2004, 14 (07) : 1162 - 1168
  • [22] Three-dimensional time-of-flight MR angiography for evaluation of intracranial aneurysms after endosaccular packing with Guglielmi detachable coils: comparison with 3D digital subtraction angiography
    Mika Okahara
    Hiro Kiyosue
    Yuzo Hori
    Masanori Yamashita
    Hirofumi Nagatomi
    Hiromu Mori
    [J]. European Radiology, 2004, 14 : 1162 - 1168
  • [23] MR angiography in the follow-up of coiled cerebral aneurysms after treatment with Guglielmi detachable coils
    Nome, T
    Bakke, SJ
    Nakstad, PH
    [J]. ACTA RADIOLOGICA, 2002, 43 (01) : 10 - 14
  • [24] Time-of-flight MR angiography at 3T versus digital subtraction angiography in the imaging follow-up of 51 intracranial aneurysms treated with coils
    Ferre, Jean-Christophe
    Carsin-Nicol, Beatrice
    Morandi, Xavier
    Carsin, Michel
    de Kersaint-Gilly, Axel
    Gauvrit, Jean-Yves
    Desal, Hubert-Armand
    [J]. EUROPEAN JOURNAL OF RADIOLOGY, 2009, 72 (03) : 365 - 369
  • [25] Intracranial aneurysms treated with Guglielmi detachable coils: long-term imaging follow-up with contrast-enhanced magnetic resonance angiography
    Gauvrit, Jean-Yves
    Caron, Sabine
    Taschner, Christian A.
    Lejeune, Jean-Paul
    Pruvo, Jean-Pierre
    Leclerc, Xavier
    [J]. JOURNAL OF NEUROSURGERY, 2008, 108 (03) : 443 - 449
  • [26] MR angiography at 3T versus digital subtraction angiography in the follow-up of intracranial aneurysms treated with detachable coils
    Majoie, CBLM
    Sprengers, ME
    van Rooij, WJJ
    Lavini, C
    Sluzewski, M
    van Rijn, JC
    den Heeten, GJ
    [J]. AMERICAN JOURNAL OF NEURORADIOLOGY, 2005, 26 (06) : 1349 - 1356
  • [27] Detection capability of the necessity for reembolization of three types of MR angiography in the follow-up of intracranial aneurysms embolized with Guglielmi detachable coils
    Ueda, F
    Matsui, O
    Suzuki, M
    Sanada, J
    Terayama, N
    [J]. RADIOLOGY, 2000, 217 : 179 - 179
  • [28] MR angiography with ultrashort echo time in cerebral aneurysms treated with Guglielmi detachable coils
    Gonner, F
    Heid, O
    Remonda, L
    Nicoli, G
    Baumgartner, RW
    Godoy, N
    Schroth, G
    [J]. AMERICAN JOURNAL OF NEURORADIOLOGY, 1998, 19 (07) : 1324 - 1328
  • [29] CONTRAST MATERIAL-ENHANCED MR ANGIOGRAPHY FOR INTRACRANIAL ANEURYSMS WITH 3D TIME-OF-FLIGHT AND 3D PHASE-CONTRAST TECHNIQUES
    CHUNG, WC
    LISTERUD, J
    ATLAS, SW
    [J]. RADIOLOGY, 1992, 185 : 121 - 121
  • [30] Follow-up of coiled cerebral aneurysms at 3T: Comparison of 3D time-of-flight MR angiography and contrast-enhanced MR angiography
    Anzalone, N.
    Scomazzoni, F.
    Cirillo, M.
    Righi, C.
    Simionato, F.
    Cadioli, M.
    Iadanza, A.
    Kirchin, M. A.
    Scotti, G.
    [J]. AMERICAN JOURNAL OF NEURORADIOLOGY, 2008, 29 (08) : 1530 - 1536