3.5 cm Artificial Urinary Sphincter Cuff Erosion Occurs Predominantly in Irradiated Patients

被引:65
|
作者
Simhan, Jay [1 ]
Morey, Allen F. [1 ]
Singla, Nirmish [1 ]
Tausch, Timothy J. [1 ]
Scott, J. Francis [1 ]
Lemack, Gary E. [1 ]
Roehrborn, Claus G. [1 ]
机构
[1] Univ Texas SW Med Ctr Dallas, Dept Urol, Dallas, TX 75390 USA
来源
JOURNAL OF UROLOGY | 2015年 / 193卷 / 02期
关键词
urinary bladder; urethra; urinary incontinence; stress; urinary sphincter; artificial; complications; POSTPROSTATECTOMY INCONTINENCE; REVISION SURGERY; IMPLANTATION; OUTCOMES; RISK; MEN; EXPLANTATION; PLACEMENT; RADIATION; FAILURES;
D O I
10.1016/j.juro.2014.07.115
中图分类号
R5 [内科学]; R69 [泌尿科学(泌尿生殖系疾病)];
学科分类号
1002 ; 100201 ;
摘要
Purpose: We analyzed our initial 100-case experience with the 3.5 cm artificial urinary sphincter cuff to identify risk factors for cuff erosion. Materials and Methods: We reviewed the records of a single surgeon, consecutive series of patients treated with 3.5 cm artificial urinary sphincter cuff placement from September 2009 to August 2013. Each patient underwent single perineal cuff placement via standardized technique. Preoperative characteristics, technical considerations and postoperative outcomes were analyzed and compared to those in a cohort of patients in whom a larger (4.0 cm or greater) artificial urinary sphincter cuff was placed during the same period. We identified clinical factors associated with an increased risk of 3.5 cm artificial urinary sphincter cuff erosion. Results: Of the 176 men who met study inclusion criteria during the 4-year period 100 (57%) received the 3.5 cm artificial urinary sphincter cuff and 76 (43%) received a larger cuff (4.0 cm or greater). The continence rate (83% vs 80%, p = 0.65) and mean followup (32 vs 25 months, p = 0.14) were similar in the 2 groups. Erosion developed in 16 of the 176 patients (9%) during the study period, of whom 13 had the 3.5 cm cuff. Of the 100 patients with the 3.5 cm cuff 52 (52%) had a history of radiation, including 11 (21%) with erosion. Cuff erosion developed only rarely in nonirradiated men (2 of 48 or 4%, p = 0.01). A history of radiation was the only significant risk factor associated with 3.5 cm cuff erosion (OR 6.2, 95% CI 1.3-29.5). Conclusions: Men with a history of radiation who underwent placement of a 3.5 cm artificial urinary sphincter cuff experienced an increased (21%) risk of cuff erosion.
引用
收藏
页码:593 / 597
页数:5
相关论文
共 50 条
  • [1] INCREASED RATES OF FAILURE WITH A 3.5CM CUFF IN THE ARTIFICIAL URINARY SPHINCTER
    Zaid, Uwais
    Lavien, Garjae
    Peterson, Andrew
    JOURNAL OF UROLOGY, 2016, 195 (04): : E1182 - E1182
  • [2] PRESENTING SYMPTOMS IN PATIENTS WITH ARTIFICIAL URINARY SPHINCTER CUFF EROSION
    Enemchukwu, Ekene
    Kaufman, Melissa
    Milam, Douglas
    JOURNAL OF SEXUAL MEDICINE, 2012, 9 : 60 - 60
  • [3] PRESENTING SYMPTOMS IN PATIENTS WITH ARTIFICIAL URINARY SPHINCTER CUFF EROSION
    Enemchukwu, Ekene
    Kaufman, Melissa
    Milam, Doug
    NEUROUROLOGY AND URODYNAMICS, 2012, 31 (02) : 258 - 258
  • [4] THE NATURAL HISTORY OF PATIENTS WITH ARTIFICIAL URINARY SPHINCTER CUFF EROSION
    Lucas, J.
    Patel, A.
    Loh-Doyle, J.
    Nikolavsky, D.
    Angulo, J.
    Martins, F.
    Simhan, J.
    JOURNAL OF SEXUAL MEDICINE, 2019, 16 (04): : S51 - S51
  • [5] EROSION RATES OF 3.5 CM ARTIFICIAL URINARY SPHINCTER CUFFS ARE SIMILAR TO LARGER CUFFS
    Shakir, Nabeel
    McKibben, Maxim
    Fuchs, Joceline
    Scott, Jeremy
    Morey, Allen
    JOURNAL OF UROLOGY, 2018, 199 (04): : E798 - E798
  • [6] Management of cuff erosion of the double cuff artificial urinary sphincter
    Bell, BB
    Mulcahy, JJ
    JOURNAL OF UROLOGY, 2000, 163 (01): : 85 - 86
  • [7] ARTIFICIAL URINARY SPHINCTER CUFF EROSION - EXPERIENCE WITH REIMPLANTATION IN 38 PATIENTS
    MOTLEY, RC
    BARRETT, DM
    UROLOGY, 1990, 35 (03) : 215 - 218
  • [8] Impact of 3.5 cm Artificial Urinary Sphincter Cuff on Primary and Revision Surgery for Male Stress Urinary Incontinence
    Hudak, Steven J.
    Morey, Allen F.
    JOURNAL OF UROLOGY, 2011, 186 (05): : 1962 - 1966
  • [9] IMPACT OF THE 3.5 CM ARTIFICIAL URINARY SPHINCTER CUFF IN PRIMARY AND REVISION SURGERY FOR MALE STRESS URINARY INCONTINENCE
    Hudak, Steven
    Valadez, Celeste
    Terlecki, Ryan
    Morey, Allen
    JOURNAL OF UROLOGY, 2011, 185 (04): : E409 - E409
  • [10] Erosion rates of 3.5-cm artificial urinary sphincter cuffs are similar to larger cuffs
    McKibben, Maxim J.
    Shakir, Nabeel
    Fuchs, Joceline S.
    Scott, Jeremy M.
    Morey, Allen F.
    BJU INTERNATIONAL, 2019, 123 (02) : 335 - 341