Adaptation of a new pneumatic nebulizer for sample introduction in ICP spectrometry

被引:32
|
作者
Groom, S [1 ]
Schaldach, G
Ulmer, M
Walzel, P
Berndt, H
机构
[1] Univ Dortmund, Dept Mech Proc Engn, D-44227 Dortmund, Germany
[2] ISAS, Inst Analyt Sci, D-44139 Dortmund, Germany
关键词
D O I
10.1039/b410772c
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A new pneumatic nebulizer, the pneumatic extension nozzle ( PEN), originally developed in technical laboratories, mainly for the production of rock and glass wool, is applied for sample introduction in ICP-OES. The droplet formation process under various operating conditions and geometries was investigated using a transparent enlarged model of the PEN. For the application in ICP spectrometry the geometry of the PEN was optimized and miniaturized with the aid of similarity theory. In comparison with a standard concentric nebulizer the miniaturized PEN generates an aerosol with a twice higher mass fraction of droplets with diameters D < 10 mu m. An empirical model for the prediction of the mean droplet diameter is presented, using the nozzle diameter as a linear size scale. This model enables a better fit to experimental data compared with existing models. Applying the miniaturised PEN to a simultaneous ICP-OES instrument lowers the detection limits up to a factor of 3.5 depending on the element. The new pneumatic extension nozzle can be easily adapted to existing ICP-OES instruments.
引用
收藏
页码:169 / 175
页数:7
相关论文
共 50 条