COMPARISON OF WIRELESS DETECTORS FOR DIGITAL RADIOGRAPHY SYSTEMS: IMAGE QUALITY AND DOSE

被引:3
|
作者
Mourik, J. E. M. [1 ,2 ]
van der Tol, P. [2 ]
Veldkamp, W. J. H. [2 ]
Geleijns, J. [2 ]
机构
[1] Sint Franciscus Vlietland Grp, Dept Radiol & Nucl Med, Kleiweg 500, NL-3045 PM Rotterdam, Netherlands
[2] Leiden Univ, Med Ctr, Dept Radiol, Albinusdreef 2, NL-2333 ZA Leiden, Netherlands
关键词
D O I
10.1093/rpd/ncv450
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
The purpose of this study was to compare dose and image quality of wireless detectors for digital chest radiography. Entrance dose at both the detector (EDD) and phantom (EPD) and image quality were measured for wireless detectors of seven different vendors. Both the local clinical protocols and a reference protocol were evaluated. In addition, effective dose was calculated. Main differences in clinical protocols involved tube voltage, tube current, the use of a small or large focus and the use of additional filtration. For the clinical protocols, large differences in EDD (1.4-11.8 mu Gy), EPD (13.9-80.2 mu Gy) and image quality (IQF(inv): 1.4-4.1) were observed. Effective dose was <0.04 mSv for all protocols. Large differences in performance were observed between the seven different systems. Although effective dose is low, further improvement of imaging technology and acquisition protocols is warranted for optimisation of digital chest radiography.
引用
下载
收藏
页码:303 / 307
页数:5
相关论文
共 50 条
  • [31] Computed radiography systems for mammography: an evaluation of image quality and dose
    CP Lawinski
    DA Goodman
    MA Whall
    Breast Cancer Research, 4 (Suppl 1)
  • [32] Management of dose and quality in digital radiography
    Busch, HP
    ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 2003, 175 (01): : 17 - 19
  • [33] Amorphous silicon technology based digital radiography of the skeleton: Image quality and potential for dose reduction in comparison with screen-film radiography
    Volk, M
    Strotzer, M
    Holzknecht, N
    Manke, C
    Lenhart, M
    Gmeinwieser, JK
    RADIOLOGY, 1999, 213P : 328 - 328
  • [34] Supine and erect abdominal radiography: A comparison of radiation dose and image quality
    Alzyoud, Kholoud
    Al-Murshedi, Sadeq
    Abualkhael, Khaleda
    Alqawasmeh, Farah
    England, Andrew
    APPLIED RADIATION AND ISOTOPES, 2022, 190
  • [35] Balancing Image Quality and Dose in Radiography
    Peck, D.
    Pasciak, A.
    MEDICAL PHYSICS, 2014, 41 (06) : 480 - 480
  • [36] Image Acquisition and Quality in Digital Radiography
    Alexander, Shannon
    RADIOLOGIC TECHNOLOGY, 2016, 88 (01) : 53 - 66
  • [37] Image Quality in Digital Industrial Radiography
    Ewert, Uwe
    Zscherpel, Uwe
    Heyne, Klara
    Jechow, Mirko
    Bavendiek, Klaus
    MATERIALS EVALUATION, 2012, 70 (08) : 955 - 964
  • [38] Image Quality Optimization vs. Patient Dose in Digital Thoracic Radiography
    Perez-Diaz, M.
    Ruiz-Gonzalez, Y.
    Machin-Linares, L.
    Ely-Andrade, M.
    Barros-Saito, J. C.
    Khoury, H. J.
    Ventosa-Almeida, J. C.
    VI LATIN AMERICAN CONGRESS ON BIOMEDICAL ENGINEERING (CLAIB 2014), 2014, 49 : 353 - 356
  • [39] IMAGE QUALITY OPTIMIZATION IN BREAST DIGITAL RADIOGRAPHY VS. PATIENT DOSE
    Bernardo, A.
    Perez-Diaz, M.
    Echevarria-Torres, W. J.
    Diaz-Amador, R.
    REVISTA CUBANA DE FISICA, 2018, 35 (1E): : E17 - E20
  • [40] The effect of added filtration on radiation dose and image quality in digital radiography of newborns
    Antonios E. Papadakis
    Vasiliki Giannakaki
    Eleftheria Hatzidaki
    John Damilakis
    Pediatric Radiology, 2023, 53 : 2060 - 2068