Determination of Cu and Cd content of groundwater colloids by solid sampling graphite furnace atomic absorption spectrometry

被引:19
|
作者
Freedman, YE
Ronen, D
Long, GL
机构
[1] WEIZMANN INST SCI,DEPT ENVIRONM SCI & ENERGY RES,IL-76100 REHOVOT,ISRAEL
[2] VIRGINIA POLYTECH INST & STATE UNIV,DEPT CHEM,BLACKSBURG,VA 24061
关键词
D O I
10.1021/es950637w
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
Cu and Cd content of standard reference materials (SRMs), minerals (quartz, calcite, kaolinite, and bentonite), and samples of aquifer colloids, obtained under natural gradient flow conditions, were analyzed by solid sampling graphite furnace atomic absorption spectrometry (GFAAS). Pyrolytically coated tubes, designed for the analysis of liquid samples, were successfully used. Recovery of analytes by solid sampling was 106-112% of the certified values of SRMs. Cu concentration in clay minerals, obtained by solid sampling, was about 9-fold (in kaolinite) and 3-fold (in bentonite) higher than that obtained by nitric acid extraction, indicating that Cu is partially incorporated into the crystal lattice of the minerals. Analysis of aquifer colloids shows that within a 6-month period the average content of Cu and Cd decreased by about 87%. Solid sampling GFAAS was demonstrated to be a convenient method for the study of metal content in small (5 mg) samples of groundwater colloids.
引用
收藏
页码:2270 / 2277
页数:8
相关论文
共 50 条
  • [21] Solid sampling-graphite furnace atomic absorption spectrometry for the direct determination of silver at trace and ultratrace levels
    Resano, M
    Aramendía, M
    García-Ruiz, E
    Crespo, C
    Belarra, MA
    ANALYTICA CHIMICA ACTA, 2006, 571 (01) : 142 - 149
  • [22] Screening of antimony in PVC by solid sampling-graphite furnace atomic absorption spectrometry
    Belarra, MA
    Belategui, I
    Lavilla, I
    Anzano, JM
    Castillo, JR
    TALANTA, 1998, 46 (06) : 1265 - 1272
  • [23] Determination of cadmium, copper and lead in mineral coal using solid sampling graphite furnace atomic absorption spectrometry
    Vale, MGR
    Silva, MM
    Welz, B
    Lima, ÉC
    SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 2001, 56 (10) : 1859 - 1873
  • [24] DETERMINATION OF COPPER IN BIOLOGICAL MICROSAMPLES BY DIRECT SOLID SAMPLING GRAPHITE-FURNACE ATOMIC-ABSORPTION SPECTROMETRY
    EBDON, L
    EVANS, EH
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1987, 2 (03) : 317 - 320
  • [25] Direct analysis of silica by means of solid sampling graphite furnace atomic absorption spectrometry
    Resano, M.
    Mozas, E.
    Crespo, C.
    Perez, J.
    Garcia-Ruiz, E.
    Belarra, M. A.
    SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 2012, 71-72 : 24 - 30
  • [26] Method for cadmium and lead longitudinal profiles determination in hair by solid sampling graphite furnace atomic absorption spectrometry
    Nomura, Cassiana Seimi
    Oliveira, Pedro Vitoriano
    ANALYTICAL METHODS, 2010, 2 (01) : 49 - 53
  • [27] Solid sampling-graphite furnace atomic absorption spectrometry for the direct determination of Au in samples of various natures
    Resano, M
    Aramendía, M
    Garcia-Ruiz, E
    Belarra, MA
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2005, 20 (05) : 479 - 481
  • [28] High purity polyimide analysis by solid sampling graphite furnace atomic absorption spectrometry
    Santos, Rafael F.
    Carvalho, Gabriel S.
    Duarte, Fabio A.
    Bolzan, Rodrigo C.
    Flores, Erico M. M.
    SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 2017, 129 : 42 - 48
  • [29] Antimony determination in marine sediments by graphite furnace atomic absorption spectrometry with slurry sampling
    Barciela-Alonso, M. C.
    Varela-Vilela, R.
    Pena-Vazquez, E.
    Bermejo-Barrera, A.
    Bermejo-Barrera, P.
    ATOMIC SPECTROSCOPY, 2007, 28 (03) : 102 - 107
  • [30] Slurry sampling for the determination of thallium in soils and sediments by graphite furnace atomic absorption spectrometry
    Dobrowolski, R.
    ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2002, 374 (7-8) : 1294 - 1300