TOF-SIMS with Argon Gas Cluster Ion Beams: A Comparison with C60+

被引:170
|
作者
Rabbani, Sadia [1 ]
Barber, Andrew M. [2 ]
Fletcher, John S. [1 ]
Lockyer, Nicholas P. [1 ]
Vickerman, John C. [1 ]
机构
[1] Univ Manchester, Sch Chem Engn & Analyt Sci, Manchester Interdisciplinary Bioctr, Manchester M13 9PL, Lancs, England
[2] Ionoptika Ltd, Southampton, Hants, England
基金
英国工程与自然科学研究理事会;
关键词
MASS-SPECTROMETRY; THIN-FILMS; SURFACES; YIELD; BOMBARDMENT; EMISSION; LAYERS; MODEL;
D O I
10.1021/ac200288v
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Time-of-flight secondary ion mass spectrometry (TOF-SIMS) is an established technique for the characterization of solid sample surfaces. The introduction of polyatomic ion beams, such as C-60, has provided the associated ability to perform molecular depth-profiling and 3D molecular imaging. However, not all samples perform equally under C-60 bombardment, and it is probably naive to think that there will be an ion beam that will be applicable in all situations. It is therefore important to explore the potential of other candidates. A systematic study of the suitability of argon gas cluster ion beams (Ar-GCIBs) of general composition Ar-n(+), where n = 60-3000, as primary particles in TOP-SIMS analysis has been performed. We have assessed the potential of the Ar-GCIBs for molecular depth-profiling in terms of damage accumulation and sputter rate and also as analysis beams where spectral quality and secondary ion yields are considered. We present results with direct comparison with C-60 ions on the same sample in the same instrument on polymer, polymer additive, and biomolecular samples, including lipids and small peptides. Large argon clusters show reduced damage accumulation compared with C-60 with an approximately constant sputter rate as a function of Ar cluster size. Further, on some samples, large argon clusters produce changes in the mass spectra indicative of a more gentle ejection mechanism. However, there also appears to be a reduction in the ionization of secondary species as the size of the Ar cluster increases.
引用
收藏
页码:3793 / 3800
页数:8
相关论文
共 50 条
  • [1] Development of an Argon Gas Cluster Ion Beam for ToF-SIMS Analysis
    Lee, Sang Ju
    Choi, Chang Min
    Min, Boo Ki
    Baek, Ji Young
    Eo, Jae Yeong
    Choi, Myoung Choul
    BULLETIN OF THE KOREAN CHEMICAL SOCIETY, 2019, 40 (09): : 877 - 881
  • [2] ToF-SIMS imaging with cluster ion beams
    Xu, J
    Ostrowski, S
    Szakal, C
    Ewing, AG
    Winograd, N
    APPLIED SURFACE SCIENCE, 2004, 231 : 159 - 163
  • [3] Investigation of molecular weight effects of polystyrene in ToF-SIMS using C60+ and Au+ primary ion beams
    Piwowar, Alan M.
    Lockyer, Nicholas
    Vickerman, John C.
    APPLIED SURFACE SCIENCE, 2008, 255 (04) : 912 - 915
  • [4] Improved mass resolution and mass accuracy in TOF-SIMS spectra and images using argon gas cluster ion beams
    Shon, Hyun Kyong
    Yoon, Sohee
    Moon, Jeong Hee
    Lee, Tae Geol
    BIOINTERPHASES, 2016, 11 (02)
  • [5] ToF-SIMS analysis of diadenosine triphosphate and didadenosine tetraphosphate using bismuth and argon cluster ion beams
    Shon, Hyun Kyong
    Cho, Young-Lai
    Lim, Choung Su
    Choi, Joon Sig
    Chung, Sang J.
    Lee, Tae Geol
    SURFACE AND INTERFACE ANALYSIS, 2014, 46 : 189 - 192
  • [6] ToF-SIMS analysis of ion implanted standard to quantify insecticide in mosquito netting with cesium and argon gas cluster sputtering beams
    Zhou, Chuanzhen
    Stevie, Fred A.
    Smith, Stephen C.
    Rading, Derk
    Zakel, Julia
    JOURNAL OF VACUUM SCIENCE & TECHNOLOGY B, 2018, 36 (03):
  • [7] Comparison of MeV monomer ion and keV cluster ToF-SIMS
    Jones, Brian N.
    Matsuo, Jiro
    Nakata, Yoshihiko
    Yamada, Hideaki
    Watts, John
    Hinder, Steven
    Palitsin, Vladimir
    Webb, Roger
    SURFACE AND INTERFACE ANALYSIS, 2011, 43 (1-2) : 249 - 252
  • [8] ToF-SIMS of OLED materials using argon gas cluster ion Beam: A promising approach for OLED inspection
    Baek, Ji Young
    Choi, Chang Min
    Lee, Sang Ju
    Min, Boo Ki
    Kang, Hwa Seung
    Choo, Dong Chul
    Sung, Ji Yeong
    Jin, Jong Sung
    Choi, Myoung Choul
    APPLIED SURFACE SCIENCE, 2020, 507
  • [9] Comparison of Bi1+, Bi3+ and C60+ primary ion sources for ToF-SIMS imaging of patterned protein samples
    Dubey, Manish
    Brison, Jeremy
    Grainger, David W.
    Castner, David G.
    SURFACE AND INTERFACE ANALYSIS, 2011, 43 (1-2) : 261 - 264
  • [10] Incident Angle Dependence in Polymer TOF-SIMS Depth Profiling with C-60 Ion Beams
    Iida, Shin-ichi
    Miyayama, Takuya
    Sanada, Noriaki
    Suzuki, Mineharu
    Fisher, Gregory L.
    Bryan, Scott R.
    E-JOURNAL OF SURFACE SCIENCE AND NANOTECHNOLOGY, 2009, 7 : 878 - 881