Initial clinical results of an in vivo dosimeter during external beam radiation therapy

被引:38
|
作者
Scarantino, CW
Rini, CJ
Aquino, M
Carrea, TB
Ornitz, RD
Anscher, MS
Black, RD
机构
[1] Sicel Technol Inc, Morrisville, NC 27560 USA
[2] Rex Canc Ctr, Raleigh, NC USA
[3] Duke Univ, Med Ctr, Durham, NC USA
关键词
implantable dosimeter; radiation therapy; dose quality assurance;
D O I
10.1016/j.ijrobp.2004.09.041
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: An implantable radiation dosimeter has been developed to monitor dose delivered at depth in patients undergoing external beam therapy. A clinical pilot study was conducted to test the safety, efficacy, and utility of the device. Methods and Materials: Ten patients, all with unresectable malignant disease, were enrolled to assess implantation risk and movement of the device in the body and to compare the in vivo measured dose to the value predicted by the treatment planning system software. Results: Migration of the sensor away from the point of original placement was noted in only 1 patient (due to unconsolidated host tissue) and no adverse events were recorded during the implantation procedure or thereafter. Daily dose measurements were recorded successfully for all sensors in all patients. Variance between measured and predicted dose values was reported as a frequency of error at the >= 5% and >= 8% levels. The error frequency at the >= 8% level was as high as 47%, 29%, and 21% for lung, prostate, and rectal tumors, respectively. Conclusions: The implantable dosimeter was found to be safe and effective in measuring dose at depth. There are many factors that can influence delivered dose, and the implantable dosimeter measures the net effect of these factors. The daily sensor readings provide a new tool for rigorous treatment quality assurance. (c) 2005 Elsevier Inc.
引用
收藏
页码:606 / 613
页数:8
相关论文
共 50 条
  • [21] Radiation Oncology-External-Beam Radiation Therapy
    Martin, Melissa C.
    [J]. HEALTH PHYSICS, 2019, 116 (02): : 184 - 188
  • [22] CHEMOTHERAPY AND EXTERNAL BEAM RADIATION AS INITIAL TREATMENT FOR RETINOBLASTOMA
    Arguis, M.
    Nieves, M.
    Sabado, C.
    Ramos, M.
    Gallego, S.
    Valdes, M.
    Sanchez de Toledo, J.
    Giralt, J.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2008, 88 : S83 - S83
  • [23] Clinical significance of increased gelatinolytic activity in the rectal mucosa during external beam radiation therapy of prostate cancer
    Hovdenak, N
    Wang, JR
    Sung, CC
    Kelly, T
    Fajardo, LF
    Hauer-Jensen, M
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2002, 53 (04): : 919 - 927
  • [24] Anesthesia for external beam radiation therapy in children n an audit of clinical practice
    Sarkar, S.
    Kumari, A.
    Mallick, I.
    Chatterjee, S.
    Achari, R.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2014, 111 : S43 - S44
  • [25] Systematic Reduction of Table Position Overrides and Radiation Incidents During External Beam Radiation Therapy
    Yu, N.
    Murray, E.
    Meier, T.
    Barrett, P.
    Suh, J.
    Chao, S.
    Xia, P.
    [J]. MEDICAL PHYSICS, 2017, 44 (06) : 3275 - 3275
  • [26] Feasibility study of radiophotoluminescent glass dosimeter for in vivo dosimetry in external photon beam radiotherapy
    Nazia Neelam Shehzadi
    Chul-Young Yi
    Ki-Hwan Kim
    Jaeuk Jang
    Ui-Jung Hwang
    Yunho Kim
    In Jung Kim
    Young Min Seong
    [J]. Journal of the Korean Physical Society, 2021, 78 : 523 - 534
  • [27] Feasibility study of radiophotoluminescent glass dosimeter for in vivo dosimetry in external photon beam radiotherapy
    Shehzadi, Nazia Neelam
    Yi, Chul-Young
    Kim, Ki-Hwan
    Jang, Jaeuk
    Hwang, Ui-Jung
    Kim, Yunho
    Kim, In Jung
    Seong, Young Min
    [J]. JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 2021, 78 (06) : 523 - 534
  • [28] A Dosimeter for In Vivo Real Time Verification of Intensity Modulated Radiation Therapy
    Price, S.
    Silvius, A. A.
    Izaguirre, E. W.
    [J]. MEDICAL PHYSICS, 2011, 38 (06)
  • [29] External-beam radiation therapy alone
    Wong, CS
    Brierley, JD
    [J]. SEMINARS IN RADIATION ONCOLOGY, 1998, 8 (01) : 3 - 12
  • [30] External beam radiation therapy for tracheobronchial amyloidosis
    Neben-Wittich, Michelle A.
    Foote, Robert L.
    Kalra, Sanjay
    [J]. CHEST, 2007, 132 (01) : 262 - 267