A Realistic Breast Phantom Proposal for 3D Image Reconstruction in Digital Breast Tomosynthesis

被引:1
|
作者
Polat, Adem [1 ]
Kumrular, Raziye Kubra [2 ]
机构
[1] Canakkale Onsekiz Mart Univ, Dept Elect Elect Engn, TR-17100 Canakkale, Turkey
[2] Univ Southampton, Inst Sound & Vibrat Res, Southampton, Hants, England
关键词
DBT; digital breast tomosynthesis; SART; FBP; breast phantom; breast cancer; breast imaging; PHASE-CONTRAST TOMOGRAPHY; ITERATIVE RECONSTRUCTION; SOFTWARE PACKAGE; ALGORITHM; FRAMEWORK; TOOLBOX; OBJECTS;
D O I
10.1177/15330338221104567
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Objectives: Iterative (eg, simultaneous algebraic reconstruction technique [SART]) and analytical (eg, filtered back projection [FBP]) image reconstruction techniques have been suggested to provide adequate three-dimensional (3D) images of the breast for capturing microcalcifications in digital breast tomosynthesis (DBT). To decide on the reconstruction method in clinical DBT, it must first be tested in a simulation resembling the real clinical environment. The purpose of this study is to introduce a 3D realistic breast phantom for determining the reconstruction method in clinical applications. Methods: We designed a 3D realistic breast phantom with varying dimensions (64(3)-512(3)) mimicking some structures of a real breast such as milk ducts, lobules, and ribs using TomoPhantom software. We generated microcalcifications, which mimic cancerous cells, with a separate MATLAB code and embedded them into the phantom for testing and benchmark studies in DBT. To validate the characterization of the phantom, we tested the distinguishability of microcalcifications by performing 3D image reconstruction methods (SART and FBP) using Laboratory of Computer Vision (LAVI) open-source reconstruction toolbox. Results: The creation times of the proposed realistic breast phantom were seconds of 2.5916, 8.4626, 57.6858, and 472.1734 for 64(3), 128(3), 256(3), and 512(3), respectively. We presented reconstructed images and quantitative results of the phantom for SART (1-2-4-8 iterations) and FBP, with 11 to 23 projections. We determined qualitatively and quantitatively that SART (2-4 iter.) yields better results than FBP. For example, for 23 projections, the contrast-to-noise ratio (CNR) values of SART (2 iter.) and FBP were 2.871 and 0.497, respectively. Conclusions: We created a computationally efficient realistic breast phantom that is eligible for reconstruction and includes anatomical structures and microcalcifications, successfully. By proposing this breast phantom, we provided the opportunity to test which reconstruction methods can be used in clinical applications vary according to various parameters such as the No. of iterations and projections in DBT.
引用
收藏
页数:9
相关论文
共 50 条
  • [21] Breast microcalcification classification: From 2D mammography to 3D digital breast tomosynthesis
    Yeh, Jinn-Yi
    Chan, Siwa
    ASIA-PACIFIC JOURNAL OF CLINICAL ONCOLOGY, 2022, 18 : 43 - 43
  • [22] 3D digital breast tomosynthesis using total variation regularization
    Kastanis, Iason
    Arridge, Simon
    Stewart, Alex
    Gunn, Spencer
    Ullberg, Christer
    Francke, Tom
    DIGITAL MAMMOGRAPHY, PROCEEDINGS, 2008, 5116 : 621 - +
  • [23] ITERATIVE BREAST TOMOSYNTHESIS IMAGE RECONSTRUCTION
    Bustamante, Veronica Mejia
    Nagy, James G.
    Feng, Steve S. J.
    Sechopoulos, Ioannis
    SIAM JOURNAL ON SCIENTIFIC COMPUTING, 2013, 35 (05): : S192 - S208
  • [24] Breast percent density estimation from 3D reconstructed digital breast tomosynthesis images
    Bakic, Predrag R.
    Kontos, Despina
    Carton, Ann-Katherine
    Maidment, Andrew D. A.
    MEDICAL IMAGING 2008: PHYSICS OF MEDICAL IMAGING, PTS 1-3, 2008, 6913
  • [25] Image Quality Phantom in Close Compliance with the ICRU-44 Breast Tissue Substitute Standard and Realistic Background Noise Pattern for Digital Breast Tomosynthesis
    Raeymaeckers, Steven
    Buls, Nico
    Van Gompel, Gert
    Breucq, Catherine
    de Mey, Johan
    DIGITAL MAMMOGRAPHY, 2010, 6136 : 556 - 560
  • [26] Enhancing tissue structures with iterative image reconstruction for digital breast tomosynthesis
    Sidky, Emil Y.
    Reiser, Ingrid S.
    Nishikawa, Robert M.
    Pan, Xiaochuan
    MEDICAL IMAGING 2014: PHYSICS OF MEDICAL IMAGING, 2014, 9033
  • [27] Improving depth resolution in digital breast tomosynthesis by iterative image reconstruction
    Roth, Erin G.
    Kraemer, David N.
    Sidky, Emil Y.
    Reiser, Ingrid S.
    Pan, Xiaochun
    MEDICAL IMAGING 2015: PHYSICS OF MEDICAL IMAGING, 2015, 9412
  • [28] Use of the Hotelling observer to optimize image reconstruction in digital breast tomosynthesis
    Sanchez, Adrian A.
    Sidky, Emil Y.
    Pan, Xiaochuan
    JOURNAL OF MEDICAL IMAGING, 2016, 3 (01)
  • [29] Statistical iterative reconstruction to improve image quality for digital breast tomosynthesis
    Xu, Shiyu
    Lu, Jianping
    Zhou, Otto
    Chen, Ying
    MEDICAL PHYSICS, 2015, 42 (09) : 5377 - 5390
  • [30] Breast tomosynthesis: Dosimetry and image quality assessment on phantom
    Meyblum, E.
    Gardavaud, F.
    Dao, T. -H.
    Fournier, V.
    Beaussart, P.
    Pigneur, F.
    Baranes, L.
    Rahmouni, A.
    Luciani, A.
    DIAGNOSTIC AND INTERVENTIONAL IMAGING, 2015, 96 (09) : 931 - 939