Hydroxyapatite-Coated Pins Versus Titanium Alloy Pins in External Fixation at the Wrist: A Controlled Cohort Study

被引:12
|
作者
Pieske, Oliver [1 ]
Pichlmaier, Leopold [1 ]
Kaltenhauser, Felicitas [1 ]
Schramm, Nicolai [2 ]
Rubenbauer, Bianka [1 ]
Greiner, Axel [1 ]
Piltz, Stefan [1 ]
机构
[1] Univ Hosp Munich, Dept Trauma Surg, D-81377 Munich, Germany
[2] Univ Hosp Munich, Dept Clin Radiol, D-81377 Munich, Germany
关键词
External fixation; Pin-track; Complication; Hydroxyapatite; Titanium-alloy; TRACK INFECTION; DISTAL RADIUS; SHORT STATURE; FRACTURES; STRENGTH; SCREWS; COMPLICATIONS; INTERFACE; MODEL; BONE;
D O I
10.1097/TA.0b013e3181e97761
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Background: The purpose of this study was to analyze whether the prevalence of pin-related complications can be reduced by the use of hydroxyapatite (HA)-coated pins in external fixators applied for unstable wrist fractures. Methods: Forty patients (160 pins) were randomized for standard uniplanar fixator treatment with the use of identically designed pins either composed of titanium-alloy (Ti6Al4V) (n = 20) or coated by HA (n = 20). Each pin site was clinically evaluated with regard to erythema, drainage, pain value, and radiologically assessed concerning loosening at T1 (mean, 9 days), T2 (mean, 43 days), and T3 (mean, 56 days). In case of pin-track complication, the patient was followed continuously. The need for antibiotics or additional surgery was documented. Bone mineral density was analyzed by Dual Energy X-ray Absorptiometry. At fixator removal (T2), the pin-extraction strength was measured by the use of a digital-torque-wrench. Results: Two minor pin-track infections requiring oral antibiotics occurred in the HA-pin group (2.7%) (p > 0.05). The vast majority of clinical pin-site parameters were comparable in both groups. At the end of the fixator therapy, there were 16 loose pins (n(Ti6AL4V-group) = 10; n(HA-group) = 6). The rate of loose pins was correlated to patient's age (p < 0.05) but not to bone mineral density values or the occurrence of pin-site infection. Finally, no significant difference between the two groups was detected with regard to the prevalence of clinical relevant pin-site complications (p = 0.80). Conclusions: In external fixation of the wrist, the use of HA-coated pins yields no clinical advantages: there is a trend toward a superior pin-bone anchorage, but a tendency of increased susceptibility for minor pin-track infections.
引用
收藏
页码:845 / 851
页数:7
相关论文
共 50 条
  • [31] POROUS TITANIUM-SURFACED PINS FOR EXTERNAL SKELETAL FIXATION
    DECAMP, CE
    BRINKER, WO
    SOUTASLITTLE, RW
    JOURNAL OF THE AMERICAN ANIMAL HOSPITAL ASSOCIATION, 1988, 24 (03) : 295 - 300
  • [32] Antimicrobial Efficacy of Surface-Coated External Fixation Pins
    Franz H. Furkert
    Jan H. Sörensen
    Jörg Arnoldi
    Bernd Robioneck
    Hartwig Steckel
    Current Microbiology, 2011, 62 : 1743 - 1751
  • [33] Antimicrobial Efficacy of Surface-Coated External Fixation Pins
    Furkert, Franz H.
    Soerensen, Jan H.
    Arnoldi, Joerg
    Robioneck, Bernd
    Steckel, Hartwig
    CURRENT MICROBIOLOGY, 2011, 62 (06) : 1743 - 1751
  • [34] Drug loading and release of Tobramycin from hydroxyapatite coated fixation pins
    Lilja, Mirjam
    Soerensen, Jan Henrik
    Brohede, Ulrika
    Astrand, Maria
    Procter, Philip
    Arnoldi, Joerg
    Steckel, Hartwig
    Stromme, Maria
    JOURNAL OF MATERIALS SCIENCE-MATERIALS IN MEDICINE, 2013, 24 (09) : 2265 - 2274
  • [35] Biomechanical and antibacterial properties of Tobramycin loaded hydroxyapatite coated fixation pins
    Soerensen, Jan Henrik
    Lilja, Mirjam
    Soerensen, Torben Christian
    Astrand, Maria
    Procter, Philip
    Fuchs, Sabine
    Stromme, Maria
    Steckel, Hartwig
    JOURNAL OF BIOMEDICAL MATERIALS RESEARCH PART B-APPLIED BIOMATERIALS, 2014, 102 (07) : 1381 - 1392
  • [36] Drug loading and release of Tobramycin from hydroxyapatite coated fixation pins
    Mirjam Lilja
    Jan Henrik Sörensen
    Ulrika Brohede
    Maria Åstrand
    Philip Procter
    Jörg Arnoldi
    Hartwig Steckel
    Maria Strømme
    Journal of Materials Science: Materials in Medicine, 2013, 24 : 2265 - 2274
  • [37] Biological fixation of hydroxyapatite-coated versus grit-blasted titanium hip stems: a canine study
    Eckardt, A
    Aberman, HM
    Cantwell, HD
    Heine, J
    ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 2003, 123 (01) : 28 - 35
  • [38] Bactericidal activity of antimicrobial coated polyurethane sleeves for external fixation pins
    Forster, H
    Marotta, JS
    Heseltine, K
    Milner, R
    Jani, S
    JOURNAL OF ORTHOPAEDIC RESEARCH, 2004, 22 (03) : 671 - 677
  • [39] Biological fixation of hydroxyapatite-coated versus grit-blasted titanium hip stems: a canine study
    Anke Eckardt
    Harold M. Aberman
    Dan H. Cantwell
    Jochen Heine
    Archives of Orthopaedic and Trauma Surgery, 2003, 123 : 28 - 35
  • [40] INCREASE OF STABILITY IN EXTERNAL FRACTURE FIXATION BY HYDROXYAPATITE-COATED BONE SCREWS
    AUGAT, P
    CLAES, L
    HANSELMANN, KF
    SUGER, G
    FLEISCHMANN, W
    JOURNAL OF APPLIED BIOMATERIALS, 1995, 6 (02) : 99 - 104