The dosimetric impact of image guided radiation therapy by intratumoral fiducial markers

被引:7
|
作者
Yu, Suhong [1 ]
Lawrenson, Lesley [1 ]
Wei, Randy [1 ]
Sehgal, Varun [1 ]
Hanna, Nevine [1 ]
Kuo, Jeffrey [1 ]
Daroui, Parima [1 ]
Ramsinghani, Nilam [1 ]
Al-Ghazi, Muthana [1 ]
机构
[1] Univ Calif Irvine, Med Ctr, Dept Radiat Oncol, Chao Family Comprehens Canc Ctr, Orange, CA USA
关键词
PANCREATIC-CANCER;
D O I
10.1016/j.prro.2015.11.007
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: Pancreatic fiducials have proven superior over other isocenter localization surrogates, including anatomical landmarks and intratumoral or adjacent stents. The more clinically relevant dosimetric impact of image guided radiation therapy (IGRT) using intratumoral fiducial markers versus bony anatomy has not yet been described and is therefore the focus of the current study. Methods and materials: Using daily orthogonal kV or cone beam computed tomography (CBCT) images and positional and dosimetric data were analyzed for 12 consecutive patients treated with fiducial based IGRT and volumetric modulated arc therapy to the intact pancreas. The shifts from fiducial to bone (Delta Fid-Bone) required to realign the daily fiducial-matched pretreatment images (kV, CBCTs) to the planning computed tomography (CT) using bony anatomic landmarks were recorded. The isocenter was then shifted by (Delta Fid-Bone) for 5 evenly spaced treatments, and the dosimetric impact of Delta Fid-Bone was calculated for planning target volume coverage (PTV50.4 and PTV47.9) and organs at risk (liver, kidney, and stomach/duodenum). Results: The Delta Fid-Bone were greatest in the superoinferior direction (Delta Fid-Bone anteroposterior, 2.7 +/- 3.0; left-right, 2.8 +/- 2.8; superoinferior, 6.3 +/- 7.9 mm; mean +/- standard deviation; P = .03). PTV50.4 coverage was reduced by 13% (fiducial plan 95 +/- 2.0 vs bone plan 82 +/- 12%; P = .005; range, 5%-52%; >5% loss in all; and >10% loss in 42% of patients), and to a lesser degree for PTV47.9 (difference, -8%; range, 1%-30%; fiducial plan 100 +/- 0.3% vs bone plan 92 +/- 7.6%; P = .003; with reductions of >5% in 66% and >10% in 33% of patients). The dosimetric impact of Delta Fid-Bone on the organs at risk was not significant. Positional shifts for kV- and CBCT-based realignments were nearly identical. Conclusion: Compared with matching by fiducial markers, IGRT matched by bony anatomy substantially reduces the PTV50.4 and PTV47.9 coverage, supporting the use of intratumoral pancreatic markers for improved targeting in IGRT for pancreatic cancer. (C) 2016 American Society for Radiation Oncology. Published by Elsevier Inc. All rights reserved.
引用
收藏
页码:276 / 283
页数:8
相关论文
共 50 条
  • [41] Validating Fiducial Markers for Image Guided Radiation Therapy for Accelerated Partial Breast Irradiation in Early Stage Breast Cancer Treatment
    Pritz, J.
    Zhang, G.
    Forster, K. M.
    Harris, E. E. R.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2009, 75 (03): : S598 - S599
  • [42] Feasibility of Gold Fiducial Markers as a Surrogate for Gross Tumor Volume Position in Image-Guided Radiation Therapy of Rectal Cancer
    van den Ende, Roy P. J.
    Kerkhof, Ellen M.
    Rigter, Lisanne S.
    van Leerdam, Monique E.
    Peters, Femke P.
    van Triest, Baukelien
    Staring, Marius
    Marijnen, Corrie A. M.
    van der Heide, Uulke A.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2019, 105 (05): : 1151 - 1159
  • [43] Infection after prostatic transrectal fiducial marker implantation for image guided radiation therapy
    Reynaud, T.
    Ben Aicha, I.
    Carignan, D.
    Pelchat, C.
    Fiset, C.
    Foster, W.
    Martin, A. G.
    Vigneault, E.
    CANCER RADIOTHERAPIE, 2023, 27 (03): : 214 - 218
  • [44] Safety and Efficacy of Percutaneous Fiducial Marker Implantation for Image-guided Radiation Therapy
    Kothary, Nishita
    Heit, Jeremy J.
    Louie, John D.
    Kuo, William T.
    Loo, Billy W., Jr.
    Koong, Albert
    Chang, Daniel T.
    Hovsepian, David
    Sze, Daniel Y.
    Hofmann, Lawrence V.
    JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2009, 20 (02) : 235 - 239
  • [45] Positioning and Dosimetric Uncertainties in Image-Guided Radiation Therapy with Respiratory Gating
    Ali, I.
    Hossain, S.
    Algan, O.
    Syzek, E.
    Ahmad, S.
    MEDICAL PHYSICS, 2016, 43 (06) : 3673 - 3673
  • [46] Fiducial markers configuration optimization in image-guided surgery
    Bao, Nan
    Chen, Yu
    Yue, Yong
    Li, Hong
    Cui, Zhiming
    Zhuang, Jinfeng
    Tian, Shi
    Kang, Yan
    BIO-MEDICAL MATERIALS AND ENGINEERING, 2014, 24 (06) : 3361 - 3371
  • [47] Fiducial markers for image-guided partial breast irradiation
    Trovo, Marco
    Polesel, Jerry
    Biasutti, Cristina
    Sartor, Giovanna
    Roncadin, Mario
    Trovo, Gaetano Mauro
    RADIOLOGIA MEDICA, 2013, 118 (07): : 1212 - 1219
  • [48] Image guided neutron therapy: A six degree correction method using implanted fiducial markers
    Wang, Y.
    Burmeister, J.
    Van den Heuvel, F.
    Yudelev, M.
    MEDICAL PHYSICS, 2007, 34 (06) : 2372 - 2372
  • [49] Validating Fiducial Markers for Image-Guided Radiation Therapy for Accelerated Partial Breast Irradiation in Early-Stage Breast Cancer
    Park, Catherine K.
    Pritz, Jakub
    Zhang, Geoffrey G.
    Forster, Kenneth M.
    Harris, Eleanor E. R.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2012, 82 (03): : E425 - E431
  • [50] Toxicity and Biochemical Failure Following Image-Guided Prostate Intensity Modulated Radiation Therapy: Fiducial Markers Versus Electromagnetic Transponders
    Chang, L.
    Li, T.
    Horwitz, E.
    Chen, D. Y. T.
    Viterbo, R.
    Kutikov, A.
    Greenberg, R.
    Buyyounouski, M. K.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2013, 87 (02): : S368 - S369