Feasibility study of navigated endoscopy for the placement of high dose rate brachytherapy applicators in the esophagus and lung

被引:2
|
作者
Weersink, Robert A. [1 ,2 ,3 ,4 ]
Qiu, Jimmy [4 ]
Martinez, Diego [2 ]
Rink, Alexandra [1 ,2 ,4 ]
Borg, Jette [1 ,2 ]
Di Tomasso, Anne [2 ]
Irish, Jonathon C. [4 ,5 ]
Jaffray, David A. [1 ,2 ,3 ,4 ,6 ]
机构
[1] Univ Toronto, Dept Radiat Oncol, Toronto, ON M5T 1P5, Canada
[2] Princess Margaret Canc Ctr, Radiat Med Program, Toronto, ON M5G 1X6, Canada
[3] Univ Toronto, Inst Biomat & Biomed Engn, Toronto, ON M5S 3G9, Canada
[4] Univ Hlth Network, Techna Inst, Toronto, ON M5G 1L5, Canada
[5] Univ Toronto, Dept Surg Oncol, Toronto, ON M5T 1P5, Canada
[6] Univ Toronto, Dept Med Biophys, Toronto, ON M5G 1L7, Canada
基金
加拿大健康研究院;
关键词
brachytherapy; electromagnetic tracking; intraluminal applicators; ELECTROMAGNETIC TRACKING; CATHETER RECONSTRUCTION;
D O I
10.1002/mp.13997
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose To evaluate the electromagnetic (EM) tracking of endoscopes and applicators as a method of positioning a high dose rate (HDR) luminal applicator. Method An anatomical phantom consisting of a rigid trachea and flexible esophagus was used to compare applicator placement measurements using EM tracking vs the traditional method using two-dimensional (2D) fluoroscopy and surface skin markers. The phantom included a tumor in the esophagus and several pairs of optically visible points inside the lumen that were used to simulate proximal and distal ends of tumors of varying lengths. The esophagus tumor and lung points were visible on a computed tomography (CT) image of the phantom, which was used as ground truth for the measurements. The EM tracking system was registered to the CT image using fiducial markers. A flexible endoscope was tracked using the EM system and the locations of the proximal and distal ends of the tumor identified and this position recorded. An EM-tracked applicator was then inserted and positioned relative to the tumor markings. The applicator path was mapped using the EM tracking. The gross tumor length (GTL) and the distance between the first dwell position and distal edge of tumor (offset) were measured using the EM tracking and 2D fluoroscopy methods and compared to the same measurements on the CT image. Results The errors in GTL using EM tracking were on average -0.5 +/- 1.7 mm and 0.7 +/- 3.6 mm for esophagus and lung measurements, similar to errors measured using the 2D fluoroscopy method of -0.9 +/- 1.2 mm and 3.4 +/- 4.4 mm. Offset measurements were slightly larger while using EM tracking relative to the fluoroscopy method but these were not statistically significant. Conclusions Electromagnetic tracking for placement of lumen applicators is feasible and accurate. Tracking of the endoscope that is used to identify the proximal and distal ends of the tumor and of the applicator during insertion generates accurate three-dimensional measurements of the applicator path, GTL and offset. Guiding the placement of intraluminal applicators using EM navigation is potentially attractive for cases with complex insertions, such as those with nonlinear paths or multiple applicator insertions.
引用
收藏
页码:917 / 926
页数:10
相关论文
共 50 条
  • [1] The importance of immobilization of gynecological applicators in high dose rate brachytherapy
    Pella, S.
    Mikko, H.
    Nicolae, D.
    Casey, C.
    Shereen, C.
    Janeil, P.
    RADIOTHERAPY AND ONCOLOGY, 2016, 119 : S936 - S936
  • [2] Vaginal Applicators in High-Dose Rate Brachytherapy for Endometrial Carcinoma: A Dosimetric Study
    Khan, Aijaz Ahmad
    Vijay, Anuj
    Wani, Shaqul Qamar
    Haq, Malik Mohib-ul
    TURK ONKOLOJI DERGISI-TURKISH JOURNAL OF ONCOLOGY, 2023, 38 (02): : 191 - 200
  • [3] High dose rate brachytherapy for superficial cancer of the esophagus
    Maingon, P
    D'Hombres, A
    Truc, G
    Barillot, I
    Michiels, C
    Bedenne, L
    Horiot, JC
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2000, 46 (01): : 71 - 76
  • [4] Dosimetric evaluation of Gammamed High Dose Rate intraluminal brachytherapy applicators
    Supe, Sanjay Sudhakar
    Varatharaj, Chandraraj
    Bijina, Themantavida Krishnan
    Bondel, Shwetha
    Sathiyan, Saminathan
    Ganesh, Kadirampatti Mani
    Arunkumar, Thangaraj
    Ravikumar, Manickam
    REPORTS OF PRACTICAL ONCOLOGY AND RADIOTHERAPY, 2007, 12 (06) : 313 - 317
  • [5] Dosimetric Evaluation of Gammamed High Dose Rate Intraluminal Brachytherapy Applicators
    Chandraraj, V.
    Supe, S.
    Manickam, R.
    MEDICAL PHYSICS, 2009, 36 (06)
  • [6] High dose rate brachytherapy with customized applicators for malignant facial skin lesions
    Jumeau, R.
    Renard-Oldrini, S.
    Courrech, F.
    Buchheit, I.
    Oldrini, G.
    Vogin, G.
    Peiffert, D.
    CANCER RADIOTHERAPIE, 2016, 20 (05): : 341 - 346
  • [7] High dose rate brachytherapy with stiff applicators in the treatment of nasopharyngeal lymphoepithelioma in childhood
    Novaes, P
    Nakamura, R
    Pellizzon, A
    Ferrigno, R
    Maia, M
    Fogaroli, R
    Salvajoli, J
    Antonelli, C
    De Camargo, B
    RADIOTHERAPY AND ONCOLOGY, 2004, 71 : S59 - S59
  • [8] 3D Printed Surface Applicators for High Dose Rate Brachytherapy
    Clarke, Scott
    Yewondwossen, Mammo
    Robar, James
    MEDICAL PHYSICS, 2016, 43 (08) : 4934 - 4935
  • [9] Dosimetric comparison between different applicators for high dose rate gynecologic brachytherapy
    Portocarrero-Bonifaz, Andres
    Syed, Salman
    Kassel, Maxwell
    McKenzie, Grant W.
    Shah, Vishwa M.
    Forry, Bryce
    Masters, Adrianna H.
    Gaskins, Jeremy T.
    Sowards, Keith T.
    Silva, Scott R.
    RADIOTHERAPY AND ONCOLOGY, 2024, 194 : S260 - S262
  • [10] Optimal Design of Multichannel Vaginal Cylinder Applicators for High Dose Rate Brachytherapy
    Park, S.
    Demanes, J.
    Kamrava, M.
    MEDICAL PHYSICS, 2012, 39 (06) : 3611 - 3611