Transcutaneous monitoring of carbon dioxide tension after cardiothoracic surgery in infants and children

被引:37
|
作者
Tobias, JD
Wilson, WR
Meyer, DJ
机构
[1] Univ Missouri, Hlth Sci Ctr, Dept Child Hlth, Columbia, MO 65212 USA
[2] Univ Missouri, Dept Anesthesiol, Columbia, MO 65212 USA
[3] Univ Missouri, Dept Surg, Columbia, MO 65212 USA
[4] Univ Missouri, Div Pediat Crit Care Pediat Anesthesiol, Columbia, MO 65212 USA
[5] Univ Missouri, Div Pediat Cardiothorac Surg, Columbia, MO 65212 USA
来源
ANESTHESIA AND ANALGESIA | 1999年 / 88卷 / 03期
关键词
D O I
10.1097/00000539-199903000-00012
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
In this prospective investigation, we evaluated the efficacy and accuracy of transcutaneous monitoring of CO2 (TC-CO2) in infants and children after cardiothoracic surgery. Cardiothoracic surgery patients whose ETCO2 and arterial CO2 values did not correlate (gradient greater than or equal to 5 mm Hg) during the first postoperative hour underwent placement of the TC electrode (30 of 33 patients). If the TC-CO2 to arterial difference was greater than or equal to 5 mm Hg,the TC-CO2 electrode was recalibrated and reapplied on another site. If the discrepancy was still greater than or equal to 5 mm Hg, the case was considered a clinical failure and no further data were collected (3 of 30 patients). If the arterial to TC gradient was <5 mm Hg, the patient was included in the data collection (27 of 30 patients). One to five sample sets (TC and arterial CO2) were collected from these patients. Statistical analysis included linear regression analysis and Bland-Altman analysis. The cohort for the study included 27 patients ranging in age from 2 days to 9 yr and in weight from 3.2 to 25 kg. A total of 101 sample sets were analyzed. The mean +/- SD absolute difference between the TC-CO2 and arterial CO2 was 1.7 +/- 1.4 mm Hg (range 0-9 mm Hg). The TC-CO2 to arterial CO2 difference was 0-2 mm Hg in 82 of 101 values (81%), 3-5 mm Hg in 18 of 101 values (18%), and >6 mm Hg in 1 of 101 values (1%). Linear regression analysis revealed a slope of 0.90, an r value of 0.9410, and an r(2) value of 0.8854 (P < 0.0001). Bland-Altman analysis revealed a bias of 0.58 mm Hg with a precision of +/-2.1 nam Hg when comparing the TC-CO2 with the arterial CO2. Implications: We conclude that, with certain caveats in mind, including the need to correlate the transcutaneous CO2 with an initial arterial CO2 value, transcutaneous CO2 monitoring can be used to estimate arterial CO2 in most neonates and children after cardiothoracic surgery.
引用
收藏
页码:531 / 534
页数:4
相关论文
共 50 条
  • [1] Transcutaneous carbon dioxide monitoring in infants and children
    Tobias, Joseph D.
    PEDIATRIC ANESTHESIA, 2009, 19 (05) : 434 - 444
  • [2] TRANS-CUTANEOUS OXYGEN AND CARBON-DIOXIDE MONITORING AFTER CARDIOTHORACIC SURGERY
    TURIEL, M
    PICCOLO, M
    MORANDI, A
    BOTTA, A
    SANTOLI, C
    MALLIANI, A
    EUROPEAN JOURNAL OF RESPIRATORY DISEASES, 1986, 69 : A170 - A170
  • [3] Transcutaneous carbon dioxide monitoring during high-frequency oscillatory ventilation in infants and children
    Berkenbosch, JW
    Tobias, JD
    CRITICAL CARE MEDICINE, 2002, 30 (05) : 1024 - 1027
  • [4] Application of Transcutaneous Carbon Dioxide Tension Monitoring with Low Electrode Temperatures in Premature Infants in the Early Postnatal Period
    Hirata, Katsuya
    Nishihara, Masahiro
    Oshima, Yukari
    Hirano, Shinya
    Kitajima, Hiroyuki
    AMERICAN JOURNAL OF PERINATOLOGY, 2014, 31 (05) : 435 - 439
  • [5] Transcutaneous carbon dioxide monitoring
    McCormack, J. G.
    ANAESTHESIA, 2007, 62 (08) : 850 - 851
  • [6] The clinical application of transcutaneous carbon dioxide monitoring during rigid bronchoscopy or microlaryngeal surgery in children
    van Wijk, Jan J.
    Gangaram-Panday, Norani H.
    van Weteringen, Willem
    Pullens, Bas
    Bernard, Simone E.
    Hoeks, Sanne E.
    Reiss, Irwin K. M.
    Stolker, Robert J.
    Staals, Lonneke M.
    JOURNAL OF CLINICAL ANESTHESIA, 2024, 100
  • [7] Ambulatory transcutaneous carbon dioxide monitoring for children with neuromuscular disease
    Shi, Jenny
    Chiang, Jackie
    Ambreen, Munazzah
    Snow, Nadia
    Mocanu, Cora
    McAdam, Laura
    Goldstein, Roger S.
    Rose, Louise
    Amin, Reshma
    SLEEP MEDICINE, 2023, 101 : 221 - 227
  • [8] Transcutaneous carbon dioxide monitoring - A reply
    Bolliger, D.
    Steiner, L. A.
    Filipovic, M.
    Seeberger, M. D.
    ANAESTHESIA, 2007, 62 (08) : 851 - 851
  • [9] TRANSCUTANEOUS CARBON-DIOXIDE MONITORING
    RENNIE, JM
    ARCHIVES OF DISEASE IN CHILDHOOD-FETAL AND NEONATAL EDITION, 1990, 65 (04): : 345 - 346
  • [10] Effects of the transcutaneous electrode temperature on the accuracy of transcutaneous carbon dioxide tension
    Sorensen, Line C.
    Brage-Andersen, Lene
    Greisen, Gorm
    SCANDINAVIAN JOURNAL OF CLINICAL & LABORATORY INVESTIGATION, 2011, 71 (07): : 548 - 552