Mechanical stability assessment of novel orthodontic mini-implant designs: Part 2

被引:19
|
作者
Hong, Christine
Peter Truong
Ha Na Song
Wu, Benjamin M. [2 ,3 ,4 ]
Moon, Won [1 ]
机构
[1] Univ Calif Los Angeles, Sch Dent, Sect Orthodont, Los Angeles, CA 90095 USA
[2] Univ Calif Los Angeles, Dept Bioengn, Los Angeles, CA 90095 USA
[3] Univ Calif Los Angeles, Calif NanoSyst Inst, Los Angeles, CA 90095 USA
[4] Univ Calif Los Angeles, Sch Engn, Los Angeles, CA 90095 USA
关键词
Mini-implants; TADs; Stability; Buccal bone thickness; Cortical bone; Micro-CT; CORTICAL BONE THICKNESS; DISTAL MOVEMENT; ANCHORAGE; ROOT; DISTALIZATION; PROXIMITY; FAILURE; LENGTH; MOLARS; SHAPE;
D O I
10.2319/031011-176.1
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Objective: To assess the mechanical stability of a newly revised orthodontic mini-implant design (N2) compared with a design introduced in Part 1 of the study (N1) and the most widely-used commercially-available design (CA). To evaluate the mean buccal bone thickness of maxillary and mandibular posterior teeth using cone-beam computed tomography (CBCT). Materials and Methods: From the CBCT scans of 20 patients, six tomographic cross-sections were generated for each tooth. Buccal bone thickness was measured from the most convex point on the bone to the root surface. CA (1.5 mm in diameter and 6 mm in length), N1, and N2 (shorter and narrower than N1) were inserted in simulated bone with cortical and trabecular bone layers. Mechanical stability was compared in vitro through torque and lateral displacement tests. Results: The bone thickness ranged from 2.26 to 3.88 mm. Maximum insertion torque was decreased significantly in N2 compared to N1. However, force levels for all displacement distances and torque ratio were the highest in N2, followed by N1 and CA (alpha = .05). Conclusions: Both torque and lateral displacement tests highlighted the enhanced stability of N2 compared with CA. Design revisions to N1 effectively mitigated N1's high insertion torque and thus potentially reduced microdamage to the surrounding bone. The N2 design is promising as evidenced by enhanced stability and high mechanical efficiency. Moreover, N2 is not limited to placement in interradicular spaces and has the capacity to be placed in the buccal bone superficial to the root surface with diminished risk of endangering nearby anatomic structures during placement and treatment. (Angle Orthod. 201181:1001-1009.)
引用
收藏
页码:1001 / 1009
页数:9
相关论文
共 50 条
  • [31] Summary of: Advances in orthodontic anchorage with the use of mini-implant techniques
    Cousley, R. R. J.
    Sandler, P. J.
    BRITISH DENTAL JOURNAL, 2015, 218 (03) : 178 - 179
  • [32] A prospective study of factors associated with orthodontic mini-implant survival
    Tsai, Chia-Chun
    Chang, Hong-Po
    Pan, Chin-Yun
    Chou, Szu-Ting
    Tseng, Yu-Chuan
    JOURNAL OF ORAL SCIENCE, 2016, 58 (04) : 515 - 521
  • [33] Bivariate optimization of orthodontic mini-implant thread height and pitch
    Shuning Shen
    Yingying Sun
    Chen Zhang
    Yongjin Yang
    Zhiren Li
    Xingwei Cai
    Yinzhong Duan
    Tao Li
    International Journal of Computer Assisted Radiology and Surgery, 2015, 10 : 109 - 116
  • [34] Primary migration of a mini-implant under a functional orthodontic loading
    Joseph W. Pittman
    Anand Navalgund
    Steve H. Byun
    Hechang Huang
    Albert H. Kim
    Do-Gyoon Kim
    Clinical Oral Investigations, 2014, 18 : 721 - 728
  • [35] Systematic review of mini-implant displacement under orthodontic loading
    Manuel Nienkemper
    Jrg Handschel
    Dieter Drescher
    International Journal of Oral Science, 2014, 6 (01) : 1 - 6
  • [36] Systematic review of mini-implant displacement under orthodontic loading
    Manuel Nienkemper
    Jrg Handschel
    Dieter Drescher
    International Journal of Oral Science, 2014, (01) : 1 - 6
  • [37] Prevalence and cause of mini-implant failure encountered by orthodontic residents
    Sreenivasagan S.
    Subramanian A.K.
    Rengalakshmi S.
    Journal of Long-Term Effects of Medical Implants, 2021, 31 (04) : 1 - 4
  • [38] Effects of surface treatment on the osseointegration potential of orthodontic mini-implant
    Jeon, Mi-Sun
    Kang, Yoon-Goo
    Mo, Sung-Seo
    Lee, Keun-Hye
    Kook, Yoon-Ah
    Kim, Seong-Hun
    KOREAN JOURNAL OF ORTHODONTICS, 2008, 38 (05) : 328 - 336
  • [39] Bone thickness of the palate for orthodontic mini-implant anchorage in adults
    Kang, Sungmin
    Lee, Shin-Jae
    Ahn, Sug-Joon
    Heo, Min-Suk
    Kim, Tae-Woo
    AMERICAN JOURNAL OF ORTHODONTICS AND DENTOFACIAL ORTHOPEDICS, 2007, 131 (04) : S74 - S81
  • [40] The impact of healing time before loading on orthodontic mini-implant stability: A histomorphometric study in minipigs
    Oltramari-Navarro, Paula V. P.
    Navarro, Ricardo L.
    Henriques, Jose Fernando C.
    Cestari, Tania M.
    Francischone, Carlos E.
    Taga, Rumio
    McNamara, James A., Jr.
    ARCHIVES OF ORAL BIOLOGY, 2013, 58 (07) : 806 - 812