Novel Protective Lead Shield and Pulse Fluoroscopy can Reduce Radiation Exposure during the ERCP Procedure

被引:16
|
作者
Kurihara, Toshio [1 ]
Ito, Takao [1 ]
Sofuni, Atsushi [1 ]
Itokawa, Fumihide [1 ]
Tsuchiya, Takayoshi [1 ]
Ishii, Kentaro [1 ]
Tsuji, Shujiro [1 ]
Ikeuchi, Nobuhito [1 ]
Moriyasu, Fuminori [1 ]
机构
[1] Tokyo Med Univ, Dept Gastroenterol & Hepatol, Shinjuku Ku, Tokyo 1600023, Japan
关键词
Radiation exposure; ERCP; Lead shield; CARDIAC-CATHETERIZATION; PERSONNEL;
D O I
10.5754/hge11764
中图分类号
R57 [消化系及腹部疾病];
学科分类号
摘要
Background/Aims: ERCP-related procedures involve radiation exposure of patients and medical staff. We developed a novel protective lead shield which is attached around the fluoroscopy generator. Here we examine levels of radiation exposure to patients, endoscopists and assistants, and evaluate the usefulness of the newly designed protective shield. Methodology: Four-hundred and seventy-one ERCP procedures were performed from April 2006 to April 2007. At first, we compared the radiation dose of consecutive fluoroscopy conditions with pulse fluoroscopy of 15 per second and then the radiation dose with and without the protective shield. Next, we measured the radiation exposure of endoscopists and assistants in the clinical setting monitored by digital dosimeter during ERCP procedure. Results: The radiation dose was the most at the 45 direction. Using pulse fluoroscopy of 15 per second the radiation dose of patients and endoscopists decreased by about half. Using both pulse fluoroscopy of 15 per second and the protective shield, the radiation dose at the endoscopist's position was reduced up to 97%. The total fluoroscopy time was 5851 minutes in the 471 ERCP cases. Using pulse 15 and the protective lead shield, the radiation exposure dose of one endoscopist and two assistants were 2430.8, 2673.9 and 1375.0 mu Sv, respectively. Conclusions: Novel protective lead shield in combination with pulse fluoroscopy can significantly reduce the radiation exposure leading to avoid unnecessary radiation exposure to patients and medical staff.
引用
收藏
页码:709 / 712
页数:4
相关论文
共 50 条
  • [41] Optimized Fluoroscopy Setting and Appropriate Project Position Can Reduce X-ray Radiation Doses Rates during Electrophysiology Procedures
    Hou Bing-Bo
    Yao Yan
    Wu Ling-Min
    Qiao Yu
    Zheng Li-Hui
    Ding Li-Gang
    Chen Gang
    Zhang Shu
    中华医学杂志英文版, 2015, 128 (09) : 1151 - 1153
  • [42] Effect of Protective Lead Curtains on Scattered Radiation Exposure to the Operator During Ureteroscopy for Stone Disease: A Controlled Trial
    Inoue, Takaaki
    Komemushi, Atsushi
    Murota, Takashi
    Yoshida, Takashi
    Taguchi, Makoto
    Kinoshita, Hidefumi
    Matsuda, Tadashi
    UROLOGY, 2017, 109 : 60 - 66
  • [43] Radiation shielding effects of lead equivalent thickness of a radiation protective apron and distance during C-arm fluoroscopy-guided pain interventions: A randomized trial
    Kim, Cho Long
    Jeong, Hae Chang
    Kim, Jae Hun
    MEDICINE, 2023, 102 (48) : E36447
  • [44] A novel nonfluoroscopic catheter visualization system (LocaLisa) to reduce radiation exposure during catheter ablation of supraventricular tachycardias
    Kirchhof, P
    Loh, P
    Eckardt, L
    Ribbing, M
    Rolf, S
    Eick, O
    Wittkampf, F
    Borggrefe, M
    Breithardt, G
    Haverkamp, W
    AMERICAN JOURNAL OF CARDIOLOGY, 2002, 90 (03): : 340 - +
  • [45] A pilot single-center prospective randomized trial to assess the short-term effect of a flashing warning light on reducing fluoroscopy time and radiation exposure during ERCP
    Zeng, Hong-Ze
    Liu, Qin
    Chen, Hong-Lin
    Liu, Wei
    Zeng, Qi-Shan
    Wu, Chun-Cheng
    Hu, Bing
    GASTROINTESTINAL ENDOSCOPY, 2018, 88 (02) : 261 - 266
  • [46] The effectiveness of a free-standing lead-shield in reducing spine surgeon radiation exposure during intraoperative 3-dimensional imaging
    Guenthner, Guy
    Eddy, Alexander
    Sembrano, Jonathan
    Polly, David W., Jr.
    Martin, Christopher T.
    SPINE JOURNAL, 2020, 20 (10): : 1685 - 1691
  • [47] The effectiveness of additional lead-shielding drape and low pulse rate fluoroscopy in protecting staff from scatter radiation during cardiac resynchronization therapy (CRT)
    Morishima, Yoshiaki
    Chida, Koichi
    Katahira, Yoshiaki
    JAPANESE JOURNAL OF RADIOLOGY, 2019, 37 (01) : 95 - 101
  • [48] The effectiveness of additional lead-shielding drape and low pulse rate fluoroscopy in protecting staff from scatter radiation during cardiac resynchronization therapy (CRT)
    Yoshiaki Morishima
    Koichi Chida
    Yoshiaki Katahira
    Japanese Journal of Radiology, 2019, 37 : 95 - 101
  • [49] Radiation exposure during pulmonary vein isolation can be dramatically reduced in German EP labs: comparison of optimized versus non-optimized fluoroscopy
    Kleemann, T.
    Brachmann, J.
    Lewalter, T.
    Kuck, K.
    Willems, S.
    Hoffmann, E.
    Eckardt, L.
    Hochadel, M.
    Senges, J.
    Zahn, R.
    EUROPEAN HEART JOURNAL, 2015, 36 : 565 - 565
  • [50] BENCH-TOP STUDY USING THE NOVEL LASER DARRT TO REDUCE RADIATION EXPOSURE DURING PERCUTANEOUS RENAL ACCESS
    Wilkinson, M.
    Ewald, J. M.
    Engelhart, S. M.
    Wagner, H.
    Cheng, J.
    Hajiha, M.
    Baldwin, D.
    JOURNAL OF INVESTIGATIVE MEDICINE, 2018, 66 (01) : 182 - 182