A novel and clinically useful dynamic conformal arc (DCA)-based VMAT planning technique for lung SBRT

被引:16
|
作者
Pokhrel, Damodar [1 ,2 ]
Visak, Justin [1 ]
Sanford, Lana [1 ]
机构
[1] Univ Kentucky, Dept Radiat Med, Med Phys Grad Program, Lexington, KY 40506 USA
[2] Univ Kentucky, Dept Radiat Med, Markey Canc Ctr, Lexington, KY 40506 USA
来源
关键词
DCA-based VMAT; FFF-beam; lung SBRT; single dose; BODY RADIATION-THERAPY; STEREOTACTIC RADIOTHERAPY; SINGLE-FRACTION; 30; GY; CANCER; BEAMS; PNEUMONITIS; TOMOGRAPHY; VALIDATION; SCHEDULES;
D O I
10.1002/acm2.12878
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose Volumetric modulated arc therapy (VMAT) is gaining popularity for stereotactic treatment of lung lesions for medically inoperable patients. Due to multiple beamlets in delivery of highly modulated VMAT plans, there are dose delivery uncertainties associated with small-field dosimetry error and interplay effects with small lesions. We describe and compare a clinically useful dynamic conformal arc (DCA)-based VMAT (d-VMAT) technique for lung SBRT using flattening filter free (FFF) beams to minimize these effects. Materials and Methods Ten solitary early-stage I-II non-small-cell lung cancer (NSCLC) patients were treated with a single dose of 30 Gy using 3-6 non-coplanar VMAT arcs (clinical VMAT) with 6X-FFF beams in our clinic. These clinically treated plans were re-optimized using a novel d-VMAT planning technique. For comparison, d-VMAT plans were recalculated using DCA with user-controlled field aperture shape before VMAT optimization. Identical beam geometry, dose calculation algorithm, grid size, and planning objectives were used. The clinical VMAT and d-VMAT plans were compared via RTOG-0915 protocol compliances for conformity, gradient indices, and dose to organs at risk (OAR). Additionally, treatment delivery efficiency and accuracy were recorded. Results All plans met RTOG-0915 requirements. Comparing with clinical VMAT, d-VMAT plans gave similar target coverage with better target conformity, tighter radiosurgical dose distribution with lower gradient indices, and dose to OAR. Lower total number of monitor units and small beam modulation factor reduced beam-on time by 1.75 min (P < 0.001), on average (maximum up to 2.52 min). Beam delivery accuracy was improved by 2%, on average (P < 0.05) and maximum up to 6% in some cases for d-VMAT plans. Conclusion This simple d-VMAT technique provided excellent plan quality, reduced intermediate dose-spillage, and dose to OAR while providing faster treatment delivery by significantly reducing beam-on time. This novel treatment planning approach will improve patient compliance along with potentially reducing intrafraction motion error. Moreover, with less MLC modulation through the target, d-VMAT could potentially minimize small-field dosimetry errors and MLC interplay effects. If available, d-VMAT planning approach is recommended for future clinical lung SBRT plan optimization.
引用
收藏
页码:29 / 38
页数:10
相关论文
共 50 条
  • [1] Dynamic Conformal Sub-Arc Vs VMAT Planning for SBRT Lung Treatment
    Yount, C.
    Sintay, B.
    Wiant, D.
    Pursley, J.
    Terrell, J.
    Maurer, J.
    Baydush, A.
    [J]. MEDICAL PHYSICS, 2013, 40 (06)
  • [2] Escalating Stereotactic Body Radiotherapy (SBRT) Dose in the Treatment of Large (&gt;5 cm) Lung Tumors Via a Novel Dynamic Conformal Arc (DCA)-Based SIB-VMAT
    Pokhrel, D.
    Critchfield, L.
    Bernard, M.
    Randall, M.
    McGarry, R.
    [J]. MEDICAL PHYSICS, 2020, 47 (06) : E766 - E767
  • [3] Dynamic Conformal Arc Therapy for Lung and Liver SBRT
    Stathakis, S.
    Licon, A.
    Saenz, D.
    Kirby, N.
    Rasmussen, K.
    Crownover, R.
    Li, Y.
    Mavroidis, P.
    Papanikolaou, N.
    [J]. MEDICAL PHYSICS, 2018, 45 (06) : E338 - E338
  • [4] Negative margin technique - a novel planning strategy to improve dose conformation in SBRT using dynamic conformal arc delivery
    Kim, Siyong
    Kim, Taeho
    Ko, Stephen J.
    Serago, Christopher
    Smith, Ashley A.
    Vallow, Laura A.
    Peterson, Jennifer L.
    Lee, Rena
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2013, 14 (05): : 79 - 89
  • [5] Novel Clinically Weight-Optimized Dynamic Conformal Arcs (WO-DC A) for Liver SBRT: A Comparison with Volumetric Modulated Arc Therapy (VMAT)
    Saglann, Yucel
    Bolukbasi, Yasemin
    Atasoy, Ali Ihsan
    Karakose, Fatih
    Budak, Mustafa
    Alpan, Vildan
    Topkan, Erkan
    Selek, Ugur
    [J]. THERAPEUTICS AND CLINICAL RISK MANAGEMENT, 2021, 17 : 1053 - 1064
  • [6] Dynamic Conformal Sub-Arc Vs 3D Planning for SBRT Lung Treatment
    Baydush, A.
    Wiant, D.
    Terrell, J.
    Pursley, J.
    Yount, C.
    Maurer, J.
    Sintay, B.
    [J]. MEDICAL PHYSICS, 2013, 40 (06)
  • [7] Robust Planning for ITV-Based Lung VMAT SBRT
    Peca, S.
    Buczek, A.
    [J]. MEDICAL PHYSICS, 2022, 49 (06) : E946 - E947
  • [8] A 3D-conformal technique is better than IMRT or VMAT for lung SBRT
    Cai, Jing
    Malhotra, Harish K.
    Orton, Colin G.
    [J]. MEDICAL PHYSICS, 2014, 41 (04)
  • [9] A novel approach to automatic planning: robust templates for lung VMAT SBRT
    Marrazzo, L.
    Arilli, C.
    Calusi, S.
    Casati, M.
    Talamonti, C.
    Bonomo, P.
    Livi, L.
    Pallotta, S.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2019, 133 : S496 - S497
  • [10] Optimal Treatment Planning Strategy Among Arc Arrangements for Prostate SBRT with VMAT Technique
    Chung, J.
    Kang, S.
    Suh, T.
    Lee, J.
    Kim, J.
    Eom, K.
    [J]. MEDICAL PHYSICS, 2016, 43 (06) : 3607 - 3607