T2-weighted Lung Imaging Using a 0.55-T MRI System

被引:19
|
作者
Campbell-Washburn, Adrienne E. [1 ]
Malayeri, Ashkan A. [3 ]
Jones, Elizabeth C. [3 ]
Moss, Joel [2 ]
Fennelly, Kevin P. [2 ]
Olivier, Kenneth N. [2 ]
Chen, Marcus Y. [1 ]
机构
[1] NHLBI, Cardiovasc Branche, Div Intramural Res, NIH,Dept Hlth & Human Serv, Bldg 10,Room BID-47,10 Ctr Dr, Bethesda, MD 20892 USA
[2] NHLBI, Pulm Branche, Div Intramural Res, NIH,Dept Hlth & Human Serv, Bldg 10,Room BID-47,10 Ctr Dr, Bethesda, MD 20892 USA
[3] NIH, Dept Radiol & Imaging Sci, Clin Ctr, Dept Hlth & Human Serv, Bethesda, MD 20892 USA
来源
RADIOLOGY-CARDIOTHORACIC IMAGING | 2021年 / 3卷 / 03期
关键词
MRI; Pulmonary; Technology Assessment; PULMONARY NODULES; IMMUNOCOMPROMISED PATIENTS; ECHO; SEQUENCES;
D O I
10.1148/ryct.2021200611
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To assess a 0.55-T MRI system for imaging lung disease and to compare image quality with clinical CT scans. Materials and Methods: In this prospective study conducted between November 2018 and December 2019, respiratory-triggered T2weighted turbo spin-echo MRI at 0.55 T was compared with clinical CT scans in 24 participants (mean age, 59 years 6 16 [standard deviation]; 18 women) with common lung abnormalities. MR images were reviewed and scored by experienced readers. Abnormal findings identified with MRI and CT were compared using the Cohen k statistic. Results: High-quality structural pulmonary MR images were attained with an average acquisition time of 11 minutes 6 3. MRI generated sufficient image quality to robustly detect bronchiectasis (k = 0.61), consolidative opacities (k = 1.00), cavitary lesions (k = 1.00), effusion (k = 0.64), mucus plug (k = 0.68), and solid scattered nodularity (k = 0.82). Diffuse disease, including ground-glass opacities (k = 0.57) and tree-in-bud nodules (k = 0.48), were the findings that were most difficult to discern using MRI, with false readings in four of 18 patients for each feature. Nodule size, which was measured independently at CT and MRI, was strongly correlated (R-2 = 0.99) for nodules with a measurement of 10 mm +/- 5 (range, 5-23 mm). Conclusion: This initial study indicates that high-performance 0.55-T MRI holds promise in the evaluation of common lung disease.
引用
收藏
页数:7
相关论文
共 50 条
  • [1] Prospects for 0.55-T MRI in Neuroimaging
    Pai, Vivek
    Maralani, Pejman Jabehdar
    RADIOGRAPHICS, 2025, 45 (03)
  • [2] Routine and Advanced Neurologic Imaging at 0.55-T MRI: Opportunities and Challenges
    Kelsey, Lauren J.
    Seiberlich, Nicole
    Morehouse, Joel
    Richardson, Jacob
    Srinivasan, Ashok
    Bapuraj, Jayapalli
    Kim, John
    Gulani, Vikas
    Mishra, Shruti
    RADIOGRAPHICS, 2025, 45 (03)
  • [3] T2-Weighted Lung Parenchymal Imaging Using a High-Performance Low Field MRI System
    Campbell-Washburn, A. E.
    Bhattacharya, I.
    Moss, J.
    Olivier, K. N.
    Fennelly, K. P.
    Jones, E. C.
    Malayeri, A. A.
    Chen, M. Y.
    AMERICAN JOURNAL OF RESPIRATORY AND CRITICAL CARE MEDICINE, 2020, 201
  • [4] Cardiac Cine MRI Using a Commercially Available 0.55-T Scanner
    Segeroth, Martin
    Winkel, David J.
    Vosshenrich, Jan
    Breit, Hanns-Christian
    Giese, Daniel
    Haaf, Philip
    Zellweger, Michael J.
    Bremerich, Jens
    Santini, Francesco
    Pradella, Maurice
    RADIOLOGY-CARDIOTHORACIC IMAGING, 2024, 6 (04):
  • [5] T2-Weighted Liver MRI Using the MultiVane Technique at 3T: Comparison with Conventional T2-Weighted MRI
    Kang, Kyung A.
    Kim, Young Kon
    Kim, EunJu
    Jeong, Woo Kyoung
    Choi, Dongil
    Lee, Won Jae
    Jung, Sin-Ho
    Baek, Sun-Young
    KOREAN JOURNAL OF RADIOLOGY, 2015, 16 (05) : 1038 - 1046
  • [6] Fast T2-weighted imaging by PSIF at 0.2 T for interventional MRI
    Chung, YC
    Merkle, EM
    Lewin, JS
    Shonk, JR
    Duerk, JL
    MAGNETIC RESONANCE IN MEDICINE, 1999, 42 (02) : 335 - 344
  • [7] High-Performance 0.55-T Lung MRI in Patient with COVID-19 Infection
    Campbell-Washburn, Adrienne E.
    Suffredini, Anthony F.
    Chen, Marcus Y.
    RADIOLOGY, 2021, 299 (02) : E246 - E247
  • [8] Abdominal T2-weighted imaging by free-breath PACE with a 3.0T MRI system
    Moroi, T.
    Mizuuchi, N.
    Maruyama, K.
    Imura, C.
    Isobe, I.
    Usagawa, T.
    Iriguchi, N.
    2007 IEEE/ICME INTERNATIONAL CONFERENCE ON COMPLEX MEDICAL ENGINEERING, VOLS 1-4, 2007, : 740 - 743
  • [9] Kidney morphology in pregnancy using T2-weighted MRI
    Conti-Ramsden, F.
    Hill, R.
    Rutherford, M.
    Railton, N.
    Chappell, L. C.
    Wiles, K.
    CLINICAL RADIOLOGY, 2022, 77 (07) : 548 - 552
  • [10] Oxygen-enhanced functional lung imaging using a contemporary 0.55 T MRI system
    Bhattacharya, Ipshita
    Ramasawmy, Rajiv
    Javed, Ahsan
    Chen, Marcus Y.
    Benkert, Thomas
    Majeed, Waqas
    Lederman, Robert J.
    Moss, Joel
    Balaban, Robert S.
    Campbell-Washburn, Adrienne E.
    NMR IN BIOMEDICINE, 2021, 34 (08)